← Product Code [LJM](/productcode/LJM) · K993108

# VARELISA RECOMBI ANA SCREEN EIA, MODELS 128 48/ 128 96 (K993108)

_Pharmacia & Upjohn Co. · LJM · Nov 12, 1999 · Immunology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K993108

## Device Facts

- **Applicant:** Pharmacia & Upjohn Co.
- **Product Code:** [LJM](/productcode/LJM.md)
- **Decision Date:** Nov 12, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5100
- **Device Class:** Class 2
- **Review Panel:** Immunology

## Indications for Use

The Varelisa® ReCombi ANA Screen EIA kit is designed for the qualitative determination of eight antinuclear antibodies in human serum or plasma to aid in the diagnosis of systemic rheumatic diseases such as SLE (Systemic Lupus Erythematosus), Scleroderma (Progressive Systemic Sclerosis), MCTD (Mixed Connective Tissue Disease), SS (Sjögren's Syndrome) and Polymyositis/ Dermatomyositis. The Varelisa ReCombi ANA Screen detects antibodies against dsDNA, RNP(68kDa, A, C), Sm(B,B',D) SS-A/Ro(52kDa, 60kDa), SS-B/La, Scl-70, Centromere and Jo-1 in a single microwell.

## Device Story

Varelisa ReCombi ANA Screen is an indirect noncompetitive enzyme immunoassay (EIA) for qualitative detection of 8 antinuclear antibodies in human serum or plasma. Microplate wells are coated with human recombinant, native affinity-purified nuclear antigens and dsDNA. Patient sample antibodies bind to these antigens; an enzyme-labeled conjugate is added to form an antigen-antibody sandwich complex. Substrate conversion produces color proportional to the concentration of bound antibodies. Used in clinical laboratories by trained personnel to aid diagnosis of systemic rheumatic diseases. Results are interpreted by healthcare providers to support clinical assessment of autoimmune conditions.

## Clinical Evidence

Bench testing only. Correlation study performed with 129 samples (10 reference sera, 20 healthy donors). Agreement was 97.5% (119/122) after excluding 7 dsDNA-positive samples. Linear regression analysis showed R^2=0.93, y=0.91x+0.05.

## Technological Characteristics

Indirect noncompetitive enzyme immunoassay (EIA). Microplate wells coated with human recombinant, native affinity-purified nuclear antigens and dsDNA. Qualitative colorimetric output proportional to antibody concentration.

## Regulatory Identification

An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).

## Predicate Devices

- Varelisa ANA-8-Screen

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS 9.

This summary of safety and effectiveness information is being submitted in accordance with the requirements of The Safety Medical Devices Act of 1990 (SMDA 1990) and 21 CFR Part 807.92.

| Assigned 510(k) Number:      | K993108                                                                                                                                                                                                                     |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date of Summary Preparation: | September 8, 1999                                                                                                                                                                                                           |
| Distributor:                 | Pharmacia & Upjohn<br>Diagnostics Division, US Operation<br>7425-248-1<br>7000 Portage Road<br>Kalamazoo, MI 49001                                                                                                          |
| Manufacturer:                | Pharmacia & Upjohn Diagnostics GmbH Co. KG<br>Munzingerstrasse 7<br>D-79111 Freiburg, Germany                                                                                                                               |
| Company Contact Person:      | Karen E.Matis<br>Manager, Regulatory Affairs and Quality<br>Management<br>Diagnostics Division<br>US Operation<br>7000 Portage Road<br>7425-248-01<br>Kalamazoo, MI 49001<br>(614) 794-3324 (Phone)<br>(614) 794-0266 (Fax) |
| Device Name:                 | Varelisa® ReCombi ANA Screen                                                                                                                                                                                                |
| Common Name:                 | Antinuclear antibody immunological test                                                                                                                                                                                     |

Classification:

| Product Name                 | Product Code | Class | CFR      |
|------------------------------|--------------|-------|----------|
| Varelisa® ReCombi ANA Screen | 82LJM        | II    | 866.5100 |

00 00078RAS9v5.doc

{1}------------------------------------------------

## Substantial Equivalence to:

Varelisa® ANA-8-ScreenAssay

#### Intended Use Statement:

The Varelisa® ReCombi ANA Screen EIA kit is designed for the qualitative determination of eight antinuclear antibodies in human serum or plasma to aid in the diagnosis of systemic rheumatic diseases such as SLE (Systemic Lupus Erythematosus), Scleroderma (Progressive Systemic Sclerosis), MCTD (Mixed Connective Tissue Disease), SS (Sjögren's Syndrome) and Polymyositis/ Dermatomyositis. The Varelisa ReCombi ANA Screen detects antibodies against dsDNA, RNP(68kDa, A, C), Sm(B,B',D) SS-A/Ro(52kDa, 60kDa), SS-B/La, Scl-70, Centromere and Jo-1 in a single microwell.

#### General Description of the Device

The Varelisa ReCombi ANA Screen is an enzyme immunoassay for the qualitative determination of antinuclear antibodies in serum or plasma. Designed as a screen assay, it detects 8 antinuclear antibodies in a single microwell. The determination of antinuclear antibodies (ANA) is of central importance for the clinical diagnosis of rheumatic diseases. The presence of ANA suggests the possibility of rheumatic autoimmune diseases. These diseases include Systemic Lupus Erythematosus, Polymyositis/ Dermatomyositis, Scleroderma, Sjögren's Syndrome and Mixed Connective Tissue Diseases.

## Varelisa® ReCombi ANA Screen, Test Principle

Varelisa ReCombi ANA Screen is an indirect noncompetitive enzyme immunoassay. The wells of a microplate are coated with human recombinant, native affinity purified nuclear antigens and dsDNA. Antibodies specific for the nuclear antigens present in a patient sample bind to these antigens.

In a second step the enzyme labeled second antibody (Conjugate) binds to the antigenantibody complex which leads to the formation of an enzyme labeled antigen-antibody sandwich complex.

The enzyme labeled antigen-antibody complex converts the added substrate to form a colored solution. The rate of color formation from the chromogen is a function of the amount of Conjugate complexed with the bound antibody and thus is proportional to the initial concentration of the respective antibodies in the patient sample.

00 00079

{2}------------------------------------------------

# Device Comparison:

A correlation study was performed comparing the new device. Varelisa ReCombi ANA Screen Assay, to the predicate device, Varelisa ANA-8-Screen. 129 samples, including 10 ANA Human Reference Sera and 20 apparently healthy blood donors were assayed using both tests.

Seven monospecific sera, equivocal and positive for dsDNA-Ab in the Varelisa ReCombi ANA Screen were excluded from the evaluation of the agreement between the assays. It is an expected result that these 7 sera would test positive with the new device and negative with the predicate device because the new device, Varelisa ReCombi ANA Screen contains antigens for the detection of dsDNA, whereas the predicate device, Varelisa ANA-8-Screen, does not detect the presence of dsDNA antibodies. With the elimination of the 7 dsDNA positive sera, the agreement in this study was 97.5% (119 out of 122 (129-7) samples). Three discordant samples were borderline either to the upper or lower limit of the equivocal zone and differ with a maximum of 0.3 Ratio.

A linear regression analysis of these data gives the following equation:

$$\begin{array}{|c|c|}\hline\hline\mathbf{y}=\mathbf{0.91x}+\mathbf{0.05}&\mathbf{R^2}=\mathbf{0.93}\\\hline\hline\end{array}$$

|            |           | Varelisa<br>ANA-8-Screen |           |          |
|------------|-----------|--------------------------|-----------|----------|
|            | n = 122   | positive                 | equivocal | negative |
| VARELISA   | positive  | 77                       | 0         | 0        |
| ReCombi    | equivocal | 1                        | 4         | 2        |
| ANA Screen | negative  | 0                        | 0         | 38       |

7 dsDNA-Ab positive samples were excluded from the evaluation

The new device Varelisa® ReCombi ANA Screen shows an excellent correlation with the predicate device, Varelisa ANA-8-Screen, and adds the ability to detect antibodies against dsDNA.

{3}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized symbol that resembles a human figure or a flame, composed of three curved lines.

NOV I 2 1999

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Karen E. Matis Manager, Regulatory Affairs and Quality Management Diagnostics Division, US Operations Pharmacia & Upjohn 7000 Portage Road 7425-248-01 Kalamazoo, Michigan 49001-0199

Re: K993108 Trade Name: Varelisa ReCombi ANA Screen EIA Regulatory Class: II Product Code: LJM Dated: September 16, 1999 Received: September 17, 1999

Dear Ms. Matis:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{4}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D. M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Varelisa® ReCombi ANA Screen 510(k) Submission Section 1. Indications For Use Statement

510(k) Number:

699308

Device Name: Varelisa® ReCombi ANA Screen

The Varelisa® ReCombi ANA Screen EIA kit is designed for the qualitative determination of eight antinuclear antibodies in human serum or plasma to aid in the diagnosis of systemic rheumatic diseases such as SLE (Systemic Lupus Erythematosus), Scleroderma (Progressive Systemic Sclerosis), MCTD (Mixed Connective Tissue Disease), SS (Sjögren's Syndrome) and Polymyositis/ Dermatomyositis. The Varelisa ReCombi ANA Screen detects antibodies against dsDNA, RNP(68kDa, A, C), Sm(B,B',D) SS-A/Ro(52kDa, 60kDa), SS-B/La, Sc1-70, Centromere and Jo-1 in a single microwell.

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use

(Per 21 CFR 801.109)

OR

Over-The-Counter Use

Utter A. Mackin'

(Division Sign-Off)
Division of Clinical Laboratory Devices K093108
510(k) Number

000.11 00

---

**Source:** [https://fda.innolitics.com/device/K993108](https://fda.innolitics.com/device/K993108)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
