K994146 · Plus Orthopedics · LWJ · Dec 11, 2000 · Orthopedic
Device Facts
Record ID
K994146
Device Name
EPF-PLUS CEMENTLESS PRESS-FIT ACETABULAR CUP
Applicant
Plus Orthopedics
Product Code
LWJ · Orthopedic
Decision Date
Dec 11, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3360
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The EPF®-PLUS Cementless Press-Fit Acetabular Cup is intended for all types of arthrosis, such as advanced destruction of the hip joint due to degenerative, post-traumatic or rheumatoid arthritis, fracture or avascular necrosis of the femoral head, sequelae of previous operations, such as internal fixation, joint reconstruction, arthrodesis, hemiarthroplasty or total hip replacement. The same considerations apply to acetabular revisions.
Device Story
EPF®-PLUS Cementless Press-Fit Acetabular Cup is a titanium orthopedic implant used in hip replacement or revision surgery. Device is a press-fit cup designed to replace the acetabulum; provides structural support for hip joint. Used by orthopedic surgeons in operating room settings. Implant is made of commercially pure titanium with a flame-sprayed plasma surface treatment to enhance bone apposition. Device is supplied sterile for single use. Surgeon implants cup into prepared acetabulum to restore joint function and alleviate pain associated with arthrosis or previous hip trauma. Benefits include improved joint stability and mobility for patients requiring hip reconstruction.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Acetabular cup made of commercially pure titanium. Materials conform to ASTM F-136-98, ASTM F 67-95 (gr. 4), F 639-93, ISO 5832-2, and ISO 5834-1 & -2. Features flame-sprayed plasma surface treatment. Supplied sterile; terminal sterilization process used.
Indications for Use
Indicated for patients with advanced hip joint destruction due to degenerative, post-traumatic, or rheumatoid arthritis, femoral head fracture or avascular necrosis, or sequelae from prior hip surgeries (e.g., internal fixation, arthrodesis, hemiarthroplasty, total hip replacement). Also indicated for acetabular revisions.
Regulatory Classification
Identification
A hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis is a device intended to be implanted to replace a portion of the hip joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum. This generic type of device includes designs which are intended to be fixed to the bone with bone cement (§ 888.3027) as well as designs which have large window-like holes in the stem of the device and which are intended for use without bone cement. However, in these latter designs, fixation of the device is not achieved by means of bone ingrowth.
{1}------------------------------------------------
DEC 1 1 2000
| | .<br>BUSINESS OF CALL COUNCE CARD |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | |
| | Characterial Comments of consistence and management manus manus manus manus manus manus manus manus marca marca marca marca marca marca marca marca marca marca marca marca ma |
| | a program de manuf manufacturer manument membron grande membre membre member of manufacturer<br>beller annum in to the provincies money con while and manufacture and more and more and more and |
| Analysis the state of the comments of the sense promotion of the more and only and the more of the many of the many of the many of | SEARCH SECTION FOR STATES FOR THE START POST FOR STORE THE CONSTITUTION CONSTITUTION CONSTITUTION CONSTITUTION CONSTITUTION OF COLLECTION COLLECTION COLLECTION COLLECTION COL<br> |
| TELLERS & LED BOOK LES SES SELE LED BELLE LE LE SECH SECORE SECTES SECTES POST AND LEA LEA LEA<br><br>I Page Banka Belle Bron Stan Streem ELEMBER State States | INEAL SERVICE BE LEE LE SE SERECE PARA BER PART CARDARI LE LES PARACE LE SER F SE P . 199 × R M |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | 11 11 1 8 8000 1 110 0000 1<br>. |
# Regulatory Authority: Safe Medical Devices Act of 1990, 21 CFR 807.92
#### 1. Device Name:
EPF®-PLUS Cementless Press-Fit Acetabular Cup Trade Name:
Common Name: Acetabular cup
Classification Name:
PROSTHESIS. HIP. SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED
## 2. Establishment Name & Registration Number:
Plus Orthopedics Name:
Number: 2086141
#### 3. Classification:
No Code Section Reference Available.
Device Class: Class II
Classification Panel: Orthopaedic and Rehabilitation Devices Panel
Product Code: 87LWJ
## 4. Special Controls:
As a Class II medical device, guidance documents and special controls are in effect.
#### Equivalent Legally Marketed Device(s): 5.
- PLUS EPF Acetabular Cup, Plus Orthopedics, K972931 .
The EPF -PLUS Cementless Press-Fit Acetabular Cup is substantially equivalent as a result of the use of identical materials, identical indications for use and patient population, identical size offerings, identical instrumentation, identical production facilities and identical sterifization method. The only difference between the original and the modified device is the application of a flame sprayed plasma surface treatment. The characteristics of the surface treatment are not new and are commonly found on many similar orthopaedic implant devices.
#### 6. Device Description:
The EPF®-PLUS Cementless Press-Fit Acetabular Cup is a surface treated titanium acetabular cup made from commercially pure titanium. The materials conform to national standards including ASTM F-136-98, ASTM F 67-95, gr. 4, F 639-93, ISO 5832-2. & ISO 5834-1 & -2.
#### Indications for Use:
"The EPF®-PLUS Cementless Press-Fit Acetabular Cup is intended for all types of arthrosis, such as advanced destruction of the hip joint due to degenerative, post-fraumatic or rheumatoid arthritis, fracture or avascular necrosis of the femoral head, sequelae of previous operations, such as internal fixation, joint reconstruction, arthrodesis, hemiarthroplasty or total hip replacement. The same considerations apply to acetabular revisions."
{2}------------------------------------------------
#### 7. Applicant Name & Address:
PLUS ORTHOPEDIC 3550 General Atomics Court Building # 15-100 San Diego, CA 92121 858.455.2400 tel - 858.455.2424 fax
#### 8. Company Contact:
Mr. Hartmut Loch Plus ORTHOPEDIC 3550 General Atomics Court Building # 15-100 San Diego, CA 92121 858.455.2400 tel - 858.455.2424 fax
## 9. Submission Correspondent:
Mr. David W. Schlerf Buckman Company, Inc. 200 Gregory Lane, Suite C-100 Pleasant Hill, CA 94553-3389 925.356.2640 vox 925.356.2654 fax
#### 10. Special Guidance Document Information:
Special Guidance documents are in effect for this device, including:
- Premarket Notification [510(k)] Applications for Orthopaedic Devices. .
- Document for Testing Orthopaedic Implants With Modified Metallic Surfaces Apposing Bone or . Bone Cement.
#### 11. Storage, Packaging & Sterilization Information:
The EPF®-PLUS Cementless Press-Fit Acetabular Cups are supplied sterile. All implants are individually packaged & labeled and supplied in boxes for ease in shipping and storage. All packaging should be inspected on arrival for evidence of shipping damage. Any product for use in the operating room must be processed, opened, handled and placed into use following accepted operating room sterile technique. The product is terminally sterilized by an appropriate sterilization process exposure. Sterility Assurance Level (SAL) is at least 100. Once opened, the product must be used or discarded. No attempt should be made at resterilization.
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a caduceus, which is a symbol often associated with medicine and healthcare. The caduceus is depicted with a staff entwined by two snakes and topped with wings.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# DEC 11 2000
PLUS Orthopedics c/o Mr. David W. Schlerf Buckman Company, Inc. 200 Gregory Lane, Suite C-100 Pleasant Hill, California 94523-3389
Re: K994146
EPF®-PLUS Cementless Press-Fit Acetabular Cup Product Code: LWJ, LZO Regulatory Class: II Dated: October 7, 2000 Received: October 17, 2000
# Dear Mr. Schlerf:
We have reviewed your Section 510(k) notification of intent to market the device referenced we nave reviewed your Section 510(x) notifically equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the stated in the elief to devices manotents, or to devices that have been that have been the Your move enactment date of the Medical Dorted Finesal. Food, Drug, and Cosmetic Act (Act). TYou may, therefore, market the device, subject to the general controls provisions of the Act. The general therefore, market the device, subject to the genirements for annual registration, listing of devices, controls provisions or the receive and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations (1 icharter Approval), it they be of Federal Regulations, Title 21, Parts 800 to 895. arrouning your at not rive slent determination assumes compliance with the current Good A substantiany equirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to inspections, the Food and Drig Frankers action. In addition, FDA may publish comply with the Grill Tegatation in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{4}------------------------------------------------
## Page 2 - Mr. David W. Schlerf
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97)." Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Mark McMellerson
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### K994146 510(k) Number:
Device Name(s):
# EPF®-PLUS Cementless Press-Fit Acetabular Cup
## Intended Use:
"The EPF®-PLUS Cementless Press-Fit Acetabular Cup is intended for all types of arthrosis, such as advanced destruction of the hip joint due to degenerative, post-traumatic or rheumatoid arthritis, fracture or avascular necrosis of the femoral head, sequelae of previous operations, such as internal fixation, joint reconstruction, arthrodesis, hemiarthroplasty or total hip replacement. The same considerations apply to acetabular revisions."
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark n Melkerson
(Division Sign-Off) (Division Sign Ora)
Division of General Restorative D 510(k) Number .
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use (Optional format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.