K993942 · Laser Dental Innovations · GEX · Jan 18, 2000 · General, Plastic Surgery
Device Facts
Record ID
K993942
Device Name
LITESABER 2000
Applicant
Laser Dental Innovations
Product Code
GEX · General, Plastic Surgery
Decision Date
Jan 18, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LiteSaber™ 2000 indications for use are for the cleared indications for the laser system to which it is attached. For example, dermatological, ENT, general surgery, GYN surgery, oral/dental, maxillofacial, orthopaedic and podiatric procedures for the incision, excision, ablation, vaporization and hemostasis of soft tissue.
Device Story
LiteSaber™ 2000 is an after-market surgical laser accessory; functions as a replacement handpiece or tip for existing surgical laser systems. Device directs laser beam by clasping clad fiber (200-400 microns) and bending distal bare fiber into smooth radius; features 360-degree rotation for lateral or angled beam delivery. No internal optics; does not affect laser spot size or transmission. Used by clinicians in surgical settings (general, plastic, ENT, GYN, dental, orthopaedic, podiatric). Handpiece is reusable (steam autoclave); tips are disposable. Benefits include improved access and beam directionality without requiring refractive/reflective contact tips.
Clinical Evidence
No clinical data. Substantial equivalence is based on design, material, and functional comparison to the predicate device.
Technological Characteristics
Mechanical laser handpiece accessory; reusable handpiece assembly; disposable pre-curved tips. Accommodates 200-400 micron diameter clad laser fibers. Sterilization: steam autoclave (270°F for 30 minutes) for handpiece; tips are single-use. No internal optics, electronics, or software.
Indications for Use
Indicated for use in general, plastic, dermatologic, orthopaedic, intraoral, maxillo-facial, and dental soft tissue surgery, including marginal and interdental gingiva, for incision, excision, ablation, vaporization, and hemostasis. Compatible with existing surgical laser systems using clad fibers (200-400 microns diameter).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Regulatory Authority: Safe Medical Devices Act of 1990, 21 CFR 807.92
Image /page/0/Picture/3 description: The image shows a sequence of handwritten characters and numbers. The sequence starts with the letter 'K', followed by the number '9' twice, then the number '3', another '9', the number '4', and ends with the number '2'. The handwriting style appears casual and slightly uneven.
- 1. Device Name:
| Device Trade Name: | LiteSaber™ 2000 |
|----------------------|-------------------------------------|
| Common Name: | Laser accessory |
| Classification Name: | Laser Instrument, Surgical, Powered |
#### Establishment Name & Registration Number: 2.
Name: Laser Dental Innovations, Inc. Number: Pending
#### Classification: 3.
| § 878.4810 | Laser surgical instrument for use in general and plastic surgery and in dermatology. |
|------------|--------------------------------------------------------------------------------------|
| § 874.4500 | Laser, ENT |
| § 874.4490 | Laser, Otolarngology |
| § 884.4550 | Laser, Surgical, Gynecological |
Device Class: Class II General and Plastic Surgery, OB/GYN, ENT, Dermatology Device Panels Classification Panel: Product Code(s): 79GEX, 77EWG, 77LMS, 85HHR respectively
- 4. Equivalent / Comparison Device:
- ー Novatip Family of Laser Handplece Tips, K971539 by Xanova Corp.
The Licesaber " 2000 tips function in the same way as the Novatip Laser Handpiece Thos. The Lites aber " 2000 operates with the same size fibers, and does not affect the spot size or the function of the laser beam. The materials used to construct the Licess ber 201 2000 are the same or are the functional equivalent of those used in the Novatip handpeice. The Licessaber™ sooo does not have a direct impact on the transmission of the laser energy because the handpiece includes no optics. The Lites aber 14 2000 is used to grasp and direct the laser fiber in clinically useful ways. Thus, it is believed that the Lices aber 14 2000 offers even less concern than the Novatip.
#### 10. Device Description:
Background. The Licesaber™ 2000 is an "after-market" accessory for use with an existing surgical laser system. The device may be used as a complete system (handpiece and tips) or just the tips may be used with an existing handpiece of the ultradent style. The LiceSaber™ 2000 is used as a direct replacement for the laser handpiece supplied as original equipment or the original equipment laser handpiece tip. The Licesaber™ 2000 is designed address the need to direct the beam in a direction other than straight through the handpiece. Further, it does so without the need for refractive or reflective contact tips. The Licesaber 2000 gently and effectively clasps the clad fiber while bending the distal bare fiber into a smooth radius. Pre-curved disposable tips channel the fiber. Precurved tips are available. The directional tips may be rotated 360 degrees. This allows the laser fiber to be directed laterally, up, down or any angle or radius in between. The tips are disposable and the handpiece is reusable.
{1}------------------------------------------------
The Licesaber™ aooo laser handpiece is intended for use in general surgery, plastic surgery, dermatologic surgery, orthopaedic surgery, intraoral, maxillo-facial or dental soft tissue surgery including the marginal and interdental gingiva. The handpiece is "universal" in nature and is intended to be used in place of the handpiece supplied or provided by the laser manufacturer. The LiceSaber™ 2000 is designed to accommodate all clad laser fibers between 200 and 400 microns diameter.
Cleared indications for Use. The Licessber™ 2000 indications for use are for the cleared indications for the laser system to which it is attached. For example, dermatological, ENT, general surgery, GYN surgery, oral/dental, maxillofacial, orthopaedic and podiatic procedures for the incision, excision, ablation, vaporization and hemostasis of soft tissue.
# 5. Applicant / Sponsor Name / Address:
Laser Dental Innovations 745 Dubanski Drive San Jose, CA 95123 408.227.4674
# 6. Company Contact:
Mr. Howard Feinberg Laser Dental Innovations 745 Dubanski Drive San Jose, CA 95123 408.227.4674
### 7. Submission Correspondent:
Mr. David W. Schlerf Buckman Company, Inc. 200 Gregory Ln., Suite C-100 Pleasant Hill, CA 94523-3389 925,358,2640 vox - 925,356.2654 fax
# 8. Sterilization Information:
The laser handpiece may be sterilized and/or re-sterilized until replacement is needed. After cleaning and inspection, standard steam autoclave processing at 270 dearees F. for 30 minutes will produce a sterility assurance level (SAL) of 10°. The tips are disposable and may not be resterilized.
The handpiece assembly must be washed, cleaned and rinsed thoroughly before resterilization. Hospital grade soap/detergent & water followed by sufface scrubbing are employed. Refer to cleaning and sterilization instructions supplied with the device. Thoroughly rinse the handpiece in clear running water. Sterifize un-assembled.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or other bird-like figure, with its wings spread and head facing left. The bird is composed of several curved lines, giving it a modern and abstract appearance.
Public Health Service
JAN 18 2000
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. David Schlerf Laser Dental Innovations c/o Buckman Company, Inc. 200 Gregory Lane Suite C-100 Pleasant Hill, California 94523-3389
K993942 Re:
> Trade Name: LiteSaber™ 2000 Regulatory Class: II Product Code: GEX Dated: October 11, 1999 Received: November 19, 1999
Dear Mr. Schlerf:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed
{3}------------------------------------------------
Page 2 - Mr. David Schlerf
predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Stupto Rluvels
Laurence E. Dillard III
James E. Dillard III Acting Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 ----------------------------------------------------------------------------------------------------------------------------------------------------------------------of
K993942 510(k) Number:
LiteSaber™ 2000 Device Name:
Indications for Use:
The Lite Saber™ 2000 indications for use are for the cleared indications for the laser system to which it is attached. For example, dermatological, ENT, general surgery, GYN surgery, oral/dental, maxillofacial, orthopaedic and podiatric procedures for the incision, excision, ablation, vaporization and hemostasis of soft tissue.
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY
Concurrence of CDRH, Office of Device Evaluation (ODE)
Stypt Rhodes
Division Sign-Off Division of General Restorative Devices K993942 510(k) Number
Prescription Use (Per 21 CFR 801.109) OR
Over - The - Counter Use ___ (Optional format 1-2-96)
LiteSaberResp.doc
r
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.