K993656 · Datex-Ohmeda, Inc. · CCK · Mar 15, 2000 · Anesthesiology
Device Facts
Record ID
K993656
Device Name
DATEX-OHMEDA TONOMETRY MODULE M-TONO
Applicant
Datex-Ohmeda, Inc.
Product Code
CCK · Anesthesiology
Decision Date
Mar 15, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1400
Device Class
Class 2
Intended Use
The Datex-Ohmeda Tonometry Module, M-TONO is intended to be used with a Datex-Ohmeda Modular monitoring systems for gastrointestinal tonometry measurements. The Datex-Ohmeda Tonometry Module, M-TONO is indicated for monitoring gastrointestinal CO2 (PgCO2) and calculation of various gastrointestinal tonometry parameters when used with a Datex-Ohmeda monitoring system. It is indicated for use in hospital patients.
Device Story
M-TONO is a plug-in parameter module for Datex-Ohmeda modular monitoring systems (AS/3, CS/3). It monitors gastrointestinal PCO2 (PgCO2) via a tonometry catheter placed in the patient's stomach or intestine. Operation involves periodic (every 10 minutes) inflation/deflation of the catheter balloon with ambient air to equilibrate with gastrointestinal CO2. The module analyzes the gas and displays PgCO2 and derived parameters on the host monitor. Used in hospital settings by qualified medical personnel. Provides real-time physiological data to assist clinicians in assessing gastrointestinal perfusion/status.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), mechanical/environmental tolerance, and software validation/verification.
Technological Characteristics
Single-width plug-in module for modular monitoring systems. Measures PgCO2 via tonometry catheter balloon inflation/deflation cycle. Complies with IEC 60601-1, IEC 60601-1-2, and IEC 60601-1-4. Connectivity via integration into AS/3 and CS/3 monitoring systems.
Indications for Use
Indicated for monitoring gastrointestinal CO2 (PgCO2) and calculating tonometry parameters (PgCO2-arterial PCO2 difference, PgCO2-end-tidal PCO2 difference, intramucosal pH) in hospital patients. For use by qualified medical personnel only.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
K052167 — DATEX-OHMEDA S/5 TONOMETRY MODULE, E TONO AND ACCESSORIES · GE Healthcare · Sep 8, 2005
K993296 — TRIP TONOMETRY CATHETER, 14F · Datex-Ohmeda, Inc. · Dec 15, 1999
K980384 — TRIP TONOMETRY CATHETER, 18F · Datex-Engstrom, Inc. · May 1, 1998
K983366 — TRIP TONOMETRY CATHETER, 16F · Datex-Engstrom, Inc. · Dec 23, 1998
K192446 — CO2 Mainstream Module · Beijing Kingst Commercial & Trade Co., Ltd. · May 21, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
K995656
Page 1 of 4
# Premarket Notification 510(k) Summary As required by section 807.92 Datex-Ohmeda Tonometry Module, M-TONO
## GENERAL COMPANY INFORMATION as required by 807.92(a)(1)
## COMPANY NAME/ADDRESS/PHONE/FAX:
Datex-Ohmeda, Inc. 3 Highwood Drive Tewksbury, MA 01876 Tel: 978-640-0460 Fax: 978-640-0469 NAME OF CONTACT: Mr. Joel Kent FDA Official Correspondent DATE: October 28, 1999
## DEVICE NAME as required by 807.92(a)(2)
## TRADE NAME:
Datex-Ohmeda Tonometry Module, M-TONO
COMMON NAME: Regional Capnometer
### CLASSIFICATION NAME:
Analyzer, Gas, Carbon-Dioxide, Gaseous-phase (per 21CFR 868.1400)
### NAME OF LEGALLY MARKETED DEVICE FOR WHICH A CLAIM OF SUBSTANTIAL EQUIVALENCE IS MADE as required by 807.92(a)(3)
The Datex-Ohmeda Tonometry Module, M-TONO is substantially equivalent to the legally marketed (predicate) Tonocap™ Monitor (K962638).
Datex-Ohmeda, Inc. Three Highwood Drive Tewksbury, MA 01876 www.datex-ohmeda.com
{1}------------------------------------------------
## DEVICE DESCRIPTION as required by 807.92(a)(4)
The Datex-Ohmeda Tonometry Module, M-TONO (later referred to as M-TONO) is a module used to monitor gastrointestinal PCO2. The user interface has been implemented in the main software of the AS/3 and CS/3 monitoring systems.
The Datex-Ohmeda Tonometry Module, M-TONO is intended to be used with a Datex-Ohmeda Modular monitoring systems for gastrointestinal tonometry measurements.
The Datex-Ohmeda Tonometry Module, M-TONO is indicated for monitoring gastrointestinal CO2 (PgCO2) and calculation of various gastrointestinal tonometry parameters when used with a Datex-Ohmeda monitoring system. It is indicated for use in hospital patients.
This device is indicated for use by qualified medical personnel only.
M-TONO is a single-width plug-in parameter module including the tonometry measurement module for a modular monitoring system. This module is designed for use in the following Datex-Ohmeda modular monitors: AS/3 Anesthesia Monitor, AS/3 Compact Monitor, CS/3 Compact Monitor and CS/3 Critical Care Monitor.
The tonometry module measures the gastrointestinal PCO2 (PgCO2) every 10 minutes utilizing a tonometry catheter placed into the patient's stomach or intestine. Initially the module fills the balloon of the Tonometrics catheter with ambient air, then repeatedly every 10 minutes deflates the balloon, analyzes and displays the PgCO2, and inflates the balloon again.
#### INTENDED USE as required by 807.92(a)(5)
The Datex-Ohmeda Tonometry Module, M-TONO is intended to be used with a Datex-Ohmeda Modular monitoring systems for gastrointestinal tonometry measurements.
The Datex-Ohmeda Tonometry Module, M-TONO is indicated for monitoring gastrointestinal CO2 (PgCO2) and calculation of various gastrointestinal tonometry parameters when used with a Datex-Ohmeda monitoring system.
It is indicated for use in hospital patients.
{2}------------------------------------------------
## SUMMARY OF TECHNOLOGICAL CHARACTERITICS OF DEVICE COMPARED TO THE PREDICATE DEVICE as required by 807.92(a)(6)
The Datex-Ohmeda Tonometry Module, M-TONO is substantially equivalent to the legally marketed (predicate) Tonocap™ Monitor (K962638). Based on the analysis and other documentation included in this 510(k) notification and attachments it is evident that the main features and indications for use of Datex-Ohmeda Tonometry Module, M-TONO is substantially equivalent to the predicate Tonocap™ Monitor (K962638).
The Datex-Ohmeda Tonometry Module, M-TONO and TONOCAP™ monitor (K962638) have the same intended use for gastrointestinal PCO2 monitoring.
The Datex-Ohmeda Tonometry Module, M-TONO and TONOCAP™ monitor (K962638) are indicated for monitoring gastrointestinal PCO2 (PgCO2) and calculation of various gastrointenstinal tonometry parameters. Both are indicated for use in hospital patients. The Datex-Ohmeda Tonometry Module. M-TONO uses Tonometrics catheters, which have separate 510(k) clearances, similar to those used by the predicate TONOCAP™ monitor (K962638).
The main differences between M-TONO and the predicate is primarily due to fact that M-TONO is a module used in modular system while predicate TONOCAP™ monitor (K962638) is a standalone configured monitor. Additionally, airway gas measurement is not made with M-TONO since another module in the AS/3 and CS/3 family already handles this function.
The M-TONO module can also measure higher PgCO2 than the predicate TONOCAP™ monitor (K962638).
The M-TONO module uses a smaller PgCO2 measurement volume than predicate TONOCAPTM monitor (K962638).
It is evident that the Datex-Ohmeda Tonometry Module, M-TONO is substantially equivalent to the predicate TONOCAP™ Monitor (K962638).
The comparison above shows that there are no questions of safety and effectiveness for the Datex-Ohmeda Tonometry Module, M-TONO.
{3}------------------------------------------------
## SUMMARY OF NONCLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as required by 807.92(b)(1)(3)
The Datex-Ohmeda Tonometry Module, M-TONO has taken into account general electrical safety requirements in the design.
The following safety standards have been followed:
- · IEC 60601-1:1988 + Amendments: A1:1991,A2:1995
- · CAN/CSA C22.2 No. 601-1-M90 (1990) + S1 (1994)+Amdt2:1998
- IEC 60601-1-2:1993
- · IEC 60601-1-4:1996
The Datex-Ohmeda Tonometry Module, M-TONO complies with the requirements of above mentioned safety standards and is therefore safe and effective for the intended use. The module has been thoroughly tested including electrical safety, electromagnetic compatibility, mechanical and environmental tolerance, software validation and verification of specifications.
#### Conclusion:
The summary above shows that there are no questions of safety and effectiveness for the Datex-Ohmeda Tonometry Module, M-TONO as compared to the predicate device.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features a stylized image of an eagle or bird with outstretched wings, symbolizing protection and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird, indicating the department's name and affiliation with the United States of America.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 1 5 2000
Mr. Joel C. Kent FDA Official Correspondent Manager, Quality and Regulatory Affairs Datex-Ohmeda, Inc. Three Highwood Drive Tewksbury, Massachusetts 01876
Re: K993656 M-TONO Tonometry Module Dated: February 17, 2000 Received: February 18, 2000 Requiatory Class: Il CFR \$868.1400/Procode: 73 CCK
#### Dear Mr. Kent:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce pror to May 26, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Aproval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalion assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requilation may result in requlatory action. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predication for your device and thus. permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Daniel G. Schultz, M.D. Captain, USPHS Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{5}------------------------------------------------
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Kaasaese
## Device Name: Datex-Ohmeda Tonometry Module, M-TONO
## Indications For Use:
The Datex-Ohmeda Tonometry Module, M-TONO is indicated for monitoring gastrointestinal CO2 (PgCO2) and calculation of various gastrointestinal tonometry parameters (gastrointestinal - arterial PCO2 difference, gastrointestinal - end-tidal PCO2 difference and intramucosal pH) when used with a Datex-Ohmeda modular monitoring system. It is indicated for use in hospital patients.
This device is indicated for use by qualified medical personnel only.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use_V OR Over-The-Counter Use _ (Per 21 CFR 801.109) (Optional Format 1-2-96) (Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devic 510(k) Number
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.