K993027 · Specs USA, Inc. · LPH · Apr 6, 2000 · Orthopedic
Device Facts
Record ID
K993027
Device Name
ESKA MODULAR HIP SYSTEM
Applicant
Specs USA, Inc.
Product Code
LPH · Orthopedic
Decision Date
Apr 6, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ESKA Modular Hip System is indicated for use in the treatment of severely disabled hip joints resulting from painful osteo-, rheumatoid, and post-traumatic arthritis, and the late stages of avascular necrosis, and for the revision of previous hip surgeries.
Device Story
The ESKA Modular Hip System is a total hip prosthesis used for replacing severely disabled hip joints. The system comprises femoral stems (anatomically designed, left/right configurations, anteverted necks), femoral heads, acetabular shells, modular liners, and bone screws. It includes fully porous, proximally porous, collared, and collarless options, as well as pre-assembled bipolar and unipolar components. The device is implanted by a surgeon in a clinical setting to provide pain relief and restore function to joints disabled by arthritis or trauma. It functions as a mechanical replacement for the hip joint.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Materials: CoCrMo alloy and UHMWPE per ASTM and ISO standards. Design: Modular hip prosthesis with porous coating for uncemented fixation. Components: Femoral stems, heads, acetabular shells, liners, and bone screws.
Indications for Use
Indicated for patients with severely disabled hip joints due to osteoarthritis, rheumatoid arthritis, post-traumatic arthritis, avascular necrosis, non-union of previous arthrodesis, or requiring revision of previous hip surgeries.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
Predicate Devices
Omnifit Hip System (Osteonics)
Horizon Hip System (BioPro)
Sentry Hip System (Howmedica)
E-Series Hip System (Howmedica)
AML Hip System (DePuy)
VitaLock Acetabular Cup System (Howmedica)
OsteoLock Acetabular Cup System (Howmedica)
PCA Acetabular Cup System (Howmedica)
Ranawat / Burstein Total Hip System (Biomet)
Omega Total Hip System (OsteoImplant Technology)
Hedrocel Acetabular Cup System (Implex)
Submission Summary (Full Text)
{0}------------------------------------------------
K995027
# APR, 6 200(
#### Summary of Safety and Effectiveness for
## ESKA Modular Hip System
This safety and effectiveness summary for the ESKA Modular Hip System is provided as required per Section 513(i)(3) of the Food, Drug and Cosmetic Act.
1. Submitter : Contact Person : ESKA Implants GmbH & Co. Thomas P. Monkus, RAC GrapengieBerstraße 34 635 South Orange Avenue D-23556 Lübeck, Germany Sarasota, FL 34236 (0) 451- 89000-0 Telephone: ( 941 ) 362-4877
Date Prepared: August 31, 1999
2. Tradename : ESKA Modular Hip System Common Name : Total Hip System Classification Name : Hip joint metal /polymer /metal semi-constrained porous coated uncemented prosthesis ( 888.3358 )
### 3. Predicate or legally marketed devices which are substantially equivalent :
- Omnifit Hip System ( Osteonics )
- Horizon Hip System ( BioPro ) .
- . Sentry Hip System and E-Series Hip System ( Howmedica )
- . AML Hip System ( DePuy )
- VitaLock, OsteoLock and PCA Acetabular Cup Systems ( Howmedica ) .
- . Ranawat / Burstein Total Hip System ( Biomet )
- Omega Total Hip System ( OsteoImplant Technology ) .
- Hedrocel Acetabular Cup System ( Implex )
#### 4. Description of the device :
The ESKA Modular Hip System is a total hip system used for the replacement of severely disabled hip joints. It consists of femoral stems and heads, acetabular shells and modular liners, bipolar femoral heads, and bone screws. The femoral stems are anatomically designed in left and right configurations, with anteverted femoral necks. There are fully porous, proximally porous. collared and collarless options. The acetabular components are porous, with various liner options and screws for supplementary fixation. Pre-assembled bipolar and unipolar components are also available.
Materials : The devices are manufactured from CoCrMo alloy and Ultra High Molecular Weight Polycthylene ( UHMWPE ) per ASTM and ISO standards.
Function : The system functions to provide pain relief and improved function to the hip that has been disabled from arthritic conditions or trauma.
#### 5. Intended Use :
The ESKA Modular Hip System is indicated for use in the treatment of severely disabled hip joints resulting from painful osteo-, rheumatoid, and post-traumatic arthritis, and the late stages of avascular necrosis, and for the revision of previous hip surgeries.
### 6. Comparison of the technological characteristics of the device to predicate and legally marketed devices :
There are no significant differences between the ESKA Modular Hip System and other systems currently being marketed which would adversely affect the use of the product. It is substantially equivalent to these other devices in design, function, material and intended use.
{1}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.
Public Health Service
APR 6 2000
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Thomas P. Monkus, RAC Director, Regulatory & Quality Systems SPECS USA. Inc. 635 South Orange Avenue Sarasota, Florida 34236
Re: K993027 Trade Name: ESKA Modular Hip System Regulatory Class: II Product Code: LPH/JDI Dated: January 12, 2000 Received: January 14, 2000
Dear Mr. Monkus:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general control provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labcling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{2}------------------------------------------------
## Page 2 - Mr. Thomas P. Monkus, RAC
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-46.59. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Duna D. Lochner
(Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
of 1 Page 1
510(k) Number (if known): K993027
Device Name: ESKA Modular Hip System
Indications For Use:
The ESKA Modular Hip System is indicated for use in the treatment of rhe Boxi Noddrar hip Ojoobarthritis, post-traumatic arthritis, avascular rheumatord archiritis of courchires, post of the dislocation of the nccrosis) non anion of reasesful arthrodeses and revision of previous hip surgeries.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Dune R. Wodner
(Division Sign-Off Avision of General Restorative Devices K94302 510(k) Number
Prescription Use (Per 21 CFR 801 OR
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.