← Product Code [EBF](/productcode/EBF) · K993566

# PROMPTFLOW (K993566)

_Espe Dental AG · EBF · Dec 6, 1999 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K993566

## Device Facts

- **Applicant:** Espe Dental AG
- **Product Code:** [EBF](/productcode/EBF.md)
- **Decision Date:** Dec 6, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3690
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

Flowable composite restorative material for: Minimally-invasive fillings Restorations of type III and V cavities (according to Black) Liner for restorations of type I and II cavities Small restorations of type I and II cavities Deciduous tooth fillings Fissure sealing Extended fissure sealing Tunnel preparation Repair of composite fillings and ceramic veneers Filling of defects and undercuts in crowns, inlays, and onlays Cementing porcelain veneers, crowns, inlays, and onlays Refacing acrylic temporaries Blockouts

## Device Story

PROMPTFLOW® is a flowable composite restorative material composed of methacrylates; intended for dental restoration of carious lesions or structural defects. Device functions by filling cavities or defects in teeth; applied by dental professionals in a clinical setting. Flowability is achieved through reduced filler content compared to the predicate REFERENCE®. Output is a hardened restorative material; used to restore tooth structure or cement dental prosthetics. Benefits include minimally invasive treatment options and versatility in restorative and repair applications.

## Clinical Evidence

No clinical data. Substantial equivalence is based on bench testing comparing physical and mechanical properties of the modified material (PROMPTFLOW®) against the predicate (REFERENCE®) and a reference flowable material (DYRACT® FLOW).

## Technological Characteristics

Composite restorative material composed of methacrylates. Flowable consistency achieved via reduced filler content. Material properties and manufacturing processes are consistent with the predicate REFERENCE®.

## Regulatory Identification

Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.

## Predicate Devices

- REFERENCE® by ESPE ([K992966](/device/K992966.md))
- DYRACT® FLOW by Dentsply ([K982395](/device/K982395.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# DEG - 6 1999

Image /page/0/Picture/1 description: The image shows a sequence of digits and a letter, all in bold black font. The sequence starts with the letter 'K', followed by the numbers '9', '3', '5', and '6'. The numbers are written in a simple, rounded style, and the letter 'K' is also in a similar style. The image is simple and clear, with the focus on the sequence of characters.

### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS IV.

## Submitter

| Name:                              | ESPE Dental AG                                            |
|------------------------------------|-----------------------------------------------------------|
| Street:                            | ESPE Platz                                                |
| ZIP-Code, City:                    | D-82229 Seefeld                                           |
| Federal State:                     | Bavaria                                                   |
| Country:                           | Germany                                                   |
| Establishment Registration Number: | 9611385                                                   |
| Contact:                           | Dr. Andreas Petermann, Manager U.S.<br>Regulatory Affairs |
| Phone:                             | 011-49-8152-700 1395                                      |
| Fax:                               | 011-49-8152-700 1869                                      |
| E-mail                             | Andreas_Petermann@ESPE.de                                 |
| Date:                              | October 19, 1999                                          |
| <b>Name of Device</b>              |                                                           |
| Proprietary Name:                  | PROMPTFLOW®                                               |
| Classification Name:               | Tooth shade resin material                                |
| Common Name:                       | Composite restorative material                            |
| <b>Predicate Device</b>            |                                                           |
| REFERENCE® by ESPE                 | K 992966                                                  |
| DYRACT® FLOW by Dentsply           | K 982395                                                  |

Description for the Premarket Notification

PROMPTFLOW® is classified as tooth shade resin material (21 C.F.R. § 872.3660) because it is a device composed of methacrylates intended to restore carious lesions or structural defects in teeth.

Image /page/0/Picture/7 description: The image shows the logo for ESPE Special 510(k) PROMPTFLOW®. The text "Special 510(k) PROMPTFLOW®" is at the top of the image. Below the text is the ESPE logo, which features a stylized tree inside of a square to the left of the word ESPE.

{1}------------------------------------------------

The recently 510(k) cleared composite restorative material REFERENCE® will be marketed under the tradename PROMPTFIL®. However, in this Special 510(k) submission the name REFERENCE® is used to avoid misunderstanding and confusion.

ESPE is submitting this Special 510(k) for modifications of its composite based restorative material REFERENCE®. The modification results in PROMPTFLOW®, a flowable composite restorative material. PROMPTFLOW® has the same fundamental scientific technology and the same intended use as REFERENCE®, therefore, we believe these modifications are eligible for the Special 510(k) process.

To achieve the flowability of PROMPTFLOW® the filler content had to be reduced. As a result, some mechanical strength characteristics decreased which leads to a reduced range of indications. However, comparison with another flowable restorative material, DYRACT® FLOW by Dentsply, shows, that PROMPTFLOW® is a suitable material for the claimed indications for use.

In this Special 510(k) Device Modification submission the chemical composition, the physical and mechanical properties, and the indications for use of both the unmodified REFERENCE® and the modified PROMPTFLOW® are compared. Furthermore, ESPE's design control activities are shortly described. DYRACT® FLOW is provided to have a predicate flowable restorative material because, as described above, the physical and mechanical properties of PROMPTFLOW® are not as strong as those of REFERENCE® due to the lower filler content.

However, the modified restorative material PROMPTFLOW® has the following similarities to the unmodified REFERENCE®:

- PROMPTFLOW® has in general the same intended use .
- PROMPTFLOW® is used by the same operating principle .
- . PROMPTFLOW® incorporates the same basic chemical design
- PROMPTFLOW® has the same shelf life .
- PROMPTFLOW® is manufactured using the same materials and processes .

In summary PROMPTFLOW® described in this submission is, in our opinion, substantially equivalent to the predicate devices.

Image /page/1/Picture/11 description: The image shows the words "Special 510(k) PROMPTFLOW®" in black font. Below this text is the word "ESPE" in a larger black font. To the left of the word ESPE is a logo of a tree.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol resembling three overlapping human figures or abstract shapes.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC - 6 1990

Mr. Andreas Petermann Manager U.S. Regulatory Affairs ESPE Dental AG ESPE Platz D2229 Seefeld,Bavaria

> Re: K993566 Promptflow® Trade Name: Regulatory Class: II Product Code: EBF Dated: October 28, 1999 Received: November 1, 1999

Dear Mr. Petermann:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any

{3}------------------------------------------------

Page 2 - Mr. Petermann

obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fday.gov/cdrh/dsmamain.html".

Sincerely yours,

Timothy A. Ulatowski

Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## STATEMENT OF INDICATIONS FOR USE

(As Required by 21 C.F.R. § 801.109)

510(k) Number:

K993566

Device Name:

Promptflow®

Indications for use:

Flowable composite restorative material for:

Minimally-invasive fillings

Restorations of type III and V cavities (according to Black)

Liner for restorations of type I and II cavities

Small restorations of type I and II cavities

Deciduous tooth fillings

Fissure sealing

Extended fissure sealing

Tunnel preparation

Repair of composite fillings and ceramic veneers

Filling of defects and undercuts in crowns, inlays, and onlays

Cementing porcelain veneers, crowns, inlays, and onlays

Refacing acrylic temporaries

Blockouts

Prescription use: 12

Over-the counter use □

(Division Sign-Off Division of Dental, Infection C and General Hospital 510(k) Numbe

---

**Source:** [https://fda.innolitics.com/device/K993566](https://fda.innolitics.com/device/K993566)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
