← Product Code [DRB](/productcode/DRB) · K992863

# LEAD STYLET,MODELS 6505,6506,6507,6508 (K992863)

_Guidant Corp. · DRB · Nov 12, 1999 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K992863

## Device Facts

- **Applicant:** Guidant Corp.
- **Product Code:** [DRB](/productcode/DRB.md)
- **Decision Date:** Nov 12, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1380
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

These stylets are wires intended to facilitate the placement of pacing leads during pacemaker implant procedures. The straight stylets are used to direct passage of a lead to the appropriate heart chamber and to wedge the lead in the trabeculae of the ventricle. The J-shaped stylets are used to direct passage of a lead in the atrium.

## Device Story

Accessory Stylet Kit consists of stainless steel wires used during pacemaker implantation. Stylets are inserted into pacing leads to provide structural support and directional control during lead placement. Straight stylets facilitate lead navigation to heart chambers and wedging into ventricular trabeculae; J-shaped stylets facilitate navigation within the atrium. Device is used by physicians in clinical settings (e.g., OR/cath lab) during lead implantation. Stylets are provided in 0.014 and 0.016 inch diameters with varying stiffness (soft/firm). Device facilitates accurate lead positioning, potentially improving procedural outcomes and patient safety.

## Clinical Evidence

No clinical data. Bench testing only. Biocompatibility and design verification testing from previous stylet submissions (P930060/S41) were referenced to support the safety and effectiveness of the subject device.

## Technological Characteristics

Materials: Chromium-nickel 304 stainless steel spring wire; polypropylene Hercules PD 701 knobs. Form factor: 0.014 and 0.016 inch diameter wires, straight or J-shaped. Packaging: Coiled polyethylene stylet ring. Sterilization: Identical to predicate devices. No software or electronic components.

## Regulatory Identification

A catheter stylet is a wire that is run through a catheter or cannula to render it stiff.

## Predicate Devices

- Accessory Stylet Kit Models 6501, 6502, 6503 and 6504 ([K905674](/device/K905674.md))
- Leads ([K893957](/device/K893957.md))
- Leads ([K932103](/device/K932103.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

NOV 1 2 1999

K992863

# 510(k) SUMMARY

- 1. Device Name: Accessory Stylet Kit Models 6504, 6505, 6506, 6507
- 2. Devices to Which Equivalence is Claimed: With the exception of a difference in length, the stylets in the new accessory kits are identical to stylets currently packaged with leads (K893957, cleared 7/14/89 and K932103, cleared 12/8/93). With respect to packaging, the new accessory kits are identical to the currently marketed Accessory Stylet Kit Models 6501, 6502, 6503 and 6504 (K905674, cleared 1/30/91).

#### 3. Intended Use:

These stylets are wires intended to facilitate the placement of pacing leads during pacemaker implant procedures.

The straight stylets are used to direct passage of a lead to the appropriate heart chamber and to wedge the lead in the trabeculae of the ventricle.

The J-shaped stylets are used to direct passage of a lead in the atrium.

# 4. Device Description:

The new accessory kits will allow separate packaging of lead stylets currently marketed with Guidant pace/sense and defibrillation leads. The stylets are provided in 0.014 and 0.016 inch diameters, with their respective "stiffness" labeled soft and firm. Each stylet package will contain two stylets of the same diameter. The stylets are made of chromium-nickel 304 stainless steel spring wire and have polypropylene Hercules PD 701 Knobs. The stylets are inserted into a coiled polyethylene stylet ring for packaging and ease of use purposes.

# 5. Summary of Technological Characteristics:

This notification concerns the separate packaging of currently marketed stylets using packaging materials, configurations and processes identical to currently marketed accessory kits. There are no changes to the design, materials, manufacturing process, packaging process, sterilization process, or shelf life of the subject device.

# 6. Summary of Substantial Equivalence:

The indications for use and packaging for the stylets are identical to the currently marketed stylets, Models 6501, 6502, 6503 and 6504 (K905674, cleared 1/30/91). All other aspects of the proposed accessory kits (Models 6505, 6506, 6507 and 6508) 

{1}------------------------------------------------

are identical to the currently marketed stylets (K893957, cleared 7/14/89 and K932103, cleared 12/8/93)

# 7. Testing Data:

This notification concerns the separate packaging of currently marketed stylets using packaging materials, configurations and processes identical to currently marketed accessory kits. There are no changes to the design, materials, manufacturing process, packaging process, sterilization process, or shelf life of the subject device. Therefore, no additional testing was deemed necessary. Biocompatibility Testing and Design Verification Testing of previous stylets were referenced in support of the proposed accessory kits (P930060/S41, approved 9/20/96).

# 8. Conclusion:

The separate stylet accessory kits, Model 6504, 6505, 6506 and 6507 are substantially equivalent to the currently marketed Guidant stylet accessory kits (K905674, cleared 1/30/91) with regard to packaging materials, configurations, processing and sterilization. The stylets are identical to the stylets currently marketed with pacing leads (K893957, cleared 7/14/89 and K932103, cleared 12/8/93).

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized emblem featuring a symbol that resembles three overlapping human figures or abstract shapes, possibly representing health and human services.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV I 2 1999

Ms. Sheryl Poganski Guidant Corporation Cardiac Rhythm Management 4100 Hamline Avenue North St. Paul, MN 55112-5798

K992863 Re: Lead Stylet, Models 6505, 6506, 6507 and 6508 Regulatory Class: II (two) Product Code: 74 DRB Dated: August 24, 1999 Received: August 25, 1999

Dear Ms. Poganski:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to leqally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such Failure to comply with the GMP regulation may result in assumptions. regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2 - Ms. Sheryl Poganski

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Clistoplystherfor

Celia M. Witten, Ph.D., M.D. Acting Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# INDICATIONS FOR USE STATEMENT

510(k) Number (if known): مامارك 3996_

Device Name: Lead Stylet

### Indications for Use:

These stylets are wires intended to facilitate the placement of pacing leads during pacemaker implant procedures.

The straight stylets are used to direct passage of a lead to the appropriate heart chamber and to wedge the lead in the trabeculae of the ventricle.

The J-shaped stylets are used to direct passage of a lead in the atrium.

# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use (Per 21 CFR 801.109) | X                                         |
|---------------------------------------|-------------------------------------------|
|                                       | OR                                        |
|                                       | Over-The-Counter (Optional Format 1-1-96) |

(Division Sign-Off)

Division of Cardiovascular, Respiratory, and Neurological Devices

510(k) Number K99281.3

2

---

**Source:** [https://fda.innolitics.com/device/K992863](https://fda.innolitics.com/device/K992863)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
