BINDAZYME ANTI-TISSUE TRANSGLUTAMINASE IGA EIA

K993612 · The Binding Site, Ltd. · MVM · Dec 21, 1999 · Immunology

Device Facts

Record IDK993612
Device NameBINDAZYME ANTI-TISSUE TRANSGLUTAMINASE IGA EIA
ApplicantThe Binding Site, Ltd.
Product CodeMVM · Immunology
Decision DateDec 21, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5660
Device ClassClass 2

Indications for Use

The in-vitro measurement of specific autoantibodies against tissue transglutaminase (tTG) present in serum for the diagnosis of Coeliac disease.

Device Story

BINDAZYME Anti-Tissue Transglutaminase IgA EIA is an in-vitro diagnostic enzyme immunoassay. It measures specific IgA autoantibodies against tissue transglutaminase (tTG) in patient serum samples. Used in clinical laboratory settings by trained laboratory personnel. The assay provides quantitative or qualitative results to assist clinicians in the diagnosis of Coeliac disease. By detecting these specific autoantibodies, the device supports the clinical evaluation of patients presenting with symptoms or risk factors for gluten-sensitive enteropathy, facilitating timely diagnosis and management.

Clinical Evidence

No clinical data provided in the document.

Technological Characteristics

Enzyme immunoassay (EIA) for the detection of IgA autoantibodies. In-vitro diagnostic device. Class II, Product Code MVM.

Indications for Use

Indicated for the in-vitro measurement of serum autoantibodies against tissue transglutaminase (tTG) to aid in the diagnosis of Coeliac disease in patients suspected of having the condition.

Regulatory Classification

Identification

A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a circular seal for the Department of Health & Human Services USA. The seal features the department's name in a circular arrangement around the perimeter. In the center of the seal is an abstract symbol that resembles three stylized human profiles facing to the right, with flowing lines suggesting movement or connection. ## DEC 2 1 1999 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 The Binding Site, Limited C/O Mr. Jay H. Geller East Tower, Suite 600 2425 West Olympic Boulevard Santa Monica, California 90404 Re: K993612 Trade Name: BINDAZYME Anti-Tissue Transglutaminase IgA EIA Regulatory Class: II Product Code: MVM Dated: October 21, 1999 Received: October 25, 1999 ## Dear Mr. Geller: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {1}------------------------------------------------ Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Steven Sutman Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## INDICATIONS FOR USE STATEMENT BINDAZYME Onti-Tissue Transglutaminase IgA Enzyme Immunoassay Device Name: Kit The in-vitro measurement of specific for Use: Indications autoantibodies against tissue transglutaminase (tTG) present in aucoantibodies againbo extremismosis of Coeliac disease. Prescription use \$\checkmark\$ Tau E. Makeri
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...