LEKSELL IMAGE GUIDANCE SURGICAL SYSTEM, VIEW SCOPE, SURGISCOPE,ELEKTA INSIGHT

K992882 · Elekta Instrument AB · HAW · Sep 24, 1999 · Neurology

Device Facts

Record IDK992882
Device NameLEKSELL IMAGE GUIDANCE SURGICAL SYSTEM, VIEW SCOPE, SURGISCOPE,ELEKTA INSIGHT
ApplicantElekta Instrument AB
Product CodeHAW · Neurology
Decision DateSep 24, 1999
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 882.4560
Device ClassClass 2

Intended Use

The Leksell® Image Guidance Surgical System (LIGS) is intended for use in cranial and spinal neurosurgical procedures to provide image guidance based on preoperative images which are visualized interactively with the aid of surgical tools.

Device Story

Leksell Image Guidance Surgical System (LIGS) provides intraoperative image guidance for cranial and spinal neurosurgery. System utilizes preoperative images to visualize surgical site interactively. Modifications include enhanced localization via infrared camera and reflective spheres on instruments, enabling wireless probe functionality. Universal Instrument Adapter allows fixation to various surgical tools. System used in clinical settings by neurosurgeons. Output consists of real-time image-guided visualization of surgical tools relative to patient anatomy, assisting in surgical navigation and decision-making. Benefits include improved accuracy and repeatability during neurosurgical procedures.

Clinical Evidence

Bench testing only. Performance testing conducted on Elekta Insight/ViewScope 2.5 system using simulated cranial and spinal models to evaluate product accuracy and repeatability based on specifications and risk analysis. All results were acceptable.

Technological Characteristics

System utilizes infrared camera-based localization with reflective spheres on instruments for wireless tracking. Includes Universal Instrument Adapter for tool fixation. Software updated to support enhanced localization. Connectivity/integration details not specified beyond system-level components.

Indications for Use

Indicated for use in cranial and spinal neurosurgical procedures to provide image guidance based on preoperative images visualized interactively with surgical tools.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Leksell® Image Guidance Surgical System Special Premarket 510(k) Notification: Device Modifications K992ff2 # SUMMARY OF SAFETY AND EFFECTIVENESS SECTION 18: This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and CFR 807.92. #### SUBMITTER INFORMATION 18.1 | a. | Company Name: | Elekta Instruments, AB | |----|-----------------------------------------|-------------------------------------------------------------------| | | b. Company Address: | Birger Jarlsgatan 53, S-103 93<br>Stockholm, Sweden | | | c. Company Phone:<br>Company Facsimile: | (011) 46 8 5872 54 00<br>(011) 46 8 5872 55 00 | | | d. Contact Person: | Sverker Glans<br>Vice President<br>Quality and Regulatory Affairs | | | e. Date Summary Prepared: | August 24, 1999 | #### DEVICE IDENTIFICATION 18.2. | a. Trade/Proprietary Name: | Leksell® Image Guidance Surgical System | |----------------------------|-------------------------------------------| | b. Classification Name: | Stereotaxic Instrument<br>21 CFR 882.4560 | #### IDENTIFICATION OF PREDICATE DEVICES 18.3 | Company | Device | 510(k) No. | Date Cleared | |---------------------------|--------------------------------------------|------------|--------------| | Elekta<br>Instruments, AB | Leksell® Image Guidance<br>Surgical System | K973684 | 03/31/98 | | BrainLAB | Vector Vision2 | K983831 | 05/19/99 | {1}------------------------------------------------ #### DEVICE DESCRIPTION 18.4 The Leksell® Image Guidance Surgical System (LIGS) is intended for use in cranial and spinal neurosurgical procedures to provide image guidance based on preoperative images which are visualized interactively with the aid of surgical tools. The LIGS System was cleared for commercial distribution under K973684 on March 31, 1998. The purpose of this special premarket 510(k) notification is to describe the changes in product specifications and system modifications. The LIGS System has not changed in its functionality or its indications for use. Technological enhancements have been made to the ViewScope/Elekta Insight Localization System and Surgical Tools. As a result of these enhancements, the software and instruction manuals have also been updated. ### SUBSTANTIAL EQUIVALENCE 18.5 The Leksell® Image Guidance Surgical System with its proposed modifications, is substantially equivalent to the current Leksell Image Guidance Surgical System and the BrainLAB Vector Vision2. The fundamental technical characteristics of the LIGS System are similar to those of the predicate devices. The functionality and the indications for use have not changed with the proposed modifications. The addition of the wireless probe feature and use of illuminator infrared camera and universal instrument adapter is equivalent to the features found in the BrainLAB Vector Vision2 predicate device. The Indications for Use for the LIGS System is also is equivalent to the BrainLAB Vector Vision2 predicate device. {2}------------------------------------------------ K992ff2 #### 18.6 INDICATIONS FOR USE The Leksell® Image Guidance Surgical System (LIGS) is indicated for use in cranial and spinal neurosurgical procedures to provide image guidance based on preoperative images which are visualized interactively with the aid of surgical tools. ## TECHNOLOGICAL CHARACTERISTICS 18.7 Modifications have been made to the Leksell® Image Guidance Surgical System in regards to the Elekta Insight/ViewScope models. The localization technology has been enhanced to include an illuminator infrared camera and reflective spheres on the instruments, allowing for wireless probe options. In addition, the surgical tools have been enhanced to include a Universal Instrument Adapter for fixation to the user's instruments. Comparison of the technological characteristics to those of the predicate devices has been provided in this submission. #### 18.8 PERFORMANCE DATA Performance testing was conducted on the Elekta Insight/ViewScope 2.5. System and component testing was completed based on product specifications and hazard effects determined from the risk analysis. Performance testing on simulated cranial and spinal models was conducted for product accuracy and repeatability. All results of testing was found to be acceptable. The Leksell® Image Guidance Surgical System with the proposed modifications performed as intended. #### 18.9 510(K) CHECKLIST This notification contains all information required by 21 CFR 807.87. A completed copy of the Premarket Notification 510(k) Reviewer's Checklist is provided in this submission. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, facing to the right. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP 2 4 1999 Elekta Instruments, AB C/o Ms. Carol Patterson Consultant Patterson Consulting Group, Inc. 21911 Erie Lane Lake Forest, California 92630 Re: K992882 Trade Name: Leksell® Image Guidance Surgical System Regulatory Class: II Product Code: HAW Dated: August 24, 1999 Received: August 26, 1999 Dear Ms. Patterson: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Ouality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4}------------------------------------------------ Page 2 - Ms. Carol Patterson This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ K992FF2 # INDICATION FOR USE To Be Assigned By FDA 510(k) Number: Leksell® Image Guidance Surgical System Device Name: Indications for Use: The Leksell® Image Guidance Surgical System (LIGS) is intended for use in cranial and spinal neurosurgical procedures to provide image guidance based on preoperative images which are visualized interactively with the aid of surgical tools. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Pallez (Division Sign-Off) of General Restorative Devices K992882 510(k) Number Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use # CONFIDENTIAL
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...