The Bayer Diagnostics PSA Immunoassay is for the quantitative serial determination of prostate specific antigen in human serum and to aid in the management (monitoring) of patients with prostate cancer.
Device Story
The Bayer Diagnostics PSA Immunoassay is an in vitro diagnostic test used on ACS:180 and ADVIA Centaur Automated Chemiluminescence Systems. It measures total PSA concentrations in human serum samples. The device utilizes a two-site sandwich immunoassay with direct chemiluminometric technology; it employs two antibodies: a polyclonal anti-goat antibody labeled with acridinium ester and a monoclonal anti-mouse antibody covalently coupled to paramagnetic particles. The system detects relative light units (RLUs), which have a direct relationship to the amount of PSA in the patient sample. The assay is performed in a clinical laboratory setting by trained personnel. Healthcare providers use the quantitative results to monitor patients for potential recurrence of prostate cancer after therapy or to assess the effectiveness of ongoing treatments. The device provides a tool for longitudinal monitoring of PSA levels in patients with known prostate cancer.
Clinical Evidence
Bench testing only. Performance data included analytical sensitivity (0.06 ng/mL) and assay range (up to 100 ng/mL). Method comparison studies (n=629 and n=661) showed strong correlation (r=0.986 and 0.990) between the subject assays and alternate methods. Precision studies (within-run, run-to-run, and total CVs) were conducted across multiple systems, with total CVs ranging from 2.6% to 6.3%. Expected results were established by analyzing serum samples from healthy subjects and patients with various malignant and benign prostate diseases.
Technological Characteristics
Two-site sandwich immunoassay; direct chemiluminometric technology. Uses acridinium ester-labeled polyclonal anti-goat antibody and monoclonal anti-mouse antibody coupled to paramagnetic particles. Measures total PSA up to 100 ng/mL. System is an automated chemiluminescence analyzer. Connectivity is integrated into the ACS:180 and ADVIA Centaur systems.
Indications for Use
Indicated for the quantitative serial determination of prostate specific antigen (PSA) in human serum to aid in the management (monitoring) of patients with prostate cancer. Not indicated for use as a screening procedure for the diagnosis of cancer.
Regulatory Classification
Identification
A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.
Special Controls
*Classification.* Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the date "APR 13 2000" in a bold, sans-serif font. The letters and numbers are printed in black ink. The date is likely extracted from a document or photograph, and it represents April 13, 2000.
Image /page/0/Picture/1 description: The image shows a handwritten string of characters, which appears to be a combination of letters and numbers. The string reads 'K994221'. The characters are written in a simple, slightly slanted style, with varying stroke thicknesses. The handwriting is clear and legible against the plain background.
# Summary of Safety and Effectiveness
As required by 21 CFR 807.92, the following 510(k) Summary is provided:
#### 1. Submitter Information
| Contact person: | William J. Pignato |
|-----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| Address:<br>63 North Street<br>Medfield, MA 02052 | Bayer Diagnostics Corporation |
| Phone:<br>FAX:<br>e-mail: | (508) 359-3825<br>(508) 359-3356<br>william.pignato.b@bayer.com |
| Date Summary Prepared: | April 5, 2000 |
| Device Information | |
| Proprietary Name:<br>Common Name:<br>Classification Name: | ADVIA: Centaur and ACS: 180 ePSA Assay<br>PSA Immunoassay<br>Reclassified to Class II, classification numbers |
#### 3. Predicate Device Information
| Name: | Immuno I PSA Immunoassay |
|---------------|--------------------------|
| Manufacturer: | Bayer Diagnostics |
The Immuno 1 PSA Immunoassay was approved by FDA as PMA #P950021 and downclassified to class II by 21 CFR 866.6010, Tumor Associated AntigenImmunological Test System on Dec 17, 1997.
unknown
#### 4. Device Description
2.
PSA is detected in the serum of males with normal, benignhypertrophic, and malignant prostate tissue. PSA is not detected in the serum of males without prostate tissue (because of radical prostatectomy or cystoprostatectomy) or in the serum of most females. The fact that PSA is unique to prostate tissue makes it a suitable marker for monitoring men with cancer of the prostate. PSA is also useful for determining possible recurrence after therapy when used in conjunction with other diagnostic indices.
Measurement of serum PSA levels is not recommended as a screening procedure for the diagnosis of cancer because elevated PSA levels also are observed in patients with benign prostatic hypertrophy. However, studies suggest that the measurement of PSA in conjunction with digital rectal examination (DRE) and ultrasound provide a better method of detecting prostate cancer than DRE alone.
{2}------------------------------------------------
PSA levels increase in men with cancer of the prostate, and after radicalprostatectomy PSA r SA levels increase in firen with cance. If prostatic tissue remains after surgery or levels routinely fall to the and course to be useful in detecting residual and early metastals nas oboured, to read PSA levels can help determine the success of recurrence of turner. Therefore, ourther treatment, such as radiation, endocrine or prostatootonly, and in the monitoring of the effectiveness of therapy.
# 5. Statement of Intended Use
The Bayer Diagnostics PSA Immunoassay is for the quantitative determination of prostate The Bayer Diagnostial PDA minutioused) to formalitoring) of patients prostate cancer specific antigen in seram to alle in the management ADVIA Automated Chemiluminescence Systems.
# 6. Summary of Technological Characteristics
The ACS:180 and ADVIA Centaur PSA assays are a two-site sandwich immunoassay using direct chemiluminometric technology, which uses constant amounts of two antibodies. The direct offermilammomotio to a polyclonal anti-goat antibody labeled with acridinium not ankibody, in the Solid Phase, is a monoclonal anti-mouse antibody, which is covalently coupled to paramagnetic particles.
A direct relationship exists between the amount of PSA present in the patient sample and the amount of relative light units (RLUs) detected by the system
### 7. Performance Data
# Sensitivity and Assay Range
The ACS: 180 PSA and ADVIA Centaur PSA assays measure total PSA concentrations up to 100 ng/mL (100 µg/L) with a minimum detectable concentration (analytical sensitivity) of 0.06 ng/mL (0.06 [g/L). Analytical sensitivity is defined as the concentration of PSA that corresponds to the RLUs that are two standard deviations greater than the mean RLUs of 20 replicate determinations of the PSA zero standard.
#### Accuracy and Method Comparison
For 629 samples in the range of 0.06 to 100 ng/mL (0.06 to 100 رو/L), the relationship between the ACS: 180 PSA assay and an alternate method is described by the equation:
ACS: 180 PSA = 0.98 (alternate method) + 0.0.118 ng/mL
Correlation coefficient (r) = 0.986
For 661 samples in the range of 0.06 to 100 ng/mL (0.06 to 100 ¡q/L), the relationship between the ADVIA Centaur PSA assay and the ACS: 180 PSA assay is described by the equation:
ADVIA Centaur PSA = 0.99 (ACS: 180 PSA) - 0.09 ng/mL
{3}------------------------------------------------
Correlation coefficient (r) = 0.990
#### Expected Results
To confirm the distribution of total PSA in patients, as shown below, serum samples from healthy subjects and patients with various malignant diseases were analyzed using the ACS: 180 PSA reagents. The patients included in this study represent a variety of disease states from active, progressive malignancy to no clinical evidence of disease. The frequency of positive PSA results is significantly lower in patients with no evidence of active disease compared to those with active disease.
| Patient<br>Diagnosis | N | 0.0-4.0<br>ng/mL | 4.1-10<br>ng/mL | 10.1-40<br>ng/mL | >40<br>ng/mL | Median<br>PSA<br>(ng/mL) |
|----------------------------------|-----|------------------|-----------------|------------------|--------------|--------------------------|
| Apparently<br>Healthy | | | | | | |
| Female | 100 | 100.0 | 0.0 | 0.0 | 0.0 | < 0.06 |
| Male < 40 | 71 | 100.0 | 0.0 | 0.0 | 0.0 | 0.73 |
| Male 40-50 | 50 | 100.0 | 0.0 | 0.0 | 0.0 | 0.53 |
| Male 50-60 | 54 | 100.0 | 0.0 | 0.0 | 0.0 | 0.61 |
| Male 60-70 | 50 | 100.0 | 0.0 | 0.0 | 0.0 | 0.85 |
| Male > 70 | 58 | 100.0 | 0.0 | 0.0 | 0.0 | 0.77 |
| Total Males | 283 | 100.0 | 0.0 | 0.0 | 0.0 | 0.71 |
| Prostate<br>Cancer | | | | | | |
| Stage A | 42 | 69.0 | 26.2 | 4.8 | 0.0 | 3.92 |
| Stage B | 50 | 60.0 | 32.0 | 8.0 | 0.0 | 3.52 |
| Stage C | 43 | 20.9 | 72.1 | 4.7 | 2.3 | 5.25 |
| Stage D* | 46 | 56.5 | 21.7 | 19.6 | 2.2 | 3.48 |
| Total Prostate | 191 | 51.6 | 38.0 | 9.3 | 1.1 | 4.04 |
| Benign<br>Diseases | | | | | | |
| Prostate<br>Hypertrophy<br>(BPH) | 152 | 46.7 | 32.9 | 20.4 | 0.0 | 4.37 |
| Genitourinary<br>(GU)a | 50 | 90.0 | 8.0 | 2.0 | 0.0 | 1.38 |
| Prostatitis | 18 | 27.8 | 5.6 | 5.6 | 61.1 | 125.9 |
| Rheumatoid<br>Factor | 5 | 100.0 | 0.0 | 0.0 | 0.0 | 0.58 |
#### % Distribution of PSA by Diagnostic Category
* Six of the these samples were from untreated patients, the remaining samples were patients under treatment
{4}------------------------------------------------
# % Distribution of PSA by Diagnostic Category
| Patient | | 0.0-4.0 | 4.1-10 | 10.1-40 | >40 | Median<br>PSA |
|------------------|----|---------|--------|---------|-------|---------------|
| Diagnosis | N | ng/mL | ng/mL | ng/mL | ng/mL | (ng/mL) |
| Other Cancers | | | | | | |
| Breast | 10 | 100.0 | 0.0 | 0.0 | 0.0 | 0.08 |
| Renal | 6 | 100.0 | 0.0 | 0.0 | 0.0 | 0.37 |
| Pulmonary | 10 | 100.0 | 0.0 | 0.0 | 0.0 | 0.08 |
| Misc. GU | 39 | 92.3 | 5.1 | 2 6 | 0.0 | 0.42 |
| Gastrointestinal | 12 | 91.7 | 0.0 | 0 0 | 8.3 | 0.90 |
| Other | 18 | 100.0 | 0.0 | 0 0 | 0.0 | 0.45 |
#### Precision
,
:
For the ACS: 180 eight samples were assayed 3 times in 6 assays, on each of 4 systems (n = 72 for each samples were abouted of the following results were
shteined; obtained:
| Mean<br>(ng/mL) | Mean<br>(ug/L) | Within-run<br>% CV | Run-to-run<br>% CV | Total<br>% CV |
|-----------------|----------------|--------------------|--------------------|---------------|
| 0.70 | 0.70 | 3.4 | 3.2 | 5.9 |
| 0.91 | 0.91 | 3.4 | 3.6 | 5.3 |
| 1.83 | 1.83 | 2.8 | 3.3 | 5.0 |
| 17.55 | 17.55 | 2.8 | 2.7 | 4.2 |
| 18.23 | 18.23 | 2.9 | 3.1 | 4.6 |
| 29.73 | 29.73 | 3.2 | 3.0 | 5.1 |
| 54.34 | 54.34 | 3.5 | 3.3 | 5.3 |
| 76.25 | 76.25 | 3.7 | 3.4 | 6.3 |
For the ADVIA Centaur six samples were assayed 3 times in 8 runs, on each of 4 systems (n = 24 for each sample), over a period of 3 days. The following results were obtained
| Mean<br>(ng/mL) | Mean<br>(µg/L) | Within-run<br>% CV | Run-to-run<br>% CV | Total<br>% CV |
|-----------------|----------------|--------------------|--------------------|---------------|
| 0.44 | 0.44 | 4.8 | 4.05 | 5.97 |
| .708 | .708 | 3.08 | 2.07 | 3.71 |
| 1.831 | 1.831 | 2.09 | 4.67 | 5.12 |
| 1.934 | 1.934 | 2.08 | 1.56 | 2.60 |
| 11.308 | 11.308 | 2.97 | 3.61 | 4.68 |
| 17.706 | 17.706 | 2.29 | 2.40 | 3.31 |
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features an abstract eagle design with three stylized lines representing the bird's body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle, indicating the department's name and national affiliation.
# APR 1 3 2000
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Mr. William J. Pignato Director or Regulatory Affairs Bayer Diagnostics Corporation 63 North Street Medfield, Massachusetts 02052
Re: K994221
> Trade Name: Bayer Diagnostics Corporation ACS: 180/ADVIA Centaur PSA Assay Regulatory Class: II Product Code: LTJ Dated: March 10, 2000 Received: March 15, 2000
Dear Mr. Pignato:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predical (or the multiple) for the multiplish for use prior to May 28, 1976, the enactment date of the Medical Innersiale United States that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Kegulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FTA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{6}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally parketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Putman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
0
{7}------------------------------------------------
Page _______ of __
510(k) Number (if known): K994221
Device Name: Bayer Diagnostics ACS:180 and ADVIA Centaur PSA Assay
# Indications for Use:
The Bayer Diagnostics PSA Immunoassay is for the quantitative serial determination of prostate specific antigen in human serum and to aid in the managenent (monitoring) of patients with prostate cancer
Qine E. Macher
(Division Sign-Off) Division of Clinical Laboratory Devi 510(k) Number -
(PLEASE DO NOT WRITE BELOW THIS LINE--CONTINUE ON ANOTHER PAGE, IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | |
|----------------------|--|
| Counter Use | |
| (Per 21 CFR 801.109) | |
| Format 1-2-96) | |
OR
Over-The-
(Optional
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.