MODIFICATION TO LASER-SHIELD II

K993582 · Xomed, Inc. · BTR · Jan 20, 2000 · Anesthesiology

Device Facts

Record IDK993582
Device NameMODIFICATION TO LASER-SHIELD II
ApplicantXomed, Inc.
Product CodeBTR · Anesthesiology
Decision DateJan 20, 2000
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5730
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Laser-Shield II is intended for endotracheal intubation. It is indicated for use for all types of surgical procedures involving carbon dioxide (10.60 microns) or KTP (532 mm) laser use (normal pulsed or continuous beam delivery in the non-contact mode), when endotracheal intubation is required to administer anesthesia gases or to overcome emergency obstruction of an airway.

Device Story

Laser-Shield II is a laser-resistant endotracheal tube used in surgical settings. The device features a silicone elastomer shaft wrapped in aluminum and fluoroplastic to provide laser resistance. It includes an inflatable cuff for tracheal sealing; the inflation assembly contains dry methylene blue, which mixes with sterile saline to provide visual detection of cuff rupture if struck by a laser. Sterile cottonoid patties are provided to be wetted and placed around the cuff to act as a heat sink. The device is operated by clinicians during procedures involving CO2 or KTP lasers to maintain airway patency and deliver anesthesia gases, protecting the patient from potential airway fires or injury during laser surgery.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Endotracheal tube with silicone elastomer shaft; laser-resistant overwraps of aluminum and fluoroplastic; inflatable cuff; inflation assembly with methylene blue indicator; sterile cottonoid patties for heat sink. Sterile, single-use device.

Indications for Use

Indicated for patients requiring endotracheal intubation for anesthesia or airway obstruction management during surgical procedures involving CO2 or KTP lasers.

Regulatory Classification

Identification

A tracheal tube is a device inserted into a patient's trachea via the nose or mouth and used to maintain an open airway.

Submission Summary (Full Text)

{0}------------------------------------------------ K993582 ### 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS K - 1.0 Date Prepared October 14, 1999 - 2.0 Submitter (Contact) Rov Berens Xomed Surgical Products Jacksonville, FL Telephone: 904-296-645 #### 3.0 Device Name Proprietary Name: Laser-Shield® II Tracheal Tube, Endotracheal Tube Common Name(s): Classification Name: Tracheal Tube #### 4.0 Device Classification Product Code 73BTR Class II Tier II Tracheal Tube 21 CFR 868.5730 #### 5.0 Device Description Xomed Laser-Shield II is an endotracheal tube provided sterile with laser resistant overwraps of aluminum and fluoroplastic covering the silicone elastomer shaft. The tube is fitted with an inflatable cuff designed to provide tracheal seal. The inflation assembly/balloon contains dry methylene blue that mixes with a liquid inflation media (sterile normal saline) to provide visible detection of inadvertent cuff rupture by the laser. Sterile cottonoid patties are also provided for wetting and placement around the positioned and inflated cuff for additional heat sink. #### 6.0 Intended Use Xomed Laser-Shield II is intended for endotracheal intubation. It is indicated for use for all types of surgical procedures involving carbon dioxide (10.60 microns) or KTP (532 mm) laser use (normal pulsed or continuous beam delivery in the non-contact mode), when endotracheal intubation is required to administer anesthesia gases or to overcome emergency obstruction of an airway. #### 7.0 Substantial Equivalence The modified Xomed Laser-Shield II is identical to the current device with the exception of the extension of the aluminum wrap under a portion of the cuff and covering of this extension with a flexible silicone sleeve. The intended use and claims for the modified device remains unchanged and no new materials are being used. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a symbol that resembles three stylized human profiles facing to the right, with flowing lines suggesting movement or connection. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JAN 2 0 2000 Mr. Roy Berens Xomed, Inc. 6743 Southpoint Drive North Jacksonville, FL 32216-0980 Re: K993582 Modification to Laser-Shield® II Endotracheal Tube Requlatory Class: II (two) Product Code: 73 BTR October 20, 1999 Dated: October 22, 1999 Received: Dear Mr. Berens: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to leqally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such Failure to comply with the GMP regulation may result in assumptions. requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {2}------------------------------------------------ ### Page 2 - Mr. Roy Berens This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Joanna VA Westerhausen for, Celia M. Witten, Ph.D., M.D. Acting Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Premarket Notification [510(k)] for Modification to Laser-Shield® II # Intended Use Statement 993582 510(k) Number (if known): K Device Name: Modification to Laser-Shield II Indications for Use: Laser-Shield II is intended for endotracheal intubation. It is indicated for use for all types of surgical procedures involving carbon dioxide (10,60 microns) or KTP (532 nm) laser use (normal pulsed or continuous beam delivery in the non-contact mode), when endotracheal intubation is required to administer anesthesia gases or to overcome emergency obstruction of an airway. > (Please do not write below this line - continue on another page if needed) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) Or Over-the-Counter Use (Optional Format 1-2-96) io.Aulestschen 51 : Kki Numt
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...