FUJI COMPUTED RADIOGRAPHY DUAL SIDE READING, BOTH SIDE READING, FCR 5501D DUAL SIDE READER
K993861 · Fujifilm Medical System U.S.A., Inc. · MQB · Feb 10, 2000 · Radiology
Device Facts
Record ID
K993861
Device Name
FUJI COMPUTED RADIOGRAPHY DUAL SIDE READING, BOTH SIDE READING, FCR 5501D DUAL SIDE READER
Applicant
Fujifilm Medical System U.S.A., Inc.
Product Code
MQB · Radiology
Decision Date
Feb 10, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Indications for Use
The indications for use of the Fuji Computed Radiography Dual Side Reading is the same as the predicate device: the identification, capture, digitization, and processing of diagnostic x-ray images, and associating patient and exam identification with the images.
Device Story
Device is a computed radiography system (FCR 5501D) consisting of an image reader and photostimulable phosphor imaging plates (IPs). System is cassetteless with IPs integrated into an upright exam stand/reader. After x-ray exposure, the IP is scanned, erased, and held for reuse; allows continuous workflow without waiting for erasure. IP uses europium-activated barium fluorohalide compounds in a crystal form with an organic binder. Key innovation is a transparent support layer allowing emission detection from both sides of the IP, increasing detective quantum efficiency (DQE) by 30-40% compared to single-side reading. Used in clinical settings by radiology staff to produce digital diagnostic images. Output is processed digital x-ray data used by clinicians for diagnostic decision-making.
Clinical Evidence
Bench testing only. Comparison of modulation transfer function (MTF) and detective quantum efficiency (DQE) against predicate devices. Results show virtually identical MTF and a 30-40% increase in relative DQE for the dual-side reading method.
Technological Characteristics
Computed radiography system. Imaging plate (ST-55 BD) features europium-activated barium fluorohalide phosphor, 320 micrometer thickness, and transparent support layer. 10-bit grayscale (1024 levels). Conforms to UL Standard 2601-1 for medical electrical equipment safety.
Indications for Use
Indicated for the identification, capture, digitization, and processing of diagnostic x-ray images, and associating patient and exam identification with the images.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
FEB 1 0 2000
Image /page/0/Picture/1 description: The image shows the Fujifilm logo. The logo consists of a stylized graphic element on the left, followed by the word "FUJIFILM" in bold, sans-serif font. The graphic element is a geometric shape with the letters "fuji" embedded within it. The text is all in black.
I - Imaging & Information
FUJI MEDICAL SYSTEMS U.S.A., INC.
419 WEST AVENUE STAMFORD, CT 06902 Telephone: 203/324-2000 Fax: 203/353-0926
K993861
Page 1 of 3
510(k) Summary [as required by 21 CFR 807.92]
#### Date Prepared [21 CFR 807.92(a)(1)]
November 12, 1999
## Submitter's Information [21 CFR 807.92(a)(1)]
Fuji Medical Systems U.S.A., Inc. 419 West Avenue Stamford, Connecticut 06902 Telephone: (203) 602-3677 Facsimile: (203) 327-6485 Contact: Joseph M. Azary
# Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
The device trade names are:
- Fuji Computed Radiography Dual Side Reading .
- . Fuji Computed Radiography Both Side Reading
- . Fuji Computed Radiography FCR 5501D Dual Side Reader
The device common or classification names are: Computed radiography system, FCR system, digital imaging system.
#### Predicate Device [21 CFR 807.92(a)(3)]
Fuji identifies the predicate devices as follows:
| Subject Device | FCR 5501D Dual Side Reader<br>(Standard Pixel Density Image<br>Mode) | FCR 5501D Dual Side Reader<br>(High Pixel Density Image Mode) |
|-----------------------------|----------------------------------------------------------------------|---------------------------------------------------------------|
| Predicate Device | FCR 9501 | FCR 9501HQ |
| 510(k) for Predicate Device | K944046 | K951373 |
| Classification | Class II, 892.1900 | Class II, 892.1900 |
| Product Code | 90LLZ and 90IXW | 90LLZ and 90IXW |
{1}------------------------------------------------
# Description of the Device [21 CFR 807.92(a)(4)]
K993861
Page 2 of 3
The Fuji Computed Radiography Dual Side Reading consists of an image reader (FCR 5501D Dual Side Reader) and Imaging Plates (described below). The image reader is cassetteless because the IPs are built-into this upright exam stand/reader. After exposure, the IP is in turn to the scanning, erasing, and holding areas. For the operator, this means that as soon as the IP is exposed , another is brought into the exposure area, without having to wait for reading and erasure.
As with other FCR image readers, the FCR 5501D Dual Side Reader will feature a photostimulable phosphor imaging plate composed of europium activated barium fluorohalide compounds in a crystal form held in an organic binder. The Imaging Plate used with the subject device (ST-55 BD) will have a transparent support, which will allow emission detection from both the support side and phosphor side of the IP.
| IP Characteristic | ST-55 BD (Dual Side IP) | ST-V, ST-VA, ST-VN |
|-----------------------|--------------------------------------------|-------------------------------|
| Thickness | 320 micrometers; thicker phosphor<br>layer | 230 micrometers |
| Protective Layer | Protective Layer is present | Protective Layer is present |
| Support Layer | Transparent Support | Non-transparent support |
| Light Shielding Layer | Not present | Light Shielding Layer present |
### Intended Use [21 CFR 807.92(a)(5)]
The indications for use of the Fuji Computed Radiography Dual Side Reading is the same as the predicate device: the identification, capture, digitization, and processing of diagnostic x-ray images, and associating patient and exam identification with the images.
### Technological Characteristics [21 CFR 807.92(a)(6)]
The subject device represents a change to the predicate devices (FCR 9501HQ) that allows for reading on both sides of the imaging plate. The technology used in the subject device is a scientific article entitled "Novel Computed Radiography System with Improved Image Quality by Detection of Emissions from Both Sides of an Imaging Plate" by Satoshi Arakawa, Wataru Itoh, Katsuhiro Kohda, Toshiaki Suzuki.
The gray scale is 10 bits (1024 gray levels) for both the predicate and subject devices. The sampling raster of the subject device is identical to that of the predicate device. The one exception is that the subject device offers additional IP reading area sizes, such as 17x17 in., 17x14 in., and 18x43cm.
# Performance Data [21 CFR 807.92(b)(1)]
Based on our scientific data we found the modulation transfer function (MTF) for the subject device is virtually identical to the MTF data for the predicate devices. Additionally, the relative detective quantum efficiency (DQE) for the subject device provides a 30-40% increase in relative DQE, as compared to the predicate device.
> 510 (k) Fuji Dual Side Reading
{2}------------------------------------------------
K993861
Page 3 of 3
The subject device was tested by Underwriters Laboratories Inc. and found to conform to the UL Standard 2601-1 Medical Electrical Equipment Part 1: General Requirements for Safety (File E179671, Project 99SC47495).
#### Conclusion [21 CFR 807.92(b)(3)]
In conclusion, the Fuji Computed Radiography Dual Side Reading are nearly identical in design, materials, construction, and function to the predicate devices. The main difference is that the subject device (image reader and imaging plate) have been modified in a manner that allows emission from both sides of the imaging plate.
For a radiographic imaging system, image quality is the principal area of concern for safety and effectiveness. The results of laboratory tests demonstrate the subject devices provide equivalent MTF and increased DOE. In fact, based on test data the dual side reading method enhances image quality.
The subject device conforms to UL Standard 2601-1 Medical Electrical Equipment Part 1: General Requirements for Safety (File E179671, Project 99SC47495).
We conclude that the subject devices are as safe and effective as the predicate device.
510 (k) Fuji Dual Side Reading
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 1 0 2000
Joseph M. Azary Regulatory Affairs Coordinator Fuji Medical Systems U.S.A., Inc. 419 West Avenue Stamford, CT 06902
Re:
K993861 Fuji Computed Radiography Image Reader with Dual Side IP Reading Dated: November 12, 1999 Received: November 15, 1999 Regulatory class: II 21 CFR 892.1630/Procode: 90 MQB
Dear Mr. Azary:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Ad (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register, Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
5 10(k) Number (if known):
Device Name: Fuji FCR 5501D
Indications For Use:
The indications for use of the FCR 5501D is the same as the predicate device: the identification, capture, digitization, and processing of diagnostic x-ray images, and associating patient and exam identification with the images.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH Office of Device Evaluation (ODE)
| (Division Sign-Off) |
|---------------------|
|---------------------|
Division of Reproductive, Abdominal, ENTDivision of Reproductive, Abdominal, ENT, and Radiological Devic 510(k) Number
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
210 (k) Fuji Dual Side Reading
---
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.