← Product Code [ITX](/productcode/ITX) · K994106

# OLYMPUS UM-S30-25R ULTRASONIC PROBE AND ASSOCIATED ANCILLARY EQUIPMENT FOR BRONCHIAL USE (K994106)

_Olympus America, Inc. · ITX · Jun 2, 2000 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K994106

## Device Facts

- **Applicant:** Olympus America, Inc.
- **Product Code:** [ITX](/productcode/ITX.md)
- **Decision Date:** Jun 2, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1570
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The Olympus UM-S30-25R Ultrasonic Probe has been designed to be used with an Olympus Endoscopic Ultrasound center, Probe Driving Unit and an endoscope for intraluminal ultrasonic imaging of the upper airways and tracheobronchial tree.

## Device Story

The UM-S30-25R is an ultrasonic probe designed for intraluminal imaging of the upper airways and tracheobronchial tree. It functions as part of a system including an Olympus Endoscopic Ultrasound center and a Probe Driving Unit. The probe is inserted through the channel of a flexible or rigid endoscope. A probe driving unit controls the mechanical rotation of the transducer, enabling 360-degree radial B-mode scanning. Imaging is performed using de-aerated water immersion or balloon methods. The device is operated by a physician in a clinical setting. The resulting ultrasound images are displayed on the ultrasound center monitor, allowing the clinician to visualize tissue structures, which aids in clinical assessment and decision-making during bronchoscopic procedures.

## Clinical Evidence

Bench testing only. The device was evaluated for compliance with IEC 60601-1 (electrical safety) and IEC 60601-1-2 (electromagnetic compatibility). No clinical data was required or provided to support the substantial equivalence determination.

## Technological Characteristics

Mechanical radial scanning transducer; 30MHz ultrasonic frequency; 2.4mm outer diameter; 2050mm length. Operates via de-aerated water immersion or balloon method. Connects to EU-M30 Ultrasound Center via MH-240 Probe Driving Unit. Complies with IEC 60601-1 and IEC 60601-1-2 standards.

## Regulatory Identification

A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.

## Predicate Devices

- UM-2R Ultrasonic Probe ([K982323](/device/K982323.md))
- UM-3R Ultrasonic Probe ([K982323](/device/K982323.md))
- EU-M30 Endoscopic Ultrasound Center ([K982610](/device/K982610.md), [K951994](/device/K951994.md), [K982323](/device/K982323.md))
- MH-240 Probe Driving Unit ([K982610](/device/K982610.md), [K951994](/device/K951994.md), [K982323](/device/K982323.md))
- MH-246R Balloon Sheath ([K982733](/device/K982733.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 510(k) SUMMARY OLYMPUS UM-S30-25R ULTRASONIC PROBE AND ASSOCIATED ANCILLARY EQUIPMENT FOR BROCHIAL USE

## A. Submitter's Name, Address, Phone and Fax Numbers

## 1. Manufacturer of the subject devices

Name & Address of manufacturer:

Registration No .: Address, Phone and Fax Numbers: of R&D Department, Endoscope Division

Olympus Optical Co., Ltd. 2-3-1 Shinjyuku Monolis Nishishinjyuku Shinjuku-ku, Tokyo, Japan 8010047 2951 Ishikawa-Cho, Hachioji-shi, Tokyo 192-8507 Japan TEL 0426-42-5101 FAX 0426-46-2786

#### B. Name of Contact Person

Name: Address, Phone and Fax Numbers:

Ms. Laura Storms-Tyler Olympus America Inc. Endoscope Division Two Corporate Center Drive Melville, New York 11747-3157 TEL: (516) 844-5474 FAX: (516) 844-5416

## C. Device Name, Common Name, Classification Name and Predicate Devices

| Device Name:         | Ultrasonic Probe UM-S30-25R<br>Endoscopic Ultrasound Center EU-M30<br>Probe Driving Unit MH-240 |
|----------------------|-------------------------------------------------------------------------------------------------|
| Common Name:         | Ultrasonic Probe                                                                                |
| Classification Name: | 21 CFR892.1570 Diagnostic Ultrasonic transduce                                                  |

{1}------------------------------------------------

| MODEL   | NAME                            | 510(k)#                              |
|---------|---------------------------------|--------------------------------------|
| UM-2R   | Ultrasonic Probe                | K982323                              |
| UM-3R   | Ultrasonic Probe                | K982323                              |
| EU-M30  | Endoscopic Ultrasound<br>Center | K982610(for urinary),                |
|         |                                 | K951994(for gastrointestinal tract), |
|         |                                 | K982323(for bronchial use)           |
| MH-240  | Probe Driving Unit              | K982610(for urinary),                |
|         |                                 | K951994(for gastrointestinal tract), |
|         |                                 | K982323(for bronchial use)           |
| MH-246R | Balloon Sheath                  | K982733 (for UM-2R/3R)               |

Predicate Device:

## D. Description of the Device(s)

The Olympus UM-S30-25R Ultrasonic Probe has been designed to be used with an Olympus Endoscopic Ultrasound center, Probe Driving Unit and an endoscope for intraluminal ultrasonic imaging of the upper airways and tracheobronchial tree. The probe is connected to the endoscopic ultrasound center through the probe driving unit. The probe is attached to the endoscope which allows the user to use probe. The probe is inserted into the patient through a channel of the endoscope. A probe driving unit controls the rotation of the transducer.

<Applicable endoscopes are as follows>

Flexible endoscope: inner channel is \$2.8mm, \$3.2mm, \$6mm and \$5.5mm. Rigid endoscopes : minimum capacity is 9Fr.

UM-S30-25R Ultrasound Probe produces a B-mode scans using the de-aerated water immersion method and balloon method. The probe offers 360 degree mechanical /radial scanning of the tissue under observation. The outer diameter of the insertion tube is 2.4mm and the length is 2050mm.

### E. Intended Use of the Device(s)

The Olympus UM-S30-25R Ultrasonic Probe has been designed to be used with an Olympus Endoscopic Ultrasound center, Probe Driving Unit and an endoscope for intraluminal ultrasonic imaging of the upper airways and tracheobronchial tree.

## F. Summary of the Technological Characteristics of the Device compared to the Predicate Device(s)

Ultrasonic Frequency

The ultrasonic frequency is changed to 30MHz.

{2}------------------------------------------------

## K994106 p.3 of 3

## G. Summary including a Brief Discussion of Non-clinical Tests and How their Results support Determination of Substantial Equivalence

#### 1. Design

Endoscopic Ultrasound Center EU-M30 and Probe Driving Unit MH-240 have been designed, manufactured and tested in compliance with IEC 60601-1 and Revised 510(k) Diagnostics ultrasound Guidance for 1993,1995. It meets the requirements of IEC 60601-1 and Revised 510(k) Diagnostics ultrasound Guidance for 1993,1995.

Ultrasonic Probe UM-S30-25R has been tested to verify electromagnetic compatibility (EMC) with IEC 60601-1-2. It meets the requirements of IEC 6060-1-1-2.

## 2. Materials

The patient contacting materials are identical to predicate device.

## H. Summary including Conclusions drawn from Non-clinical Tests

When compared to the predicate device, the Olympus UM-S30-25R Ultrasonic probe and associated ancillary equipment do not incorporate any significant changes in intended use, method of operation, material, or design that could affect the safety or effectiveness.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure, represented by three curved lines that suggest the shape of a bird in flight.

Public Health Service

JUN - 2 2000

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Laura Storms-Tyler Director, Regulatory Affairs Olympus America, Inc. Two Corporate Center Drive Melville, New York 11747-3157

K994106 Re: UM-S30-25R Ultrasonic Probe (for Bronchial Use) Regulatory Class: II 21 CFR 892.1570/Procode: 90 ITX Dated: March 30, 2000 Received: April 3, 2000

Dear Ms. Storms-Tyler:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranion.

This determination of substantial equivalence applies to the following transducer intended for use with the Olympus EU-M30 Endoscopic Ultrasound Center, as described in your premarket notification:

Probe Model Number

UM-S30-25R Ultrasonic Probe

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic QS inspections, the FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification does not affect any obligation you may have under sections 531 and 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

{4}------------------------------------------------

## Page -2- Ms. Storms-Tyler

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

If you have any questions regarding the content of this letter, please contact Paul M. Gammell, Ph.D. at (301) 594-1212.

Sincerely yours,

Daniel G. Schu

Daniel G. Schultz, M.I Captain, USPHS Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# K994106

510(k) Number (if known):

Olympus UM-S30-25R Ultrasonic Probe and Device Name: associated ancillary equipment for bronchial use

Indications for Use:

The OLYMPUS UM-S30-25R ULTRASONIC PROBE have been designed to be used with an OLYMPUS Endoscopic Ultrasound System for intraluminal sonographic imaging of the upper airways and tracheobronchial tree.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

and the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the contribution of the contribution of th

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use<br>(Per 21 CFR 801.109)                              |         | OR | Over-The-Counter Use     |  |
|-----------------------------------------------------------------------|---------|----|--------------------------|--|
| (Division Sign-Off)                                                   |         |    | (Optional Format 1-2-96) |  |
| Division of Reproductive, Abdominal, ENT,<br>and Radiological Devices |         |    |                          |  |
| 510(k) Number                                                         | K994106 |    |                          |  |

{6}------------------------------------------------

K994106

CONFIDENTIAL

Appendix F

### Diagnostic Ultrasound Indications for Use Form OLYMPUS UM-S30-25R ULTRASONIC PROBE (Subject Device) Intended Use: Diagnostic ultrasound imaging or fluid flow analysis or the human body as follows:

|                                                  | Mode of Operation |   |   |     |     |                      |                          |                                      |                               |                        |
|--------------------------------------------------|-------------------|---|---|-----|-----|----------------------|--------------------------|--------------------------------------|-------------------------------|------------------------|
| Clinical Application                             | A                 | B | M | PWD | CWD | Color<br>Doppl<br>er | Amplit<br>ude<br>Doppler | Color<br>Veloc<br>ity<br>Imagi<br>ng | Combin<br>ed<br>(Speci<br>fy) | Other<br>(Speci<br>fy) |
| Ophthalmic                                       |                   |   |   |     |     |                      |                          |                                      |                               |                        |
| Fetal                                            |                   |   |   |     |     |                      |                          |                                      |                               |                        |
| Abdominal                                        |                   |   |   |     |     |                      |                          |                                      |                               |                        |
| Intraoperative (Specify)                         |                   |   |   |     |     |                      |                          |                                      |                               |                        |
| Intraoperative Neurological                      |                   |   |   |     |     |                      |                          |                                      |                               |                        |
| Pediatric                                        |                   |   |   |     |     |                      |                          |                                      |                               |                        |
| Small Organ (Specify)                            |                   |   |   |     |     |                      |                          |                                      |                               |                        |
| Neonatal Cephalic                                |                   |   |   |     |     |                      |                          |                                      |                               |                        |
| Adult Cephalic                                   |                   |   |   |     |     |                      |                          |                                      |                               |                        |
| Cardiac                                          |                   |   |   |     |     |                      |                          |                                      |                               |                        |
| Transesophageal                                  |                   |   |   |     |     |                      |                          |                                      |                               |                        |
| Transrectal                                      |                   |   |   |     |     |                      |                          |                                      |                               |                        |
| Transvaginal                                     |                   |   |   |     |     |                      |                          |                                      |                               |                        |
| Transurethral                                    |                   |   |   |     |     |                      |                          |                                      |                               |                        |
| Intravascular                                    |                   |   |   |     |     |                      |                          |                                      |                               |                        |
| Peripheral Vascular                              |                   |   |   |     |     |                      |                          |                                      |                               |                        |
| Laparoscopic<br>Musculo-skeletal<br>Conventional |                   |   |   |     |     |                      |                          |                                      |                               |                        |
| Musculo-skeletal Superficial                     |                   |   |   |     |     |                      |                          |                                      |                               |                        |
| Other (Specify)                                  |                   | N |   |     |     |                      |                          |                                      |                               |                        |

N = new indication; P = previously cleared by FDA; E = added under Appendix E  

Additional Comments:the upper airway and (bronchial use),

ﻨﺘ

(PLEASE DO NOT WRITE BELOW THE LINE-CONTIINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evalution (ODE)

Prescription Use (Per 21 CFR 801.109)

E-3

Yamil A. de Leon

(Division Sign Off)

(Division Sign-Off) Division of Reproductive, Abdominal, EN and Radiological Devices 510(k) Number

{7}------------------------------------------------

## K 994106

Image /page/7/Picture/1 description: The image shows the word "CONFIDENTIAL" in all caps, printed in bold black font. The word is enclosed in a rectangular box, also in black. The image appears to be a stamp or label indicating the confidential nature of a document or information.

bendix F

## Diagnostic Ultrasound Indications for Use Form OLYMPUS EU-M30 ENDOSCOPIC ULTRASOUND CENTER (ancillary system) Intended Use: Diagnostic ultrasound imaging or fluid flow analysis or the human body as follows:

|                                                  | Mode of Operation |   |   |     |     |                      |                              |                                      |                               |                        |
|--------------------------------------------------|-------------------|---|---|-----|-----|----------------------|------------------------------|--------------------------------------|-------------------------------|------------------------|
| Clinical Application                             | A                 | B | M | PWD | CWD | Color<br>Doppl<br>er | Amplit<br>ude<br>Dopple<br>r | Color<br>Veloc<br>ity<br>Imagi<br>ng | Combin<br>ed<br>(Speci<br>fy) | Other<br>(Speci<br>fy) |
| Ophthalmic                                       |                   |   |   |     |     |                      |                              |                                      |                               |                        |
| Fetal                                            |                   |   |   |     |     |                      |                              |                                      |                               |                        |
| Abdominal                                        |                   |   |   |     |     |                      |                              |                                      |                               |                        |
| Intraoperative (Specify)                         |                   |   |   |     |     |                      |                              |                                      |                               |                        |
| Intraoperative Neurological                      |                   |   |   |     |     |                      |                              |                                      |                               |                        |
| Pediatric                                        |                   |   |   |     |     |                      |                              |                                      |                               |                        |
| Small Organ (Specify)                            |                   |   |   |     |     |                      |                              |                                      |                               |                        |
| Neonatal Cephalic                                |                   |   |   |     |     |                      |                              |                                      |                               |                        |
| Adult Cephalic                                   |                   |   |   |     |     |                      |                              |                                      |                               |                        |
| Cardiac                                          |                   |   |   |     |     |                      |                              |                                      |                               |                        |
| Transesophageal                                  |                   | P |   |     |     |                      |                              |                                      |                               |                        |
| Transrectal                                      |                   | P |   |     |     |                      |                              |                                      |                               |                        |
| Transvaginal                                     |                   |   |   |     |     |                      |                              |                                      |                               |                        |
| Transurethral                                    |                   | P |   |     |     |                      |                              |                                      |                               |                        |
| Intravascular                                    |                   |   |   |     |     |                      |                              |                                      |                               |                        |
| Peripheral Vascular                              |                   |   |   |     |     |                      |                              |                                      |                               |                        |
| Laparoscopic<br>Musculo-skeletal<br>Conventional |                   |   |   |     |     |                      |                              |                                      |                               |                        |
| Musculo-skeletal Superficial                     |                   |   |   |     |     |                      |                              |                                      |                               |                        |
| Other (Specify)                                  |                   | P |   |     |     |                      |                              |                                      |                               |                        |

N = new indication; P = previously cleared by FDA; E = added under Appendix EN = new indication; P = previously cleared by FDA; E = added under Appendix E Additional Comments:

Previously cleared indication : Intraluminal ultrasound for gastrointestinal tract and surround organs, upper air and tracheobronchial tree and urinary tract.

(PLEASE DO NOT WRITE BELOW THE LINE-CONTIINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evalution (ODE)

Prescription Use (Per 21 CFR 801.109)

F-3

David li. Sigomm

(Division Sign-Off) (Division Sign-Off)
Division of Reproductive, Abdominal, ENT and Radiological Dev 510(k) Number

---

**Source:** [https://fda.innolitics.com/device/K994106](https://fda.innolitics.com/device/K994106)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
