TALON ANCHOR SNAP-PAK

K993327 · Mitek Products · DZL · Dec 16, 1999 · Dental

Device Facts

Record IDK993327
Device NameTALON ANCHOR SNAP-PAK
ApplicantMitek Products
Product CodeDZL · Dental
Decision DateDec 16, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.4880
Device ClassClass 2
AttributesTherapeutic

Intended Use

The TALON™ Anchor Snap-Pak is used for the fixation of non-absorbable braided surgical suture to bone. This product is intended for the following indications: Repair and reconstruction or reattachment of tendons and ligaments, muscles and soft tissue flaps to the parietal, temporal ridge, frontal, maxilla, zygoma and periorbital bones of the skull.

Device Story

TALON™ Anchor Snap-Pak is a sterile bone fixation implant; used to anchor or lock non-absorbable braided surgical suture within pre-drilled bone sites; secures soft tissue to bone. Used in cranio/maxillofacial surgical procedures; operated by surgeons. Device functions as a mechanical anchor; provides fixation for tendons, ligaments, muscles, and soft tissue flaps to skull bones. Benefits include stable reattachment of soft tissue to bone during repair and reconstruction.

Clinical Evidence

Bench testing only. Performance was evaluated via pull-out force testing using preserved human cadaver skulls to compare the strength of the TALON™ Anchor Snap-Pak against the predicate Mitek Micro Anchor.

Technological Characteristics

Sterile bone fixation implant; mechanical anchor design for suture fixation; intended for use in pre-drilled bone sites. Materials and sterilization methods are consistent with predicate Mitek anchors.

Indications for Use

Indicated for fixation of non-absorbable braided surgical suture to bone for repair, reconstruction, or reattachment of tendons, ligaments, muscles, and soft tissue flaps to the parietal, temporal ridge, frontal, mandible, maxilla, zygoma, and periorbital bones of the skull.

Regulatory Classification

Identification

An intraosseous fixation screw or wire is a metal device intended to be inserted into fractured jaw bone segments to prevent their movement.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ DEC 1 6 1999 ## 510(k) Summary | Trade Name: | TALON™ Anchor Snap-Pak | |----------------------|-----------------------------------------------------------------------------------------------------------------| | Sponsor: | Mitek Products | | | 60 Glacier Drive | | | Westwood, MA 02090 | | | Registration #1221934 | | Contact: | Paula E. Bulger | | | Regulatory Affairs Project Manager | | | Mitek Products | | | 60 Glacier Drive | | | Westwood, MA 02090 | | | Fax: (781) 461-9166 | | | Phone: (781) 251-2700 | | Device Generic Name: | Staple, Fixation, Bone | | Classification: | According to Section 513 of the Federal Food, Drug, and<br>Cosmetic Act, the device classification is Class II. | | Product Code: | JDR (21 CFR 888.3030) | | Predicate Devices: | K931782 - Mitek LS Anchor<br>K962511, K982420 - Mitek Micro Anchor | Product Description: The device described in this 510(k) is a sterile implant used to anchor or lock suture within pre-drilled bone sites and firmly secure soft tissue to bone. ## Indications for Use: The TALON™ Anchor Snap-Pak is used for the fixation of non-absorbable braided surgical suture to bone. This product is intended for the following indications: Repair and reconstruction or reattachment of tendons and ligaments, muscles and soft tissue flaps to the parietal, temporal ridge, frontal, maxilla, zygoma and periorbital bones of the skull. ## Safety and Performance: The following safety and performance data has been provided to support substantial equivalence of TALON™ Anchor Snap-Pak: | Performance testing: Pull-out force (preserved human cadaver skull) | |---------------------------------------------------------------------| | Strength comparison (TALON vs. Mitek Micro Anchor) | ## Conclusion: Based on safety and performance data, similarities in design, operating principle, materials, biocompatibility and sterilization method, the TALON™ Anchor Snap-Pak has been shown to be substantially equivalent to predicate devices under the Federal Food, Drug and Cosmetic Act. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure with outstretched arms. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC I 6 1999 Ms. Paula E. Bulger Regulatory Affairs Project Manager Mitek Products 60 Glacier Drive 02090 Westwood, Massachusetts K993327 Re : > Talon™ Anchor Snap-Pak Trade Name: Requlatory Class: II Product Code: DZL October 1, 1999 Dated: Received: October 4, 1999 Dear Ms. Bulger: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ಗಳ substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any {2}------------------------------------------------ Page 2 - Ms. Bulger obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA findinq of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in yitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Patricio Cucente /Koz Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ Page 1_ of __ of _ 510(k) Number (if known): K993327 Device Name: __ TALON™ Anchor Snap-Pak Indications for Use: The TALON™ Anchor Snap-Pak is used for the fixation of non-absorbable braided surgical suture to bone. This product is intended for the following indications: Cranio/Maxillofacial: Repair and reconstruction or reattachment of tendons and ligaments, muscles and soft tissue flaps to the parietal, temporal ridge, frontal, mandible, maxilla, zygoma and periorbital bones of the skull. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-the -Counter Use ________________________________________________________________________________________________________________________________________________________ Susan Runner (Division Sign-Off) Division of Dental, Infection Control, ಇಂಡ General Hospital Devi . in | Number __
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...