WAKO IGM II- HA, IMMUNOLOBULIN CALIBRATOR SET, IMMUNOLOBULIN STANDARD
K993924 · Wako Chemicals USA, Inc. · CFN · Jan 14, 2000 · Immunology
Device Facts
Record ID
K993924
Device Name
WAKO IGM II- HA, IMMUNOLOBULIN CALIBRATOR SET, IMMUNOLOBULIN STANDARD
Applicant
Wako Chemicals USA, Inc.
Product Code
CFN · Immunology
Decision Date
Jan 14, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5510
Device Class
Class 2
Intended Use
The Wako IgM II - HA test is an in vitro assay for the quantitative determination of immunoglobulin M in serum.
Device Story
Wako IgM II-HA is an in vitro turbidimetric immunoassay for quantitative measurement of IgM in human serum. Sample is mixed with buffer and anti-human IgM antibody; specific antigen-antibody binding creates insoluble aggregates, increasing turbidity. Turbidity is measured optically; signal is proportional to IgM concentration. Used in clinical laboratory settings by trained personnel. Output provides quantitative IgM levels to assist clinicians in diagnosing and monitoring liver disease, infections, and immune disorders. Benefits include improved sensitivity, precision, and faster assay turnaround compared to traditional single radial immunodiffusion (SRID) methods.
Clinical Evidence
Bench testing only. Performance was validated through comparison studies against the predicate device (Wako IgM HA-Direct), yielding a correlation coefficient of 0.996 and a regression equation of y = 0.662x - 6.64. Precision studies demonstrated acceptable day-to-day reproducibility. The minimum detectable level is 10 mg/dL.
Technological Characteristics
Turbidimetric immunoassay. Reagents include buffer solution and anti-human IgM antibodies. Measurement principle relies on optical detection of turbidity resulting from antigen-antibody complex formation. Standalone in vitro diagnostic assay.
Indications for Use
Indicated for the quantitative determination of immunoglobulin M (IgM) in human serum to aid in the diagnosis, monitoring, and prognosis of conditions such as chronic liver diseases, infectious diseases, lymphocytosis, multiple myeloma, and primary/secondary immune failure.
Regulatory Classification
Identification
An immunoglobulins A, G, M, D, and E immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the immunoglobulins A, G, M, D, an E (serum antibodies) in serum. Measurement of these immunoglobulins aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.
Predicate Devices
Wako IgM HA-Direct
Related Devices
K993928 — WAKO-IGG II-HA, IMMUNOGLOBULIN CALIBRATOR SET, IMMUNOLOBULIN STANDARD · Wako Chemicals USA, Inc. · Jan 18, 2000
K993927 — WAKO IGA II-HA, IMMUNOGLOBULIN CALIBRATOR SET, IMMUNOGLOBULIN STANDRAD · Wako Chemicals USA, Inc. · Jan 14, 2000
{0}------------------------------------------------
JAN 1 4 2000
VO. 2797 P. 11
K993924
Image /page/0/Picture/3 description: The image contains the word "Wako" in a bold, sans-serif font. The text is white and set against a black background. The letters are slightly condensed, giving the word a compact appearance.
Wako Chemicals USA, Inc.
1600 Bellwood Road, Richmond, VA 23237 U.S.A.
Telephone (804) 271-7677 Telex 293208 WAKO UR(RCA) Facsimile (804) 271-7791
### 510(K) Summary of Safety and Effectiveness
The Wako IgM II - HA test is an in vitro assay for the quantitative determination of immunoglobulin M in serum,
#### Summary:
Immunoglobulin M (IgM) is the most primitive immunoglobulin and the only immunoglobulin that a neonate synthesizes. In adult serum, it accounts for 5 to 10% of the total circulating immunglobulins. Most of the serum IgM is a pentamer of five IgM monomers. Each monomer has a molecular weight of 185,000. The monomers are attached to each other with the J chain. IgM is an efficient complement activator, has high affinity to bacteria and erythrocytes, and plays a major role in protection from Gram-negative bacterial infection.
The quantification of immunoglobulins in serum is important for the diagnosis, monitoring and prognosis of chronic liver diseases, infectious diseases, lymphocytosis, multiple myeloma, primary and secondary immune failure, etc. The conventional test, single radial immunodiffusion (SRID), has been widely used. In recent years, there have been many reports on the use of turbidity or light scattering for the measurement of antigen-antibody complexes formed in solution. Advantages over conventional methods include increased sensitivity, better precision, and shortened assay time. The Wako IgM II-HA test is a highly specific reagent based on turbidimetric immunoassay. 112
#### Principle:
When a sample is mixed with the Buffer solution and Anti-IgM, IgM in the sample combines specifically with anti-human IgM antibody in the Anti-IgM to yield an insoluble aggregate that causes increased turbidity. The degree of turbidity can be measured optically and is proportional to the amount of IgM in the sample.
The safety and effectiveness of the Wako IgM II-HA is demonstrated by its substantial equivalency to Wako IgM HA-Direct product. Both test systems arc used to measure IgM in serum. In comparison studies against the predicate assay, a correlation coefficient of 0.996 and a regression equation of y = 0.662 x - 6.64 was obtained. Precision studies indicate acceptable values can be obtained on a day to day basis. The minimum detectable level of this method is 10 mg/dL.
#### References:
1. Burtis, C.A. and Ashwood, E.R., Ed.: Tietz Textbook of Clinical Chemistry, 2nd Ed., Saunders, Philadelphia, 1994.
2. Tsubaki, K. et al., Japanese J. Clin. Chem., 14,185-191 (1985).
3. DG Klinische Chemie Mitteilungen 26 (1995) Heft 5.
laTina Mallum
September 30, 1999
LaTonya Mallory Wako Diagnostics 1600 Bellwood Road, Richmond, VA 23237
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the top half of the circle. Inside the circle is a stylized image of three wavy lines, which are meant to represent the department's mission of promoting health and well-being.
## JAN 1 4 2000
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Tonya Mallory Senior Manager Wako Chemicals USA. Inc. 1600 Bellwood Road Richmond, Virginia 23237
Re: K993924 Trade Name: Wako IgM II HA Immunoglobulin Calibrator Set Regulatory Class: II Product Code: CFN Dated: September 30, 1999 Received: November 18, 1999
Dear Ms. Mallory:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of-the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{2}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven Autman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
of Page_
| 510(k) Number (if known): | K993924 |
|---------------------------|-----------|
| Device Name: | IgM TI HA |
# Indications For Use:
Measurement . ભાંવેડ ingu0515 body's metabolism and five to vesist intections ager
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation
Titus E. Malone
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K993924
Prescription Use_ (Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.