RAPID DRUG SCREEN 5-PANEL DRUG SCREENING TEST FOR COCAINE, MARIJUANA, OPIATES, AMPHETAMINE AND METHAMPHETAMINE
K993796 · American Bio Medica Corp. · LDJ · Dec 21, 1999 · Clinical Toxicology
Device Facts
| Record ID | K993796 |
| Device Name | RAPID DRUG SCREEN 5-PANEL DRUG SCREENING TEST FOR COCAINE, MARIJUANA, OPIATES, AMPHETAMINE AND METHAMPHETAMINE |
| Applicant | American Bio Medica Corp. |
| Product Code | LDJ · Clinical Toxicology |
| Decision Date | Dec 21, 1999 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.3870 |
| Device Class | Class 2 |
Intended Use
The Rapid Drug Screen 5-Panel test with Methamphetamine is used for the qualitative detection of the following abused substances in human urine: d-Amphetamine, Benzoyl ecgonine, Cannabinoids, Opiates and Methamphetamines. This immunoassay is a simplified qualitative screening method that provides only a preliminary result for use in determining the need for additional or confirmatory testing, i.e., gas chromatography/ mass spectrometry (GC/MS). "Rapid Drug Screen" 5-panel with cocaine, marijuana, opiates, amphetamine and methamphetamine is a lateral flow immunoassay for the simulatenous detection in urine of five abused drugs at stated detectable limits. (Each assay occupies a seperate channel). It is intended for use in the qualitative detection of d-Amphetamine (1000 ne/ml). Benzoyl ecgonine (300 ng/ml), Cannabinoids (50 ng/ml), Methamphetamines (1000 ng/ml) and Opiates (300 ng/ml). "Rapid Drug Screen" is intended for professional use. It is not intended for over the counter sale to non-professionals. The assays are easy to perform, but should not be used without proper supervision. These immuno-assays are simplified qualatative screening methods that provides only a preliminary result for use in the need for aditional or confirmatory testing, i.e., gas-chromatography/mass spectrometry (GC/MS). "Rapid Drug Screen" provides only a preliminary analytical test result. A more specific alternate chemical method must be used to obtain a more confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical and professional judgement should be applied to any drug of abuse test result, particulary when preliminary positive results are used. "Rapid Drug Screen" is not intended as a point of care test.
Device Story
Lateral flow immunoassay for qualitative detection of five drugs of abuse in human urine; d-Amphetamine, Benzoyl ecgonine, Cannabinoids, Opiates, Methamphetamines. Device consists of membrane strip with immobilized drug conjugates and dried colloidal gold-antibody complexes. Principle: competitive immunoassay; drug in urine competes with immobilized conjugate for limited antibody binding sites. Negative sample: colloidal gold-antibody complex binds to immobilized conjugate, forming visible line. Positive sample: drug in urine saturates antibody binding sites, preventing line formation. Control line confirms test validity. Professional use only; requires clinical judgment for interpretation; preliminary results require GC/MS confirmation.
Clinical Evidence
Bench testing only. Reproducibility evaluated using control urines at concentrations above and below cut-offs. Each sample tested four times, twice daily, for five days. All concentrations verified by GC/MS. Results confirmed performance reproducibility.
Technological Characteristics
Lateral flow immunoassay; membrane strip with immobilized drug conjugates; colloidal gold-antibody complex; five separate channels for simultaneous detection; qualitative visual readout; no electronic components.
Indications for Use
Indicated for professional use in the qualitative detection of d-Amphetamine, Benzoyl ecgonine, Cannabinoids, Opiates, and Methamphetamines in human urine. Not for OTC sale or point-of-care use. Provides preliminary results requiring GC/MS confirmation.
Regulatory Classification
Identification
A cannabinoid test system is a device intended to measure any of the cannabinoids, hallucinogenic compounds endogenous to marihuana, in serum, plasma, saliva, and urine. Cannabinoid compounds include delta-9-tetrahydrocannabinol, cannabidiol, cannabinol, and cannabichromene. Measurements obtained by this device are used in the diagnosis and treatment of cannabinoid use or abuse and in monitoring levels of cannabinoids during clinical investigational use.
Special Controls
*Classification.* Class II (special controls). A cannabinoid test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Predicate Devices
- American Bio Medica "Rapid Drug Screen" 5-Panel test kit with Methamphetamine (K-984525)
Related Devices
- K984525 — RAPID DRUG SCREEN 5-PANEL WITH METHAMPHETAMINE TEST · American Bio Medica Corp. · Feb 26, 1999
- K993961 — RAPID DRUG SCREEN 3-PANEL TEST FOR COCAINE, MARIJUANA (THC) (CANNABINOIDS) AND OPIATES (3 PANEL TEST FOR CTO) · American Bio Medica Corp. · Jan 24, 2000
Submission Summary (Full Text)
{0}------------------------------------------------
American Bio Medica Corporation Rapid Drug Screen 5-Panel Test with Methamphetamine 510(k) Summary
K993796
DEC 2 1 1999
## 510(k) Summary
Submitter's Name/Address: American Bio Medica Corporation 300 Fairview Avenue Hudson, N. Y. 12534
Date of Preparation of this Summary:
Device Trade or Proprietary Name:
Device Common/Usual Name or Classification Name:
Classification Number/Class:
Contact Person: Henry J. Wells Vice President of Product Development Phone: 800-227-1243 Fax: 518-822-0391
November 1999
Rapid Drug Screen 5-Panel with Methamphetamine Rapid Drug Screen 5-Panel with Methamphetamine
[no classification number]/Class II
This 510(k) Summary is being submitted in accordance with the requirements of 21 CFR 807.92.
The assigned 510(k) is:
Predicate Devices: American Bio Medica "Rapid Drug Screen" 5-Panel test kit with Methamphetamine (K-984525)
#### Test Description:
All of the assays employed in the Rapid Drug Screen panels are based on the same principle of highly specific reaction between antigens and antibodies.
Each assay is a one-step, immunoassay in which a specially labeled drug (drug conjugate) competes with drug which may be present in the sample for the limited number of binding sites on the antibody. The test device consists of a membrane strip onto which a drug conjugate has been immobilized. A colloidal gold-antibody complex is dried at one end of the membrane. In the absence of any drug in the urine sample, the colloidal gold-antibody complex moves with the urine by capillary action to contact the immobilized drug conjugate. An antibody-antigen reaction occurs forming a visible line in the "test" area. The formation of a visible line in the test area occurs when the test is below the cut-off for the drug.
{1}------------------------------------------------
When drug is present in the urine sample, the drug or metabolite will compete with the immobilized drug conjugate in the test area for the limited antibody sites on the colloidal gold-labeled antibody complex.. If sufficient amount of drug is present, it will fill all of the available binding sites, thus preventing attachment of the labeled antibody to the drug conjugate. An absence of a color band (line) in the test area is indicative of a positive result.
A control band (line), comprised of a different antibody/antigen reaction, is present on the membrane strip. The control line is not influenced by the presence or absence of drug in the urine, and therefore, should be present in all reactions.
A negative urine will produce two colored bands, and a positive sample will produce only one band.
#### Intended Use:
The Rapid Drug Screen 5-Panel test with Methamphetamine is used for the qualitative detection of the following abused substances in human urine: d-Amphetamine, Benzoyl ecgonine, Cannabinoids, Opiates and Methamphetamines. This immunoassay is a simplified qualitative screening method that provides only a preliminary result for use in determining the need for additional or confirmatory testing, i.e., gas chromatography/ mass spectrometry (GC/MS).
#### Performance Characteristics:
The Rapid Drug Screen 5-Panel test will detect drugs of abuse in human urine at the following levels:
| d-Amphetamine | 1000 ng/ml |
|------------------|------------|
| Benzoyl ecgonine | 300 ng/ml |
| Cannabinoids | 50 ng/ml |
| Methamphetamines | 1000 ng/ml |
| Opiates | 300 ng/ml |
Reproducibility was evaluated using control urines containing concentrations above and below the stated cut-off. Negative controls were also used. All concentrations were verified by GC/MS. Each sample was tested four times, twice daily,
for five days. The results confirmed the reproducibility of the Rapid Drug Screen 5-Panel with Methamphetamine performance.
{2}------------------------------------------------
### Conclusion:
The "Rapid Drug Screen" 5-Panel with Methamphetamine is substantially equivalent to the previously cleared 5-Panel with Methamphetamine (K-984525) as demonstrated by results obtained in the studies. All of the five analytes in the kit have been cleared by the 510(k) process. There is no evidence of cross-reactivity when the five colloidal gold-antibody complexes are mounted in a common device side-by-side with physical separation of the individual channels.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of human figures.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
DEC 21 1999
Mr. Henry Wells Vice President, Product Development American Bio Medica Corporation 300 Fairview Avenue Hudson, New York 12534
Re: K993796
> Trade Name: Rapid Drug Screen 5-Panel Test with Methamphetamines Regulatory Class: II Product Code: LDJ, DKZ, LAF, DJG, DIO Dated: November 3, 1999 Received: November 9, 1999
Dear Mr. Wells:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
#### Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Rapid Drug Screen 5-Panel Test with Methamphetamines
Indications For Use:
"Rapid Drug Screen" 5-panel with cocaine, marijuana, opiates, amphetamine and methamphetamine is a lateral flow immunoassay for the simulatenous detection in urine of five abused drugs at stated detectable limits. (Each assay occupies a seperate channel). It is intended for use in the qualitative detection of d-Amphetamine (1000 ne/ml). Benzoyl ecgonine (300 ng/ml), Cannabinoids (50 ng/ml), Methamphetamines (1000 ng/ml) and Opiates (300 ng/ml).
"Rapid Drug Screen" is intended for professional use. It is not intended for over the counter sale to non-professionals. The assays are easy to perform, but should not be used without proper supervision. These immuno-assays are simplified qualatative screening methods that provides only a preliminary result for use in the need for aditional or confirmatory testing, i.e., gas-chromatography/mass spectrometry (GC/MS).
"Rapid Drug Screen" provides only a preliminary analytical test result. A more specific alternate chemical method must be used to obtain a more confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical and professional judgement should be applied to any drug of abuse test result, particulary when preliminary positive results are used.
"Rapid Drug Screen" is not intended as a point of care test.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Joan Cooper
(Division Sign-Off) Division of Clinical Laboratory Devices 11 510(k) Number.
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use