METHAFILCON A SOFT HYDROPHILIC CONTACT LENSES, FREQUENCY 55, FREQUENCY 55 UV (TO BE MARKETED YR. 2000), FREQUENCY 55 ASP
K993252 · CooperVision, Inc. · MVN · Nov 24, 1999 · Ophthalmic
Device Facts
Record ID
K993252
Device Name
METHAFILCON A SOFT HYDROPHILIC CONTACT LENSES, FREQUENCY 55, FREQUENCY 55 UV (TO BE MARKETED YR. 2000), FREQUENCY 55 ASP
Applicant
CooperVision, Inc.
Product Code
MVN · Ophthalmic
Decision Date
Nov 24, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Indications for Use
Methafilcon A lenses are intended for use as a daily wear lens for the correction of refractive ametropia (myopic, hyperopic and astigmatism) in not-aphakic persons with non-diseased eyes.
Device Story
Soft hydrophilic contact lenses composed of methafilcon A (random copolymer of hydroxyethylmethacrylic acid). Lenses correct refractive ametropia (myopia, hyperopia, astigmatism). Frequency 55 UV variant includes benzotriazole UV-absorbing monomer to block UVA/UVB radiation. Lenses are tinted for visibility. Used by patients with non-diseased eyes; prescribed by eye care practitioners. Lenses are cleaned, rinsed, and disinfected for frequent/planned replacement or discarded after single use for disposable wear. Benefit includes vision correction and UV protection for the eye.
Clinical Evidence
No clinical or non-clinical tests were performed. Substantial equivalence is based on identical design, material, and chemical composition to previously cleared predicate devices.
Technological Characteristics
Material: methafilcon A (random copolymer of hydroxyethylmethacrylic acid). Frequency 55 UV includes benzotriazole UV-absorbing monomer. Tinted with Reactive Blue No. 4 and/or Reactive Yellow 86. Soft hydrophilic contact lens form factor. Daily wear, frequent/planned replacement, or disposable use.
Indications for Use
Indicated for daily wear correction of refractive ametropia (myopia, hyperopia, astigmatism) in aphakic and not-aphakic persons with non-diseased eyes. Encore Toric indicated for astigmatism up to 12.00 diopters; others for astigmatism up to 2.00 diopters. Frequency 55 UV includes UV protection. Prescribed for frequent/planned replacement or disposable (single use) wear.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
{0}------------------------------------------------
1. Submitter's Name:
# NOV 2 4 1999
Image /page/0/Picture/2 description: The image shows the logo for CooperVision. The logo is black and white and features the company name in a stylized font. The "C" in Cooper is made up of two circles, one inside the other. The "n" in Vision is connected to a black square with a white "c" inside.
CooperVision, Inc. 711 North Road Scottsville, NY 14546 Phone: (716) 385-6810 FAX: (716) 889-5688
Bonnie Tsymbal Phone: (716) 264-3210 (716) 889-5688 FAX:
October 5, 1999
- 3. Date Summary Prepared:
- 4. Name of Device:
2. Contact Person:
- Frequency 55 Trade Name: . Frequency 55 UV Frequency 55 Aspheric Encore Encore Toric CooperFlex Onevue . Common Name: Soft Contact Lens Soft Hydrophilic Contact Lens . Classification Name: (Per 21 CFR §886.5925) 5. Legally Marketed Device: Same as Trade Name
- 6. Description of Device:
Frequency 55, Frequency 55 Aspheric, Encore, CooperFlex and Onevue (methafilcon A) Soft (hydrophilic) Contact Lenses are available as spherical lenses. Encore Toric (methafilicon A) Soft (hydrophilic) Contact Lenses are available as astigmatic (toric) lenses. The lens material, methafilcon A, is a random copolymer of hydroxyethylmethacrylic acid. The lenses are tinted from edge to edge for visibility purposes with the color additive, Reactive Blue No. 4.
Frequency 55 UV (methafilcon A) Soft (hydrophilic) Contact Lenses are available as spherical lenses. The lens material, methafilcon A, is a random copolymer of hydroxyethylmethacrylate and methacrylic acid. A benzotriazole UV absorbing monomer is used to block UV radiation. the average transmittance characteristics are less than 10% in the UVB range if 280 to 315 nm and less than 10% in the UVA range of 315 to 380 nm. The lenses are tinted aqua from edge to edge for visibility purposes with the color additives, C.I. Reactive Blue No. 4 and C.I. Reactive Yellow 86.
K993252
(716) 385-6810
Fax (716) 889-5688
{1}------------------------------------------------
- 7. Intended Use:
Methafilcon A lenses are intended for use as a daily wear lens for the correction of refractive ametropia (myopic, hyperopic and astigmatism) in not-aphakic persons with non-diseased eyes.
- 8. Technological Characteristics:
Design, material, and chemical composition are unchanged from those cleared in K971164, K973063, and K982997.
- 9. Summary of Non-Clinical/Clinical Tests:
No Non-Clinical or Clinical tests were performed. The lenses are the same as those cleared in the above premarket notifications.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The logo is in black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Nov 2 4 1999
Ms. Bonnie Tsymbal Coopervision, Inc. 711 North Road Scottsville, NY 14546
Re: K993252
Trade Name: Frequency 55, Frequency 55 UV, Frequency 55 Aspheric, Encore, Encore Tonc, CooperFlex and Onevue (methafilcon A) Soft (Hydrophilic) Contact Lenses for Daily Wear
Regulatory Class: II Product Code: LPL, MVN (Disposable Use) Dated: November 5, 1999 Received: November 8, 1999
Dear Ms. Tsymbal:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 - Ms. Bonnie Tsymbal
. . This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours.
ARalph Rosenthal
A. Ralph Rosenthal, M.I Director Division of Ophthalmic Devices Office of Device Evaluation
Approval, it may be subject to such accurional controls. Existing major regulations arrecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
### Indication for Use Statement
| 510(k) Number: | K993252 |
|----------------|-----------------------|
| Device Name: | Frequency 55 |
| | Frequency 55 UV |
| | Frequency 55 Aspheric |
| | Encore |
| | Encore Toric |
| | CooperFlex |
| | Onevue |
## Indications for Use:
- Frequency 55, Frequency 55 UV, Frequency 55 Aspheric, Encore, CooperFlex a 1. Onevue are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic and not-aphakic persons with non-diseased eyes.
The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with usual acuity.
Frequency 55 lenses with UV absorbing monomer helps protect against transmission of harmful UV radiation to the cornea and into the eye.
- 2. Encore Toric lenses are indicated for daily for the correction of refractive (myopia, hyperopia, and astigmatism) in aphakic and not-aphakic persons with nondiseased eyes. The lenses may be worn by persons who have astigmatism of 12.00 diopters or less.
### Frequent/Planned Replacement Wear
When prescribed for Frequent/Planned Replacement Wear, the Frequency 55, Frequency 55 Aspheric, Frequency 55 UV, Encore, Encore Toric, CooperFlex and Onevue lenses be cleaned, rinsed and disinfected each time they are removed from the eye and discarded
after the recommended wearing period prescribed by the eye care practitoner.
#### Disposable Wear
When prescribed for Disposable Wear, the wearing time prescribed by the eye practitioner is for daily wear (single use). Patients should be instructed to discard the lenses at each removal.
srfut
**Prescription Use**
(Per 21 CFR 801.109)
Division Sign-Off ision of Ophthalm Number K99
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.