BAHO AUTOCLAVABLE CYSTOSCOPE

K992983 · Innovative Endoscopy Components, LLC · FAJ · Oct 7, 1999 · Gastroenterology, Urology

Device Facts

Record IDK992983
Device NameBAHO AUTOCLAVABLE CYSTOSCOPE
ApplicantInnovative Endoscopy Components, LLC
Product CodeFAJ · Gastroenterology, Urology
Decision DateOct 7, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Like the predicate devices, the BAHO autoclavable Cystoscope is indicated for use in the visual examination of body cavities, hollow organs and canals and using additional accessories, to perform various diagnostic and therapeutic procedures.

Device Story

BAHO autoclavable Cystoscope is a rigid endoscope consisting of an eyepiece, body, and insertion tube. The insertion tube features a double-walled surgical steel construction with integrated illumination fibers and an optical rod lens system. The device is used by physicians in clinical settings to provide visualization of internal body cavities and hollow organs. By connecting to external light sources and imaging systems, it enables the clinician to perform diagnostic and therapeutic procedures. The device is designed to be autoclavable for sterilization and reuse. It functions similarly to standard rigid cystoscopes, providing direct optical visualization to assist in clinical decision-making and patient treatment.

Clinical Evidence

Bench testing only. No clinical data provided. Substantial equivalence is based on design, material composition, and functional comparison to legally marketed predicate devices.

Technological Characteristics

Rigid endoscope; double-walled surgical steel insertion tube; optical rod lens system; integrated illumination fibers; autoclavable for sterilization. No software or electronic components.

Indications for Use

Indicated for visual examination of body cavities, hollow organs, and canals, and for use with accessories to perform diagnostic and therapeutic procedures in patients requiring endoscopic intervention.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## OCT 7 1999 ## SUMMARY OF SAFETY AND EFFECTIVENESS 4992983 1 ## BAHO INNOVATIVE MEDIZINTECHNIK, GmbH Autoclavable Cystoscopes ## 1) Submitter Information : | Name and Address : | Innovative Endoscopy Components, LLC<br>1112 Weston Road, PMB 227<br>Ft. Lauderdale, FL 33326 | |----------------------|-----------------------------------------------------------------------------------------------| | Telephone : | (954) 217-8780 | | Fax : | (954) 217-8781 | | E-mail : | IECflorida@aol.com | | FDA Registration No. | 1064152 | 2) Manufacturer Information : BAHO Innovative Medizintechnik GmbH Im Stoeckacker 7 D- 79224 Umkirch Germany | Telephone : | (011)49-7665-99094 | |------------------------|--------------------| | Fax : | (011)49-7665-99095 | | FDA Registration No. | 9615001 | | FDA Owner/Operator No. | 9026517 | 3) Device Name : | Classification Name | Cystoscope<br>Identical Materials,Components and Procedures as<br>BAHO autoclavable Laparoscope 510(k) Number K982276 | |-----------------------|-----------------------------------------------------------------------------------------------------------------------| | Classification Number | 876.1500 ( Class II Device ) | | Proprietary Name | BAHO autoclavable Cystoscope | 4) Predicate Device : Storz Model # 27005 Wolf Model # 8650.../8660... | Manufacturer : | Karl Storz | Richard Wolf | |----------------|-------------------|---------------------| | | Mittelstrasse 8 | Postfach 1164 | | | Postfach 230 | D-75434 Knittlingen | | | D-4200 Tuttlingen | Germany | | | Germany | | {1}------------------------------------------------ The above identified predicate devices have the basic design and technological features equivalent to those in the proposed BAHO autoclavable Cystoscope. K992983 pg 2 of 2 - 5) Description of Device : The BAHO autoclavable Cystoscope has a basic design similar to those legally available for sale in the U.S.A. Like others legally sold in the U.S.A.,it consists of an evepiece.body and insertion tube. The insertion tube is double walled surgical steel with illumination fibers in between. The insertion tube contains the optical rod lens system. ( identical to BAHO autoclavable Laparoscope 510(k) Number K982276 ) - 6) Indication for use : Like the predicate devices the BAHO autoclavable Cystoscope is indicated for use in the visual examination of body cavities, hollow organs and using additional accessories, to perform various diagnostic and therapeutic procedures. - 7) Description of safety and substantial equivalence : The biological safety of the BAHO autoclavable Cystoscope has been defined through the selection of materials that demonstrated appropriate levels of biocompatibility which constitute the building blocks of the proposed device. These materials are similar or identical to those used for the manufacturing of the predicate devices as well as other brands legally sold in the U.S.A. and other approved BAHO endoscopes ( Arthroscope, Laparoscope ) - 8) Summary In summary : biocompatibility, function, indications and designs have been developed to ensure the safety of this device.It is substantially equivalent to commercially approved cystoscopes available for sale in the U.S.A. {2}------------------------------------------------ Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 7 1999 OCT Mr. Gerald Goigitzer Managing Director Innovative Endoscopy Components, LLC 1112 Weston Road. Suite 227 Ft. Lauderdale, Florida 33326 Re: K992983 BAHO Autoclavable Cystoscope Dated: August 21, 1999 Received: September 3, 1999 Regulatory Class: II 21 CFR §876.1500/Procode: 78 FAJ Dear Mr. Goigitzer: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If vour device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and . Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control . provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ of Page / 510(k) Number (if known): K992983 Device Name: BAHO AUTOCLAVABLE CYSTOSCOPE Indications For Use: Like the predicate devices, the BAHO autoclavable Cystoscope is indicated for use in the visual examination of body cavities, hollow organs and canals and using additional accessories, to perform various diagnostic and therapeutic procedures. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96) Yhird le. Sypm **(Division Sign-Off)** Division of Reproductive, Abdominal, ENT, and Radiological Devic 510(k) Number.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...