ENABLE ENDOSCOPIC BIPOLAR SCISSORS

K992996 · Enable Medical Corp. · GEI · Nov 24, 1999 · General, Plastic Surgery

Device Facts

Record IDK992996
Device NameENABLE ENDOSCOPIC BIPOLAR SCISSORS
ApplicantEnable Medical Corp.
Product CodeGEI · General, Plastic Surgery
Decision DateNov 24, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The ENABLE Endoscopic Bipolar Scissors intended use is substantially equivalent to the intended use statements of CardioThoracic, Symbiosis, Everest and ENABLE Bipolar Scissors. I bipolar scissors described in this section cut and coagulate soft tissue through the use of .polar technology. The ENABLE Endoscopic Bipolar Scissors are intended for use during laparoscopic and general surgical procedures.

Device Story

Device consists of handle connected to scissor blades with integrated electrodes; utilizes bipolar electrosurgical technology (RF energy) to cauterize blood vessels by heating tissue proteins. Input: RF energy from external electrosurgical generator (e.g., Valleylab Force 2). Output: thermal coagulation and mechanical cutting of soft tissue. Used in laparoscopic and general surgery; operated by surgeons. Current flows between electrodes on blades; no grounding pad required, reducing patient burn risk. Device fits standard 5mm cannula; length 33-48cm. Benefits include precise tissue management and elimination of return-electrode-related burns.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on bench testing and adherence to recognized industrial and international standards, including ISO 10993 (biocompatibility), ISO 11607 (packaging), ISO 11137 (sterilization), ANSI/AAMI HF18 (electrosurgical devices), IEC 60601-2-2 (safety of high-frequency surgical equipment), and ASTM F1079 (surgical scissors).

Technological Characteristics

Materials: stainless steel blades. Energy: bipolar RF (15-35 watts). Dimensions: 33-48cm length, fits 5mm cannula. Standards: ISO 10993, ISO 11607, ISO 11137, ANSI/AAMI HF18, IEC 60601-2-2, ASTM F1079. Sterilization: Gamma irradiation.

Indications for Use

Indicated for cutting and controlling bleeding through coagulation of soft tissue during laparoscopic and general surgical procedures.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ENABLE Medical Corporation Endoscopic 5100k) # Nov 2 4 1999 # 510(k) SUMMARY ENABLE ENDOSCOPIC BIPOLAR SCISSORS 510(k) NOTIFICATION K 992996 #### GENERAL INFORMATION | Manufacturer: | ENABLE Medical Corporation<br>6345 Centre Park Drive<br>West Chester, OH 45069-3863<br>(513) 755-7600<br>(513) 755-7676<br>Est. Reg. No. 1530251 | |----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Mark L. Friedman. Ph.D.<br>Vice President of Quality Assurance & Regulatory Affairs<br>ENABLE Medical Corporation | | Date Prepared: | [to be added after 510(k) process] | | DEVICE DESCRIPTION | | | Classification: | Class II | | Trade Name: | ENABLE Endoscopic Bipolar Scissors | | Generic/Common Name: | Electrosurgical cutting and coagulation device and<br>accessories<br>21CFR878.4400 | #### PREDICATED DEVICES - Symbiosis Bipolar Scissors (K950286 and K951387) し、 - Everest Medical Bipolar Scissors (K945975 and K955001) 2. - CardioThoracic Systems MIDCAB/SVH Bipolar Scissors (K963930) 3. - ENABLE Medical Corporation Bipolar Scissors (K972558 and K981219) 4. #### INTENDED USE The ENABLE Endoscopic Bipolar Scissors intended use is substantially equivalent to the intended use statements of CardioThoracic, Symbiosis, Everest and ENABLE Bipolar Scissors. I bipolar scissors described in this section cut and coagulate soft tissue through the use of .polar technology. The ENABLE Endoscopic Bipolar Scissors are intended for use during laparoscopic and general surgical procedures. {1}------------------------------------------------ P.05 # PRODUCT DESCRIPTION The ENABLE Endoscopic Bipolar Scissors are substantially equivalent to the CardioThoracic Systems. Symbiosis, Everest Medical and ENABLE Bipolar Scissors in that each consist of a handle which is connected to a pair of scissor blades with electrodes on each blade. All the devices utilize the same bipolar electrosurgical technology. i.e. radio frequency (RF) energy. 10 caurerize the blood vessels by heating proteins in the tissue to a temperature where the proteins congeal. The ENABLE Endoscopic Bipolar Scissors contain both a negative and positive electrode on each blade. The function of the ENABLE and predicate devices is the same; current flows from a negatively charged pole through the tissue to a positively charged pole. The ENABLE, CardioThoracic. Symbiosis and Everest Bipolar Scissors are connected to a similar electrosurgical generator that supplies RF bipolar energy. The ENABLE. CardioThoracic and Everest Bipolar Scissors are all labeled for use with the Valleylab Force 2 Electrosurgical Generator or equivalent electrosurgical units. All four bipolar scissors listed in this notification are labeled for operation at a comparable range (15-35 watts). Labeling for the ENABLE Endoscopic Bipolar Scissors are provided with each product. As with other bipolar instruments. there is no need for a grounding pad for the return electrode, therefore. the current does not travel through the patient to the grounding pad as it does with monopolar instruments. Patient burns seen with monopolar devices due to electrical current passage are eliminated. The ENABLE Endoscopic Bipolar Scissors and the listed predicate devices are of similar size with an overall length range of 33 to 48cm (13 to 19 inches) and a scissors blade length range of approximately 0.7 inches. The scissors shaft is designed to fit a standard 5mm cannula. The bipolar devices utilize similar materials of construction including stainless steel scissors blades. All of the materials utilized in the ENABLE Endoscopic Bipolar Scissors are standard medical device materials that are used in a variety of tissue contact medical devices. The ENABLE Endoscopic Bipolar Scissors will meet the following industrial/international standards. | ISO 10993/EN 30993 | Biological Evaluation of Medical Devices | |----------------------------|-------------------------------------------------------------------------------------------------------------| | ISO 11607 | Packaging for Terminally Sterilized Medical Devices | | ISO 11137 | Sterilization of Health Care Products, Sterilization of<br>Gamma Irradiation | | ANSI/AAMI HF18 | Electrosurgical Devices | | IEC 60601-2-2/EN 60601-2-2 | Medical Electrical Equipment: Particular<br>Requirements for Safety of High Frequency<br>Surgical Equipment | | ASTM F1079 | Standard Specification for Insert and Non-insert<br>Surgical Scissors | {2}------------------------------------------------ Nov-22-99 10:52A ENABLE Medical Corporation # SUMMARY As contained in this 510(k) summary, the ENABLE Endoscopic Bipolar Scissors are substantially equivalent to the predicate devices identified. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines extending from the bottom of the profiles, resembling a wave or a stylized bird. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 NOV 2 4 1999 Mark L. Friedman, Ph.D. Vice President of Quality Assurance and Regulatory Affairs Enable Medical Corporation 6345 Centre Park Drive West Chester, Ohio 45069 Re: K992996 > Trade Name: Endoscope Bipolar Scissors Regulatory Class: II Product Code: GEI Dated: August 31, 1999 Received: September 7, 1999 Dear Dr. Friedman: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4}------------------------------------------------ Food and Drug Administration 510(k) Notification - The Turbo 7000™ System September 3, 1999 K992994 510(k) Number: Device Name: The Turbo 7000™ System #### Intended Use: The Turbo 7000™ System's intended use is the cutting and removal of bone and tissue in general ENT, head & neck, and otoneurologic procedures. Otology procedures would include: - mastoidectomy and ● - mastoidotomy. . Sinus applications would embody: - septoplasty and . - . procedures such as - . the removal of septal spurs, - polypectomy, . - antrostomy, . - ethmoidectomy/sphenoethmoidectomy, . - frontal sinus trephination and irrigation, . - frontal sinus drill out, . - endoscopic DCR and . - trans-spehnoidal procedures. . Nasopharyngeal/Laryngeal procedures would comprise: - adenoidectomy, . - . tracheal, - laryngeal polyectomy, . - laryngeal lesion debulking and . - tonsillectomy. . Head and neck procedures would encompass: - . soft tissue shaving, - rhinoplasty (narrowing of the bony vault and revision of the bony pyramid), . - removal of fatty (adipose) tissue (lipo debridement) in the maxillary and mandibular regions of . the face, and - . acoustic neuroma removal. Additionally, an irrigation pump is integrated into the controller unit to provide irrigant in conjunction with the procedures. **Prescription Use** (Per 21 CFR 801.109) Page 3 (Division Sign-Off) Division of Ophthalmic Devices 510(k) Number K992994 {5}------------------------------------------------ Page_ of 1 1 K992996 510(k) Number (if known): Enable Endoscopic Bipolar Scissors Device Name:_ Indications For Use: The Enable Endoscopic Bipolar Scissors is intended to cut and and control blooding through coagulation during The Enable Endoscopic Biporal Sersboro culation during tissue and control bleeding through coagulation during the control control currical procedures. tissue and control breeding checugical bely (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Mark n Milkerson (Division Sign-Off) Division of General Restorative Devices 510(k) Number Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...