WICHITA FUSION NAIL FULLY THREADED SELF-TAPPING SCREW

K993603 · Howmedica Osteonics Corp. · HSB · Nov 19, 1999 · Orthopedic

Device Facts

Record IDK993603
Device NameWICHITA FUSION NAIL FULLY THREADED SELF-TAPPING SCREW
ApplicantHowmedica Osteonics Corp.
Product CodeHSB · Orthopedic
Decision DateNov 19, 1999
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3020
Device ClassClass 2
AttributesTherapeutic

Indications for Use

This device is intended for use with the Wichita™ Fusion Nail in cases for intramedullary knee arthrodesis. Arthrodesis is performed to relieve pain and to stabilize a knee joint that has been severely damaged as a result of trauma, infection, or failed previous surgeries (including previous total joint replacement). More specific indications include treatment of the sequelae of septic arthritis in patients who are not candidates for total knee arthroplasty, irretrievably failed total knee arthroplasty (either septic or aseptic), Charot arthropathy, painful degenerative knee in patients who are not candidates for total knee arthroplasty, arthrodesis of the knee for salvage in tumor surgery, delayed or non-union after previous arthrodesis of the knee, and any other condition where arthrodesis is the treatment of choice.

Device Story

Wichita™ Fusion Nail Fully Threaded Self-Tapping Screw is an orthopedic fixation component used in conjunction with the Wichita™ Fusion Nail for intramedullary knee arthrodesis. The device is a modification of the predicate Wichita™ partially threaded screw; it features a fully threaded design with fine threads, a reduced head length, and a rounded head profile. It is intended for surgical use by orthopedic surgeons to stabilize damaged knee joints. The screw provides mechanical fixation to facilitate bone fusion, relieving pain and restoring stability in patients where total knee arthroplasty is not viable.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Material: Wrought Vitallium® (CoCr) alloy conforming to ASTM F1537. Design: Fully threaded, self-tapping, fine thread profile, rounded head. Dimensions: Includes 25mm length option. Energy source: None (mechanical).

Indications for Use

Indicated for patients requiring intramedullary knee arthrodesis due to trauma, infection, failed total knee arthroplasty (septic/aseptic), Charcot arthropathy, painful degenerative knee, tumor salvage, or non-union, specifically for those who are not candidates for total knee arthroplasty.

Regulatory Classification

Identification

An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ NOV 1 9 1999 Special 510(k) Summary - Device Modification Summary of Safety and Effectiveness for the Wichita™ Fusion Nail Fully Threaded Self-Tapping Screw Wichita™ Fusion Nail Fully Threaded Self-Proprietary Name: Tapping Screw Intramedullary Fixation Rod Common Name: Intramedullary Fixation Rod, Classification Name and Reference: 21 CFR §888.3020 Proposed Requiatory Class: Class II Device Product Code: OR (87) HSB Jennifer A. Daudelin, Regulatory Affairs For Information contact: Howmedica Osteonics Corp. 359 Veterans Boulevard Rutherford, NJ 07070-2584 (201) 507-7283 Fax: (201) 507-6870 This Special 510(k) submission is intended to address a design modification to the predicate Wichita™ screw. The predicate device is a partially threaded screw manufactured from wrought Vitallium® (CoCr) Alloy which conforms to ASTM standard F1537. This device was found substantially equivalent via the 510(k) process. The subiect device is being modified from a partially threaded screw with course thread design to a fully threaded screw with fine thread design. The screw head length is being reduced by 1mm, and the head profile is changing from the square to the rounded design. Additionally, a 25mm screw will be added to the current product line. The intended use of the Wichita™ Fully Threaded Self-Tapping Screw is identical to the Wichita™ Partially Threaded Self-Tapping Screw. This device is intended for use with the Wichita™ Fusion Nail in cases for intramedullary knee arthrodesis. Arthrodesis is performed to relieve pain and to stabilize a knee joint that has been severely damaged as a result of trauma, infection, or {1}------------------------------------------------ ## K 9 93603 failed previous surgeries (including previous total joint replacement). More specific indications include treatment of the sequelae of septic arthritis in patients who are not candidates for total knee arthroplasty, irretrievably failed total knee arthroplasty (either septic or aseptic), Charot arthropathy, painful degenerative knee in patients who are not candidates for total knee arthroplasty, arthrodesis of the knee for salvage in tumor surgery, delayed or non-union after previous arthrodesis of the knee, and any other condition where arthrodesis is the treatment of choice. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird with three lines representing its wings or feathers. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 NOV 1 9 1999 Ms. Elizabeth A. Staub Vice President Regulatory Affairs, Quality Assurance and Clinical Research Stryker Howmedica Osteonics 359 Veterans Boulevard Rutherford, New Jersey 07070-2584 K993603 Re: Wichita™ Fusion Nail Fully Threaded Trade Name: Self-Tapping Screw Requlatory Class: II Product Code: нав Dated: October 4, 1999 Received: October 25, 1999 Dear Ms. Staub: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ Page 2 - Ms. Elizabeth A. Staub This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA deberized in formalence of your device to a legally rinding of babban device results in a classification for your marketed produced or mits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in regaradiagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to Other general premarket notification" (21 CFR 807.97). information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, James E. Dillard III Acting Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use K993603 510(k) Number (if known): Device Name: Wichita™ Fusion Nail Fully Threaded Self-Tapping Screw Indications for Use: This device is intended for use with the Wichita™ Fusion Nail in cases for intramedullary knee arthrodesis is performed to relieve pain and to stabilize a knee joint that has been severely damaged as a result of trauma, infection, or failed previous surgeries (including previous total joint replacement). More specific indications include treatment of the sequelae of septic arthritis in patients who are not candidates for total knee arthroplasty, irretrievably failed total knee arthroplasty (either septic or aseptic), Charot arthropathy, painful degenerative knee in patients who are not candidates for total knee arthroplasty, arthrodesis of the knee for salvage in tumor surgery, delayed or non-union after previous arthrodesis of the knee, and any other condition where arthrodesis is the treatment of choice. ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use > (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96) bcotteleto
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.