K994101 · Techlab, Inc. · KHW · Dec 21, 1999 · Microbiology
Device Facts
Record ID
K994101
Device Name
E. HISTOLYTICA II
Applicant
Techlab, Inc.
Product Code
KHW · Microbiology
Decision Date
Dec 21, 1999
Decision
SESE
Submission Type
Special
Regulation
21 CFR 866.3220
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The E. histolytica II is an enzyme immunoassay for the rapid detection of the adhesin of E. histolytica in human fecal specimens. It is indicated for use with fecal specimens from patients with diarrhea or dysentery to determine the presence of E. histolytica gastrointestinal infection. The test can be used for fecal specimens submitted for routine clinical testing from adults or children. Conventional microscopy is not a routine prerequisite for use of the test. FOR IN VITRO DIAGNOSTIC USE.
Device Story
E. histolytica II is an ELISA-based in vitro diagnostic test for detecting E. histolytica adhesin (galactose-inhibitable lectin) in human stool. The device utilizes microtiter wells coated with polyclonal goat antibodies to capture the adhesin, followed by a monoclonal mouse antibody conjugate for detection. The test is performed in clinical laboratories as a rapid, sensitive alternative to time-consuming culture and zymodeme analysis. By specifically identifying pathogenic E. histolytica without cross-reacting with nonpathogenic E. dispar, the device assists clinicians in diagnosing amebiasis. Results are interpreted by laboratory personnel to guide clinical management of patients presenting with diarrheal symptoms.
Clinical Evidence
Clinical evaluation compared E. histolytica II results against zymodeme analysis of cultured Entamoeba isolates and the predicate E. histolytica TEST. Study demonstrated >97% correlation with zymodeme analysis. The device showed higher sensitivity than the predicate for detecting E. histolytica trophozoites and adhesin in stool specimens from endemic areas.
Technological Characteristics
ELISA-based immunoassay. Components: microtiter wells coated with polyclonal goat antibody, monoclonal mouse antibody conjugate, positive control, diluent, substrate, wash solution, and stop solution. Standalone diagnostic kit for laboratory use.
Indications for Use
Indicated for adults and children with diarrhea or dysentery to detect E. histolytica adhesin in fecal specimens to diagnose gastrointestinal infection.
Regulatory Classification
Identification
Entamoeba histolytica serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Entamoeba histolytica in serum. Additionally, some of these reagents consist of antisera conjugated with a fluorescent dye (immunofluorescent reagents) used to identify Entamoeba histolytica directly from clinical specimens. The identification aids in the diagnosis of amebiasis caused by the microscopic protozoan parasite Entamoeba histolytica and provides epidemiological information on diseases caused by this parasite. The parasite may invade the skin, liver, intestines, lungs, and diaphragm, causing disease conditions such as indolent ulcers, an amebic hepatitis, amebic dysentery, and pulmonary lesions.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.
{1}------------------------------------------------
12/21/99 K994101
SUMMARY OF SAFETY AND EFFECTIVENESS (**E. histolytica II**)
TE
E
#### 1. Name of Manufacturer
TechLab. Inc. Corporate Research Center 1861 Pratt Drive, STE 1030 Blacksburg, VA 24060-6364
# 2. Establishment Registration
Federal ID # 54-1527427 Initial Registration of Medical Device Establishment, #1122855
## 3. Trade Name
- E. histolytica II
## 4. Common Name
E. histolytica ELISA
# 5. Class of Device
This device is classified in Class II.
# 6. Performance Standards
No performance standards have been developed for this device under 514 of the Food, Drug, and Cosmetic Act.
# 7. Safety and Effectiveness
The E. histolytica II can be used to detect adhesin (also referred to as galactose-inhibitable lectin) produced by strains of E. histolytica. It does not cross-react with the adhesin from E. dispar (formerly known as nonpathogenic E. histolytica). The test can be used to detect the adlaesip in fest specimens from persons suspected of having amebiasis. The kit, which includes ready-to-use reagents, contains microtiter wells coated with polycloual antibody, positive control reagent, monclonal-antibody conjugate, diluent, one component substrate, wash solution, and stop solution, The microtiter wells coated with polyclonal antibody "capture" the adhesin and the monoclonal anthologyconjugate serves as the "detecting" antibody. The polyclonal antibody used to monthcound is prepared from hyperimmune antiserum developed in goats. The monoclonal antibody wedia prepare the conjugate is prepared from mouse ascites fluid
{2}------------------------------------------------
#### SUMMARY OF SAFETY AND EFFECTIVENESS (E. histolytica II) (cont'd)
The E. histolytica II is to be used in an ELISA format and is substantially equivalent to culturing and zymodeme analysis that are used in some clinical laboratories as diagnostic aids. It also is more sensitive than our E. histolytica TEST (K955895), which has been cleared previously for in vitro diagnostic use. Culturing is used to obtain the isolate and zymodeme analysis is used to examine the enzyme profile of the isolate. Zymodeme analysis must be used to determine if the isolated Entamoeba strain is pathogenic. The major disadvantages are that culturing and zymodeme are timeconsuming and labor-intensive. Only a few clinical laboratories around the world are capable of performing this type of analysis. The E. histolytica II offers a major advantage to clinical laboratories because it is rapid, easy-to-perform, and it is more sensitive than our E. histolytica TEST, which was the first test to offer clinical labs a simpler and easier alternative format than zymodeme analysis for the specific detection of pathogenic E. histolytica. Like the E. histolytica TEST, the E. histolytica II is highly specific for pathogenic Entamoe a
The E. histolytica II is different from TechLab's Entumoeba TEST and the Alexon ProSpecT Entamoeba histolytica Test. These two tests detect both E. histolytica (formerly referred to as pathogenic E. histolytica) and E. dispar (formerly referred to as nonpathogenic E. histolytica). They do not distinguish between E. histolytica and E. dispar. The E. histolytica II is very similar to TechLab's E. histolytica TEST. Both are highly specific for the adhesin of E. histolytica and neither cross-reacts with the adhesin of E. dispar. Both have the same procedure and utilize an ELISA format. The difference between the E. histolytica II and the E. histolytica TEST is that the E. histolytica II offers increased sensitivity. Although all of these tests serve as diagnostic aids for amebiasis, the E. histolytica II is specific for pathogenic strains and offers higher sensitivity.
The E. histolytica II was used to analyze stool specimens in areas where amebiasis is endemic, and the results were compared with zymodeme analysis of Entamoeba isolates cultured from these specimens and with the E. histolytica TEST. It is important to remember that culture, without the aid of zymodeme analysis, does not distinguish between E. dispar. Zymodeme analysis, which is available only in a sclect number of clinical laboratories around the world, is the only method that distinguishes these species. For the purpose of our studies, zymodeme analysis represents the "gold standard". The results of our clinical evaluations show that the E. histolvica II exhibits a correlation of >97% when compared with zymodeme analysis, and that it is more sensitive than the E. histolytica Test for detecting E. histolytica trophozoites and adhesin. Our findings demonstrate that the test is useful for the detection of pathogenic E. histolytica in stool specimens.
\$\beta\$
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle with three curved lines representing its wings. The eagle is positioned above the acronym "HHS". The text "OF HEALTH & HUMAN SERVICES - USA" is arranged in a semi-circle above the eagle.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
DEC 2 1 1999
David M. Lyerly, Ph.D. Vice President TechLab. Inc. VPI Research Park 1861 Pratt Drive, Suite 1030 Blacksburg, Virginia 24060-6364
Re: K994101 Trade Name: E. histolytica II Regulatory Class: II Product Code: KHW Dated: November 22, 1999 Received: December 6, 1999
#### Dear Dr. Lyerly:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will veify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Eederal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Javs or regulations.
{4}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its to lifee number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven Autman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
2
{5}------------------------------------------------
K 494101
Page__________________________________________________________________________________________________________________________________________________________________________
510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________ E. histolytica II
Device Name:_
Indications For Use:
The E. histolytica II is an enzyme immunoassay for the rapid detection of the adhesin of E. histolytica in human fecal specimens. It is indicated for use with fecal
adhesin of E. histolytica in human fecal speciments asteming the presence of E adhesin of E. Alstolyned in illulian lecal specifical or dysentery to determine the presence of E.
specimens from patients with diarrhead or dysenters of essed for facel subs specimens from patients with diamica the cysent for tecal specimens submitted for
histolytica gastrointestinal infection. The test can becases miss not a routine clinical testing from adults or children. Conventional microscopy is not a
routine clinical testing from adults or children. Conventional microscopy is not a
routine prerequisite for use of the test. FOR IN VITRO DIAGNOSTIC USE.
| riller in the first for the first for the first for the first for the first for the film<br>ﻣﻨﺬ ﺑﻨﻴﺔ ﺍﻟﺘﻲ ﺗﻘﺮﻳﺒﺎ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ<br>، | 0<br>De<br>C | F<br>D<br>A<br>/<br>C<br>D |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|----------------------------|
| <br>, | ్రిక్ | f (<br>H<br>/<br>0 |
| | | D<br>E |
| 产品<br>ប្រ | 0<br>i | f<br>D<br>H<br>C |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | | |
|--------------------------------------------------------|---------|--------------------------|
| Wordy Dators | | |
| (Division Sign/Off) | | |
| Division of Clinical Laboratory Devices | | |
| 510(k) Number | K994101 | |
| Prescription Use | OR | Over-The-Counter Use |
| (Per 21 CFR 801.109) | | (Optional Format 1-2-96) |
2, \$k-13
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.