K993671 · Candela Corp. · GEX · Jan 24, 2000 · General, Plastic Surgery
Device Facts
Record ID
K993671
Device Name
CANDELA SCLERO LONG PULSE DYE LASER
Applicant
Candela Corp.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jan 24, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Candela Sclero Long Pulse Laser is indicated for the following uses: General Surgery: Photocoagulation of benign cutaneous vascular lesion and benign cutaneous lesions. DermatologyPlastic Surgery: For treatment of benign cutaneous vascular lesion, such as facial and leg telangiectasia, rosacea, port wine stains, hemangiomas, spider angioma, poikiloderma of Civatte and benign cutaneous lesion, such as warts, scars, striae. Gynecology: Photocoagulation of benign cutaneous lesion and benign vascular lesion in gynecology Podiatry: Treatment of benign cutaneous lesion, such as warts. The Candela Sclero Long Pulse Laser includes an integrated Dynamic Cooling Device, which is indicated for: 1. the reduction in pain. 2. cooling of the skin prior to laser treatment
Device Story
Pulsed, flash lamp-excited dye laser system; treats benign cutaneous vascular/non-vascular lesions. Input: electrical energy; Output: laser energy delivered via optical fiber to handpiece. Integrated Dynamic Cooling Device (DCD) sprays cryogen onto skin prior to laser pulse; minimizes thermal damage; reduces pain. Operated by clinicians via onboard control panel. Safety interlocks protect patients/operators. Clinical benefit: precise photocoagulation of lesions with reduced thermal injury and pain.
Clinical Evidence
Bench testing only. Device conforms to Laser Performance Standard (21 CFR 1040), UL 544 electrical safety standards, and European Union Medical Device Directives. No clinical trial data provided.
Technological Characteristics
Flash lamp-excited dye laser; high voltage power supply; optical fiber delivery system; integrated cryogen-based Dynamic Cooling Device. Electrical safety per UL 544; laser safety per 21 CFR 1040. Embedded processor control system.
Indications for Use
Indicated for patients requiring photocoagulation or treatment of benign cutaneous vascular lesions (e.g., telangiectasia, rosacea, port wine stains, hemangiomas, spider angioma, poikiloderma of Civatte) and benign cutaneous lesions (e.g., warts, scars, striae) across general surgery, dermatology, plastic surgery, gynecology, and podiatry.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Candela SPTL Long Pulse/Tunable Pulse Dye Laser (K954934)
{0}------------------------------------------------
# K993671
## 510(k) Summary
### General Information:
This 510(k) is to provide notification of substantial equivalence for the Candela Sclero Long Pulse Laser System, which is substantially equivalent to previously marketed devices.
| Classification: | Class II (21 CFR § 878.4810 Laser Surgical Instrument for use in General<br>and Plastic Surgery and in Dermatology) |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Dermatology Laser, Long Pulse Dye Laser System |
| Predicate Devices: | Candela SPTL Long Pulse/Tunable Pulse Dye Laser cleared under 510(k)<br>number K954934, Candela SPTL (K931762), Coherent Argon/Dye Laser<br>(K882160), Cynosure LPIR, Apogee 40 (K971737), Coherent VersaPulse<br>Laser (K972347), Instruments for Medicine and Diagnostics Spectrum<br>VeinLase (K981952), and Altus CoolGlide (K991234) |
### Description:
The Sclero LP laser is a pulsed, flash lamp excited dye, medical laser, controlled by an embedded processor, to be used for the treatment of telangiectasia, port wine stains, benign cutaneous lesions and other benign cutaneous vascular lesions. The laser system has a Dynamic Cooling Device built into it which provides a short burst of cryogen spray prior to firing the laser output energy is delivered via an optical fiber to a handpiece on to the skin. The cryogen which is housed within the laser enclosure is delivered via a hose to a nozzle located in the handpiece. The Dynamic Cooling Device functions to cool the skin prior to laser treatment, minimizing thermal damage to skin during laser treatment and reducing pain associated with laser treatment.
The Candela Sclero Long Pulse system is comprised of the following components:
- High voltage power supply (HVPS) 1.
- Flashlamp laser head 2.
- Circulatory System 3.
- 4. Dve fluid system
- న్. Optical delivery system
- Software control system 6.
- Dynamic Cooling Device 7.
The Candela Sclero Long Pulse Lasers are equipped with safety interlock systems to protect patients and operators. Users of the device make selections from an onboard control panel to regulate operation during treatment.
### Indication for Use:
The Candela Sclero Long Pulse Laser System is indicated for use in photocoagulation of benign cutaneous vascular lesion, benign cutaneous lesion, and benign cutaneous vascular lesion in gynecology.
### Testing:
As a laser product, the Sclero Long Pulse is required to conform and does conform to the Laser Performance Standard (21 CFR 1040). In addition the device conforms to the UL 544 electrical safety standard and the Essential Requirements of the European Union Medical Device Directives.
### Summary of Substantial Equivalence:
The Candela Sclero Long Pulse Laser has the same intended use, utilizes the same operating principles and match key design aspects, including similar spot size, the same wavelength and / or the same maximum delivered power as the predicate devices.
On the basis of similarities in methods of assembly, method of operation, intended uses that its Candela Sclero Long Pulse Laser System is substantially equivalent to the predicate devices.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with three curved lines representing its wings.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 2 4 2000
Ms. Joan Clifford Clinical Research Manager Candela Corporation 530 Boston Post Road Wayland, Massachusetts 01778
Re: K993671
> Trade Name: Candela Sclero Long Pulse Laser System Regulatory Class: II Product Code: GEX Dated: October 29, 1999 Received: November 1, 1999
Dear Ms. Clifford:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{2}------------------------------------------------
Page 2 – Ms. Joan Clifford
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours.
Sincerely yours,
James E. Dillard III Acting Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Page 1 of 1.
S10(k) NUMBER (IF KNOWN): K993671
Candela Sclero Long Pulse DEVICE NAME:
INDICATIONS FOR USE:
The Candela Sclero Long Pulse Laser is indicated for the following uses:
General Surgery: Photocoagulation of benign cutaneous vascular lesion and benign cutaneous lesions.
DermatologyPlastic Surgery: For treatment of benign cutaneous vascular lesion, such as facial and leg telangiectasia, rosacea, port wine stains, hemangiomas, spider angioma, poikiloderma of Civatte and benign cutaneous lesion, such as warts, scars, striae.
Gynecology: Photocoagulation of benign cutaneous lesion and benign vascular lesion in gynecology
Podiatry: Treatment of benign cutaneous lesion, such as warts.
The Candela Sclero Long Pulse Laser includes an integrated Dynamic Cooling Device, which is indicated for:
510(k) Number .
- 1. the reduction in pain.
r -
- 2. cooling of the skin prior to laser treatment
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter-Use
(Optional Format 1-2-96)
(Division Sign-Off)
Division of Genera) Restorative Devices
510(k) Number K993671
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.