← Product Code [LZO](/productcode/LZO) · K993889

# EXETER CERAMIC FEMORAL HEADS (K993889)

_Howmedica Osteonics Corp. · LZO · Feb 11, 2000 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K993889

## Device Facts

- **Applicant:** Howmedica Osteonics Corp.
- **Product Code:** [LZO](/productcode/LZO.md)
- **Decision Date:** Feb 11, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3353
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Exeter Ceramic Femoral Heads are intended to be used with modular femoral components in primary and secondary cementless total hip replacement procedures. These devices are intended to articulate with a polyethylene cup or a metal backed polyethylene cup component to reconstruct painful and/or severely disabled hip joints resulting from osteoarthritis, rheumatoid arthritis, post-traumatic arthritis, avascular necrosis, subcapital fracture, or a revision of a failed femoral prosthesis. The Exeter Ceramic Femoral Heads are intended to be used with the femoral components of the previously released Exeter hip stems - and Exeter II Total Hip System (K891454).

## Device Story

Modular femoral head component for total hip replacement; affixes to femoral stem; articulates with polyethylene or metal-backed polyethylene acetabular cup. Used in primary/secondary cementless hip arthroplasty. Operates by restoring joint function in patients with degenerative or traumatic hip conditions. Surgeon-implanted in clinical/OR setting. Modification of existing zirconia head design involving slight change in taper angle. Provides mechanical articulation surface for hip reconstruction.

## Clinical Evidence

Bench testing only. Ultimate compression strength testing performed; all samples exceeded 46 kN, meeting FDA Guidance Document requirements for Ceramic Ball Hip Systems.

## Technological Characteristics

Modular femoral head; Zirconia material; 5° 43' taper; designed for articulation with Exeter hip stems. Mechanical device; no software or electronics.

## Regulatory Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

## Predicate Devices

- V40™ Zirconia Femoral Heads ([K972361](/device/K972361.md))
- Exeter II Total Hip System ([K891454](/device/K891454.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K993889

FEB 1 1 2000

## 510(k) Summary

## Device: Exeter Ceramic Femoral Heads

This device is a modular femoral head component which is affixed to a femoral stem component and articulates with a polyethylene acetabular cup or a metal backed polyethylene acetabular cup to reconstruct painful and/or severely disabled hip joints resulting from osteoarthritis, theumatoid arthritis, post-traumatic arthritis, avascular necrosis, subcapital fracture or a revision of a failed femoral prosthesis. The Exeter Ceramic Femoral Heads described in this submission are a modification of a previously cleared Zirconia head (K972381) and Orthinox® head (K891454). The modification is a slight change in the taper angle of the device.

The substantial equivalence is based on an equivalence in intended use, materials, design, operational principles, and relative indications and contraindications to devices currently in commercial distribution including: V40™ Zirconia Femoral Heads (K972361) and Exeter II Total Hip System (K891454).

All of the named devices are intended to be used as the modular femoral head component of a total hip replacement. The basic design of these devices is generally the same, with varying diameters and lengths of internal tapers to accommodate individual patient needs. The material used in the manufacture of these femoral heads is also the same as the V40™ Zirconia Femoral Heads.

Assembly of all the named femoral head components to an appropriate femoral stem component requires similar instrumentation and preparation. All of the named Exeter Ceramic Femoral Heads are intended to articulate with the femoral components of the previously released Exeter Orthinox® (Rex 734 stainless steel) hip sterns with a 5° 43' taper. Relative indications and contraindications for all of the zirconia heads named are the same.

Testing of the Exeter Ceramic Femoral Heads included ultimate compression strength testing. All heads tested for ultimate compression strength exceeded loads greater than 46 kN as specified in the FDA Guidance Document for the Proparation of Premarket Notifications for Ceramic Ball Hip Systems.

For information contact:

Nancy J. Rieder Rutherford Regulatory Affairs Howmedica Osteonics Corp. 359 Veterans Boulevard Rutherford, NJ 07070 Phone: Fax: (201) 507-6870 E-mail: NRieder@HowOst.com

{1}------------------------------------------------

Public Health Service

FEB 1 1 2000

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Elizabeth A. Staub Vice President Quality Assurance, Regulatory Affairs and Clinical Research Howmedica Osteonics Corporation 359 Veterans Boulevard Rutherford, New Jersey 07070

Re: K993889 Trade Name: Exeter Ceramic Femoral Heads Regulatory Class: II Product Code: LZO Dated: November 11, 1999 Received: November 16, 1999

Dear Ms. Staub:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general control provisions of the Act. The general control provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls, Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (OS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

{2}------------------------------------------------

Page 2 - Ms. Elizabeth A. Staub

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

James E. Dillard III Acting Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

Page 1 of 1

## 510(k) Number (if known): K993889

Device Name: Exeter Ceramic Femoral Heads

Indications for Use:

The Exeter Ceramic Femoral Heads are intended to be used with modular femoral components in primary and secondary cementless total hip replacement procedures. These devices are intended to articulate with a polyethylene cup or a metal backed polyethylene cup component to reconstruct painful and/or severely disabled hip joints resulting from osteoarthritis, rheumatoid arthritis, post-traumatic arthritis, avascular necrosis, subcapital fracture, or a revision of a failed femoral prosthesis. The Exeter Ceramic Femoral Heads are intended to be used with the femoral components of the previously released Exeter hip stems - and Exeter II Total Hip System (K891454).

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use<br>(Per 21 CFR 801.109) | <div style="text-align:center;">X</div> |
|------------------------------------------|-----------------------------------------|
|------------------------------------------|-----------------------------------------|

OR

| Over-The-Counter Use<br>(Optional Format 1-2-96) |  |
|--------------------------------------------------|--|
|--------------------------------------------------|--|

| (Division Sign-Off)                     |         |
|-----------------------------------------|---------|
| Division of General Restorative Devices |         |
| 510(k) Number                           | K993889 |

---

**Source:** [https://fda.innolitics.com/device/K993889](https://fda.innolitics.com/device/K993889)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
