BAXTER PERCUTANEOUS MECHANICAL THROMBECTOMY (PMT) DEVICE, MODELS PMT0606FD AND PMT1206FD

K993816 · Baxter Edwards · MCW · May 4, 2000 · Cardiovascular

Device Facts

Record IDK993816
Device NameBAXTER PERCUTANEOUS MECHANICAL THROMBECTOMY (PMT) DEVICE, MODELS PMT0606FD AND PMT1206FD
ApplicantBaxter Edwards
Product CodeMCW · Cardiovascular
Decision DateMay 4, 2000
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4875
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Baxter Percutaneous Mechanical Thrombectomy (PMT) Device is designed for percutaneous removal of thrombus using aspiration and an internal helical screw to fragment and remove thrombus. The Baxter PMT device is indicated for removal of thrombus in synthetic dialysis access grafts.

Device Story

Baxter PMT Device is a 6-French catheter for percutaneous thrombus removal in synthetic dialysis access grafts. Device features stainless steel housing with internal helical screw attached to flexible drive shaft; battery-operated motor rotates screw at ~3,100 rpm. Central lumen allows guidewire placement or fluid infusion. Device fragments and aspirates thrombus. Used in clinical settings by medical professionals. Output is physical removal of thrombus, restoring graft patency. Benefits include minimally invasive thrombus clearance.

Clinical Evidence

Device underwent design validation, functional testing, animal testing, and clinical testing. Clinical studies demonstrated the device is as safe and effective as the predicate device for thrombus removal in synthetic dialysis access grafts.

Technological Characteristics

6-French catheter; stainless steel housing; helical screw; flexible drive shaft; battery-operated motor (approx. 3,100 rpm); central lumen for guidewire/infusion. Mechanical aspiration and fragmentation principle.

Indications for Use

Indicated for removal of thrombus in synthetic dialysis access grafts.

Regulatory Classification

Identification

An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Baxter Healthcare Corporation 17221 Red Hill Avenue (Irvine) PO Box 11150 Santa Ana, CA USA 92711-1150 Santa Ana, CA USA 92711-1150 949.250.2500 : : : : . # MAY = 4 2000 ## raxter #### 510(k) Summary | Submitter: | Baxter Healthcare Corporation, CardioVascular Group | |-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Diane Peterson, Senior Regulatory Affairs Associāte | | Date Prepared: | November 9, 1999 | | Trade Name: | Baxter Percutaneous Mechanical Thrombectomy (PMT)<br>Device | | Classification Name: | Peripheral Atherectomy Catheter | | Predicate Devices: | MICROVENA Amplatz Thrombectomy Device (ATD) | | Device Description: | The Baxter Percutaneous Mechanical Thrombectomy<br>(PMT) Device consists of a 6-French catheter<br>terminating at a stainless steel housing in which a small<br>diameter helical screw is attached to a drive shaft. A<br>battery-operated motor rotates a flexible drive shaft<br>connected to the helical screw at approximately 3,100<br>rpm. A central lumen within the device is provided for<br>guidewire placement or infusion of fluids. | | Intended Use: | The Baxter Percutaneous Mechanical Thrombectomy<br>(PMT) Device is designed for percutaneous removal of<br>thrombus using aspiration and an internal helical screw<br>to fragment and remove thrombus. The Baxter PMT<br>device is indicated for removal of thrombus in synthetic<br>dialysis access grafts. | | Comparative<br>Analysis | It has been demonstrated (through clinical studies) that<br>the Baxter Percutaneous Mechanical Thrombectomy<br>(PMT) Device is as safe and effective as the predicate<br>device for the removal of thrombus within synthetic<br>dialysis access grafts. | | Functional/Safety<br>Testing: | The Baxter Percutaneous Mechanical Thrombectomy<br>(PMT) Device has successfully undergone design<br>validation, as well as functional, animal and clinical<br>testing. | | Conclusion: | The Baxter Percutaneous Mechanical Thrombectomy<br>(PMT) Device is substantially equivalent to the<br>predicate device. | {1}------------------------------------------------ Image /page/1/Picture/1 description: The image contains the words "Public Health Service" in bold, black font. The words are stacked on top of each other, with "Public Health" on the top line and "Service" on the bottom line. The text appears to be extracted from a document or sign. Image /page/1/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing its body and wings. The symbol is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. MAY = 4 2000 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Diane Peterson Senior Regulatory Affairs Associate Baxter Healthcare Corporation Cardio Vascular Group 17221 Red Hill Avenue (Irvine) P.O. Box 11150 Santa Ana, CA 92711-1150 Re: K993816 Baxter Percutaneous Mechanical Thrombectomy Device Requlatory Class: II (two) Product Code: 74 MCW Dated: February 22, 2000 Received: February 23, 2000 Dear Ms. Peterson: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act {2}------------------------------------------------ Page 2 - Ms. Diane Peterson for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described This lected will afrow you co boying . The FDA finding of substantial In your 510(x) premained to a legally marketed predicate device equivalence of your device of a your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling II you desire specific advanced additionally 809.10 for in vitro regulation (21 Crk fart over the addition of Compliance at (301) diagnostic devices), please contact the Office promotion and diagnostic devisor), Iv, for questions on the promotion and 594-4640. Addicionariff 10- quise contact the Office of compliance at (301) 594-4639. Also, please note the regulation entitled, (301) 394 1039. There to premarket notification" (21CR 807.97). "Misbranding by reference to premarket notification the 7ct mi "Misbranding by reference to promation on your responsibilities under the Act may ocher general information of Small Manufacturers Assistance at its be obtained from the Brillers. I (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, sincerely yours, Christopher James E. Dillard III Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ Page of 510(k) Number (if known): Kggz816 Device Name: Baxter Percutaneous Mechanical Thrombectomy (PMT) Device Indications for Use: Baxter Percutaneous Mechanical Thrombectomy (PMT) Device is indicated for removal of thrombus in synthetic dialysis access grafts. #### (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office Of Device Eva Of Device Evaluation (QDE) Chito M.V. Kinta (Division Sign-Off) Division of Cardiovascular, Respira and Neurological Devices 510(k) Number Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________ (Optional Format 1-2-96) ### CONFIDENTIAL
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...