K992937 · Nobel Biocare Uas, Inc. · DZE · Feb 29, 2000 · Dental
Device Facts
Record ID
K992937
Device Name
DENTAL IMPLANT
Applicant
Nobel Biocare Uas, Inc.
Product Code
DZE · Dental
Decision Date
Feb 29, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Selected Brånemark System® implant products (those identified in Table 7.1, below) are intended to be placed in the upper or lower jaw to support prosthetic devices. such as artificial teeth, and to restore a patient's chewing function. This may be accomplished using either a two stage surgical procedure or a single stage surgical procedure. If a single stage surgical procedure is used, these implants may be loaded immediately following insertion - PROVIDED - at least four implants are placed, and are splinted with a bar. These implants must be placed predominantly in the anterior mandible (between the mental foramina) where good initial stability of the implants, with or without bi-cortical anchorage, can most often be obtained.
Device Story
Brånemark System® dental implants are threaded, root-form devices fabricated from ASTM grade 1 commercially pure titanium. Used by dental surgeons in clinical settings to support prosthetic devices like artificial teeth and restore chewing function. Implants are placed in the upper or lower jaw via single or two-stage surgical procedures. For immediate loading (within 2-3 weeks), at least four implants must be placed and splinted with a bar, specifically in the anterior mandible to ensure initial stability. The device provides a stable foundation for prosthetic restoration, benefiting patients by enabling immediate functional use of dental prostheses.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on design, material, and intended use similarities to legally marketed predicate devices.
Technological Characteristics
Threaded, root-form dental implants; material: ASTM grade 1 commercially pure titanium; diameters: 3.75 mm and 4.0 mm; lengths: 10 mm to 21 mm; sterilization: dry heat (glass ampoule) or steam (blister pack).
Indications for Use
Indicated for patients requiring dental prosthetic support in the upper or lower jaw. For immediate loading, indicated for patients receiving at least four implants splinted with a bar, placed predominantly in the anterior mandible between the mental foramina where good initial stability is achievable.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
FEB 2 9 2000
Nobel Biocare USA, Inc. 510(k) Notification: Immediate Loading Indication June 21, 1999
K992937
## Section 7 510(k) Summary
### A. Manufacturer Information:
Nobel Biocare USA. Inc. Submitter's Name: Address: 22725 Savi Ranch Parkway U.S. Representative/Distributor Yorba Linda, CA 92887, USA
Contact's Name: Contact's Telephone No .: Date Prepared:
Jeff Hausheer, Ph.D., Regulatory Affairs Specialist 714-282-4800, extension 7832 June 1999
Address-Manufacturer:
Nobelpharma Production AB Dimbovagen 2 Karlskoga S-691-51, Sweden Manufacturer Registration Number: 9611993
### B. Device Name:
Common Name:
Dental Implant
Trade Name:
Brånemark System® Implants Indicated For Loading Immediately Following Implant Placement (see table below) (i.e., within 2 to 3 weeks following implant placement)
| Brånemark System®: Implant Product<br>Family Name | Dimensions: Diameter &<br>Range of Lengths (mm) |
|---------------------------------------------------|-------------------------------------------------|
| Standard Series Implants (3.75) | 3.75 x 10-20 mm |
| Standard Series Implants (4.0) | 4.0 x 10-18 mm |
| Mk II Self-Tapping Implants (3.75) | 3.75 x 10-18 mm |
| Mk II Self-Tapping Implants (4.0) | 4.0 x 10-18 mm |
| Self-Tapping Implants | 3.75 x 10-18 mm |
| Conical Self-Tapping Implants | 3.75 x 13-21 mm |
| Mk IV Self-Tapping Implants | 4.0 x 10-18 mm |
### Classification:
| Classification Name: | Endosseous Dental Implant |
|--------------------------|---------------------------|
| Classification Number: | DZE |
| Classification Citation: | 21 CFR 872.3640 |
Page 1 of 3
{1}------------------------------------------------
# Section 7 510(k) Summary (continued)
### C. Device Description:
Description: Brånemark System® dental implants are threaded, root-form implants fabricated from ASTM grade 1 "commercially pure" titanium. They are available in diameters of 3.75 mm and 4.0 mm, and are available in lengths ranging from 10 mm to 21 mm. Research studies have demonstrated that titanium is biocompatible.
### D. Intended Use:
Indications For Use: Selected Brånemark System® implant products (those identified in Table 7.1, below) are intended to be placed in the upper or lower jaw to support prosthetic devices. such as artificial teeth, and to restore a patient's chewing function. This may be accomplished using either a two stage surgical procedure or a single stage surgical procedure.
If a single stage surgical procedure is used, these implants may be loaded immediately following insertion - PROVIDED - at least four implants are placed, and are splinted with a bar. These implants must be placed predominantly in the anterior mandible (between the mental foramina) where good initial stability of the implants, with or without bi-cortical anchorage, can most often be obtained.
| Brånemark System®<br>Implant Product Family | Diameter | Range of Lengths | 510(k) |
|---------------------------------------------|----------|------------------|---------|
| Mk II Self-Tapping | 3.75 mm | 10-18 mm | K925762 |
| Mk II Self-Tapping | 4.0 mm | 4.0 x 10-18 mm | K945398 |
| Standard Series | 3.75 mm | 3.75 x 10-20 mm | K925765 |
| Standard Series | 4.0 mm | 4.0 x 10-18 mm | K925764 |
| Self-Tapping | 3.75 mm | 3.75 x 10-18 mm | K925762 |
| Conical Self-Tapping | 3.75 mm | 3.75 x 13-21 mm | K925760 |
| Mk IV Self-Tapping | 4.0 mm | 4.0 x 10-18 mm | K974828 |
| Table 7.1 |
|--------------------------------------------------------------------|
| 510(k) Notification: |
| Brånemark System® Implant Products Indicated For Immediate Loading |
{2}------------------------------------------------
Nobel Biocare USA, Inc. 510(k) Notification: Immediate Loading Indication June 21, 1999
# Section 7: 510(k) Summary (continued)
#### E. Comparison to the Predicate Device(s): .
### Comparison Of The Predicate Devices And The Branemark System® Implant Products For Which Clearance Of An Immediate Load Indication Is Sought
| | Predicate Product | Predicate Product | Submitted Product |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic | Straumann ITI Implant (510(k)<br>K984104) | Sargon Cylindro-Blade Implant [510(k)<br>K930071] | Brånemark System® Implant Products** |
| Intended Use | intended to be placed in the<br>maxillary and/or mandibular<br>arches to support prosthetic<br>restorations in edentulous or<br>partially edentulous patients. | Intended to act as a replacement for<br>missing teeth by providing the means for<br>fixation of dentures, removable bridgework,<br>or prosthetic teeth. | (Functionally, the same): intended to be placed<br>in the upper or lower jaw to support prosthetic<br>devices such as artificial teeth, and to restore<br>patient's chewing function. |
| Indication | Immediate Load | Immediate Load | Same |
| Design | threaded, root-form implant | a threaded, tapered, non-solid (hollow),<br>flanged root form implant | threaded, root-form implants |
| Placement Method | Single stage surgery | Single stage surgery | Same |
| Material | Commercially pure titanium | Titanium alloy | Commercially pure titanium |
| Coating | None | None | None |
| Length (mm, min.-<br>max.) | Unknown | 10mm to 18 mm | 10mm to 21 mm |
| Diameter (mm) | Unknown | Available in 1 tapered design only, ran<br>Available in one tapered design only; its<br>diameter ranges from 3.8 mm to 4.1<br>ranging from 3.8 mm to 4.1 mm | 3.75 mm and 4.0 mm |
| Precautions &<br>Warnings | Unknown | None | None |
| Packaging | Unknown | Vial | Glass ampoule in peel-open blister pack |
| Provided Sterile? | Unknown | Unknown | Yes (dry heat = glass ampoule; steam = blister<br>pack) |
= See Table 2.A.1 for identification, dimensions, and 510(k) numbers of each product
Page 3 of 3
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling three overlapping, curved shapes that could be interpreted as abstract representations of human figures or flowing elements.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 2 9 2000
Dr. Jeff Hausheer Requlatory Affairs Specialist Nobel Biocare USA, Inc. 22895 Eastpark Drive Yorba Linda, CA 92887
Re: K992937 Brånemark System® Implants Trade Name: Requlatory Class: III Product Code: DZE Dated: December 2, 1999 Received: December 6, 1999
Dear Dr. Hausheer:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, qood manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਉ substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of
{4}------------------------------------------------
Page 2 -Dr. Hausheer
the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA findinq of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4690. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Patricio Cavente/for
Timothy A. Ulatowski Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Nobel Biocare USA. Inc. 510(k) Notification: Immediate Loading Indication June 21, 1999
### Section 10 Indications for Use
Page 1 of 1
50
### 510(k) Number (if known): K ------
Device Name:
Selected Brånemark System® Implant Products:
| Mk II Self-Tapping | K925762 |
|----------------------|---------|
| Mk II Self-Tapping | K945398 |
| Standard Series | K925765 |
| Standard Series | K925764 |
| Self-Tapping | K925762 |
| Conical Self-Tapping | K925760 |
| Mk IV Self-Tapping | K974828 |
Indications For Use: Selected Brånemark System® implant products (those identified in the preceding section, "Device Name") are intended to be placed in the upper or lower jaw to support prosthetic devices, such as artificial teeth, and to restore a patient's chewing function. This may be accomplished using either a two stage surgical procedure or a single stage surqical procedure.
If a single stage surgical procedure is used, these implants may be loaded immediately following insertion - PROVIDED - at least four implants are placed, and are splinted with a bar. These implants must be placed predominantly in the anterior mandible (between the mental foramina) where good initial stability of the implants, with or without bi-cortical anchoragel can most often be obtained.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) n of Dental, Infection
Numbers Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use __ L OR Over-The-Counter Use (Per 21 CFR 801.109) (Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.