SYSMEX AUTOMATED COAGULATION ANALYZER, MODEL CA-1500

K993299 · Dade Behring, Inc. · GKP · Dec 7, 1999 · Hematology

Device Facts

Record IDK993299
Device NameSYSMEX AUTOMATED COAGULATION ANALYZER, MODEL CA-1500
ApplicantDade Behring, Inc.
Product CodeGKP · Hematology
Decision DateDec 7, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.5400
Device ClassClass 2

Indications for Use

The intended use of the Sysmex® CA-1500 is as a fully automated, computerized blood plasma coagulation analyzer for in vitro diagnostic use in clinical laboratories. The instrument uses citrated human plasma to perform the following parameters and calculated parameters: Clotting Analysis Parameters - Prothrombin Time (PT) . - Activated Partial Thromboplastin Time (APTT) . - Fibrinogen (Clauss) . Chromogenic Analysis Parameters - Antithrombin III . Calculated Parameters - . PT Ratio - PT INR - PT % ● - Extrinsic Factors (II, V, VII, X) ● Intrinsic Factors (VIII, IX, XI, XII) . - Derived Fibrinogen ● - Factor Assays % Activity .

Device Story

Sysmex CA-1500 is a fully automated, computerized blood plasma coagulation analyzer; used in clinical laboratories by trained personnel. Device accepts citrated human plasma samples; performs clotting analysis (PT, APTT, Fibrinogen) and chromogenic analysis (Antithrombin III). System calculates parameters including PT ratio, INR, PT %, and various extrinsic/intrinsic factor activities. Output provided to healthcare providers for clinical decision-making regarding patient coagulation status and therapy monitoring. Automated processing replaces manual coagulation testing; improves throughput and consistency in clinical settings.

Clinical Evidence

Bench testing and clinical correlation studies performed. Comparison studies against predicate devices (CA-6000, BCT) using 66-165 patient samples per assay showed strong correlation (r = 0.945 to 0.999). Precision studies (n=40 per test) demonstrated low variability, with total %CV ranging from 0.6% to 9.2% across various control levels. No clinical prospective trials on patient outcomes reported; evidence based on analytical performance and method comparison.

Technological Characteristics

Automated coagulation analyzer; utilizes citrated human plasma. Performs clotting and chromogenic assays. System is computerized/automated. Connectivity and specific materials not detailed. Software-based analysis for coagulation parameters.

Indications for Use

Indicated for in vitro diagnostic use in clinical laboratories to perform coagulation and chromogenic analysis on citrated human plasma samples. Used for assessing clotting parameters (PT, APTT, Fibrinogen) and chromogenic parameters (Antithrombin III) and calculating related factors in patients requiring coagulation monitoring.

Regulatory Classification

Identification

A coagulation instrument is an automated or semiautomated device used to determine the onset of clot formation for in vitro coagulation studies.

Special Controls

*Classification.* Class II (special controls). A fibrometer or coagulation timer intended for use with a coagulation instrument is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ DEC - 7 1999 K993299 # 510(k) Summary of Safety and Effectiveness Information Sysmex ® Automated Coagulation Analyzer CA-1500 September 30, 1999 Dade Behring Inc. 1851 Delaware Parkway Miami, FL 33125 Contact Person: Radames Riesgo at 305.636.7727 or by facsimile at 305.637.6887. Trade or Proprietary Name: Sysmex® Automated Coagulation Analyzer CA-1500 | Common or Usual Name: | Automated Coagulation Instruments | | |-----------------------|--------------------------------------------------------------------------------------------|---------| | Classification Name: | Coagulation instrument (21 CFR §864.5400) | | | Registration Number: | | | | | Manufacturing Site<br>Sysmex Corporation<br>Kobe, Japan | 9613959 | | | Importer<br>Sysmex Corporation of America<br>One Wildlife Way<br>Long Grove, IL 60047-9596 | 1422681 | | | Distributor<br>Dade Behring Inc.<br>Glasgow Site<br>P.O. Box 6101<br>Newark, DE 19714-6101 | 2517506 | The CA-1500 is substantially equivalent in intended use and technological characteristics to the Sysmex® Automated Coagulation Analyzer CA-6000, Sysmex Corporation, Kobe, Japan, which was cleared by FDA under Document Control Nos. K964139 and K992321; or the Behring Coagulation Timer (BCT), Dade Behring, Marburg, Germany which was cleared by FDA under Document Control No. K955278. As demonstrated by clinical correlation studies, the performance claims of the proposed device are similar to the predicate devices. During those studies, specimens were evaluated from apparently healthy individuals and from patients with different pathological conditions which are expected to affect the results for a particular assay. The following summary shows the results of the comparison studies between the proposed and the predicate devices. {2}------------------------------------------------ 510(k) Summary of Safety and Effectiveness Information Sysmex® Automated Coagulation Analyzer CA-1500 Page 2 | Test | Predicate Device | Sample Number (n) | Coefficient of Correlation (r) | Regression Equation | |----------------------------------------------|------------------|-------------------|--------------------------------|---------------------| | Prothrombin Time<br>(Innovin®, Seconds) | CA-6000 | 165 | 0.999 | Y = 0.97X + 0.09 | | Prothrombin Time<br>(Innovin®, INR) | CA-6000 | 165 | 0.999 | Y = 0.99X + 0.01 | | Prothrombin Time<br>(Thromborel® S, Seconds) | BCT | 163 | 0.998 | Y = 1.04X - 0.51 | | Prothrombin Time<br>(Thromborel® S, %) | BCT | 161 | 0.996 | Y = 1.08X - 2.44 | | Derived Fibrinogen | BCT | 149 | 0.945 | Y = 1.17X - 0.54 | | Activated Partial Thromboplastin Time | CA-6000 | 128 | 0.995 | Y = 1.00X + 0.15 | | Fibrinogen (Clauss) | CA-6000 | 115 | 0.985 | Y = 0.96X + 0.41 | | Factor VII | CA-6000 | 122 | 0.997 | Y = 1.04X - 1.67 | | Factor VIII | CA-6000 | 66 | 0.990 | Y = 0.96X + 4.25 | | Antithrombin III | BCT | 104 | 0.998 | Y = 0.97X - 4.72 | #### Summary of Method Comparison Studies Between CA-1500 and CA-6000 or BCT #### Summary of Precision Studies Sysmex® Automated Coagulation Analyzer CA-1500 | Sysmex® Automated Coagulation Analyzer CA-1500 | | | | | | | |-------------------------------------------------------------|-------------------|----|--------------|----------------------|-----------------------|--------------| | Test | Control<br>Level | N | Mean | Within<br>Run<br>%CV | Between<br>Run<br>%CV | Total<br>%CV | | Prothrombin Time<br>(Dade® Innovin®, Seconds) | CPN<br>Path. Pool | 40 | 11.5<br>23.3 | 0.4<br>1.2 | 0.5<br>0.9 | 0.6<br>1.4 | | Prothrombin Time<br>(Dade® Innovin®, INR) | CPN<br>Path. Pool | 40 | 1.1<br>2.2 | 0.4<br>1.2 | 0.5<br>0.9 | 0.6<br>1.4 | | Prothrombin Time<br>(Thromborel® S, Seconds) | CPN<br>Path. Pool | 40 | 11.8<br>28.4 | 0.6<br>0.6 | 2.0<br>3.0 | 2.1<br>3.0 | | Prothrombin Time<br>(Thromborel® S, %) | CPN<br>Path. Pool | 40 | 95.0<br>23.4 | 0.9<br>0.8 | 3.4<br>4.1 | 3.6<br>4.2 | | Derived Fibrinogen<br>(Dade® Innovin® on CA-1500) | CPN<br>Path. Pool | 40 | 1.8<br>5.0 | 3.7<br>2.4 | 3.2<br>1.2 | 4.7<br>2.6 | | Activated Partial Thromboplastin<br>Time (Dade® Actin® FSL) | CPN<br>Path. Pool | 40 | 28.1<br>39.5 | 2.1<br>0.5 | 1.4<br>1.9 | 2.4<br>2.0 | | Fibrinogen (Clauss)<br>(Dade® Thrombin Reagent) | CPN<br>CPP | 40 | 2.7<br>0.7 | 1.9<br>8.4 | 2.7<br>4.9 | 3.2<br>9.2 | | Factor VII<br>(Dade® Innovin®) | CPN<br>CPP | 40 | 91.6<br>32.5 | 2.5<br>2.3 | 1.5<br>2.1 | 2.7<br>3.0 | | Factor VIII<br>(Dade® Actin® FSL) | CPN<br>CPP | 40 | 89.1<br>26.0 | 1.6<br>1.8 | 3.3<br>4.3 | 3.6<br>4.6 | | Antithrombin III<br>(Berichrom® Antithrombin III(A)) | CPN<br>CPP | 40 | 97.4<br>32.3 | 1.3<br>2.7 | 4.6<br>7.6 | 4.8<br>8.1 | {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings or feathers. The eagle is facing to the left. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the eagle. DEC - 7 1999 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Mr. Radames Riesgo Manager, Regulatory Affairs Biology Dade Behring. Inc. 1851 Delaware Parkway Miami, Florida 33125 > K993299 Trade Name: Sysmex® Automated Coagulation Analyzer CA-1500 Regulatory Class: II Product Code: GKP Dated: September 30, 1999 Received: October 1, 1999 Dear Mr. Riesgo: Re: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4}------------------------------------------------ Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours. Steven Sutman Steven I. Gutman, M.D. M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ 510(k) Number (if known): K9932999 Device Name: Sysmex® Automated Coagulation Analyzer CA-1500 ### Indications for Use: The intended use of the Sysmex® CA-1500 is as a fully automated, computerized blood plasma coagulation analyzer for in vitro diagnostic use in clinical laboratories. The instrument uses citrated human plasma to perform the following parameters and calculated parameters: Clotting Analysis Parameters - Prothrombin Time (PT) . - Activated Partial Thromboplastin Time (APTT) . - Fibrinogen (Clauss) . Chromogenic Analysis Parameters - Antithrombin III . Calculated Parameters - . PT Ratio - PT INR - PT % ● - Extrinsic Factors (II, V, VII, X) ● Intrinsic Factors (VIII, IX, XI, XII) . - Derived Fibrinogen ● - Factor Assays % Activity . (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) ## Concurrence of CDRH, Office of Device Evaluation (ODE) Peter E. Madini Clinical Laboratory Device Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter-Use (Optional Format 1-2-96)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...