← Product Code [QPB](/productcode/QPB) · K992855

# XPS STRAIGHTSHOT OR MODEL 2000 MICRODEBRIDER SYSTEM OR XPS POWERSCULPT (K992855)

_Xomed, Inc. · QPB · Jan 21, 2000 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K992855

## Device Facts

- **Applicant:** Xomed, Inc.
- **Product Code:** [QPB](/productcode/QPB.md)
- **Decision Date:** Jan 21, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.5040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The XPS/PowerSculpt System with reciprocating adapter and suction cannula is indicated for the removal of soft tissue and fluid from the body during general surgical procedures including suction lipoplasty for aesthetic body contouring.

## Device Story

System consists of power control unit, handpiece, reciprocating adapter, and suction cannulae. Device transforms rotary action of handpiece into reciprocating motion via adapter to facilitate cannula movement through adipose tissue. Used in surgical settings by surgeons to perform tissue debridement and aspiration. Reciprocating motion simulates manual surgeon movement, reducing manual effort and aiding tissue penetration. Output is physical removal of soft tissue and fluid via suction. Benefits include reduced surgeon fatigue and improved efficiency in adipose tissue dissection.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on technological characteristics and design similarities to predicate devices.

## Technological Characteristics

System comprises an AC-powered control unit, handpiece, removable reciprocating adapter, and suction cannulae. The adapter converts rotary handpiece motion to reciprocating motion. Design is similar to existing powered aspiration devices. No specific material standards or software algorithm details provided.

## Regulatory Identification

A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.

## Special Controls

*Classification.* Class II (special controls). Consensus standards and labeling restrictions.

## Predicate Devices

- MicroAire PAD-100 Power Aspiration Device ([K981922](/device/K981922.md))
- NuMED Power Cannula (510k unknown)
- Wells Johnson Company Aspiration Cannulae ([K832520](/device/K832520.md))

## Reference Devices

- XPS handpieces ([K984363](/device/K984363.md))
- XPS rasping adapter ([K983025](/device/K983025.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Xomed, Inc. David Timlin Manager Regulatory Affairs 6743 Southpoint Dr. North Jacksonville, Florida 32216-0980

June 8, 2021

Re: K992855

Trade/Device Name: Xps Straightshot Or Model 2000 Microdebrider System Or Xps Powersculpt Regulation Number: 21 CFR 878.5040 Regulation Name: Suction lipoplasty system Regulatory Class: Class II Product Code: OPB

Dear David Timlin:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated January 21, 2000. Specifically, FDA is updating this SE Letter because FDA has created a new product code to better categorize your device technology.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Cindy Chowdhury, OHT4: Office of Surgical and Infection Control Devices, 240-402-6647, Cindy.Chowdhury@fda.hhs.gov.

Sincerely,

# Cindy Chowdhury -S

Cindy Chowdhury, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three stripes representing its wings.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 21 2000

Mr. David Timlin Manager, Regulatory Affairs Xomed, Inc. 6743 Southpoint Drive North Jacksonville, Florida 32216-0980

Re: K992855 Trade Name: XPS/PowerSculpt System Regulatory Class: II Product Code: MUU Dated: December 8, 1999 Received: December 14, 1999

Dear Mr. Timlin:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{2}------------------------------------------------

Page 2 -- Mr. David Timlin

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Hinet Sagon

Ser James E. Dillard III Acting Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

| 510(k) Number (if known): | K992855                  |
|---------------------------|--------------------------|
| Device Name:              | XPS / PowerSculpt System |
| Indications for Use:      |                          |

The XPS/PowerSculpt System with reciprocating adapter and suction cannula is indicated for the removal of soft tissue and fluid from the body during general surgical procedures including suction lipoplasty for aesthetic body contouring.

(Please do not write below this line - continue on another page if needed)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

Or

Over-the-Counter Use

(Optional Format 1-2-96)

Russell Sage

(Division Sion-Division of Ger 510(k) Numb

052

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the word "XOMED" in bold, black letters. To the right of the word is a circle with two arrows pointing towards each other. There is a small "R" in a circle in the upper right corner of the circle with the arrows.

K992855 (P.1 of 2)

510(k) Summary

- 1.0 Date Prepared August 20, 1999
- 2.0 Submitter (Contact) David Timlin Xomed Surgical Products Jacksonville, FL (904) 279-7532

# 3.0 Device Name

| Proprietary Name:    | XPS StraightShot or Model 2000 Microdebrider System<br>or XPS PowerSculpt<br>(The proposed product tradename has not been finalized<br>and may be changed at a later date) |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name(s):      | Electrical surgical debriders, Tissue aspirator, Suction<br>lipoplasty system and cannula                                                                                  |
| Classification Name: | Surgical instrument, AC powered motors and<br>accessories / attachments<br>Suction lipoplasty system                                                                       |

### 5.0 Device Classification

Surgical instrument, AC powered motors and accessories / attachments Procode 87HWE Class II ; 21 CFR 878.4820 Tier 1

Suction lipoplasty system Procode 79MFF Class II ; 21CFR 878.5040 Tier 2

## 6.0 Device Description

The Power Control Unit, handpiece and principle of operation remain essentially the same as described in K984363. The standard blades that are used for tissue debridement and aspiration will be replaced by suction cannulae equivalent in design to predicate suction lipoplasty cannulae. Lastly, a removable adapter, similar to the adapter cleared for rasping in K983025, will be provided to translate the rotary action of the handpiece to a reciprocating action. The

{5}------------------------------------------------

992855 (R.A of 3)

resulting reciprocation will facilitate the movement of the cannula through the adipose tissue.

#### 7.0 Intended Use

The Xomed XPS / PowerSculpt System with reciprocating adapter and suction rno Achea XF Or the removal of soft tissue and fluid from the body during general surgical procedures including suction lipoplasty for aesthetic body contouring.

#### Substantial Equivalence 8.0

The XPS/PowerSculpt System, with expanded indications for suction lipoplasty, is substantially equivalent to the following predicate devices that are currently marketed to dissect and remove adipose tissue:

| Powered Systems<br>MicroAire PAD-100 "Power Aspiration Device"<br>(MicroAire Surgical Instruments) | K981922      |
|----------------------------------------------------------------------------------------------------|--------------|
| NuMED Power Cannula<br>(NuMED or United American Medical)                                          | 510k unknown |
| Aspiration Cannulae<br>Various cannulae (and aspirators)<br>(Wells Johnson Company)                | K832520      |

Like the predicate powered systems, the XPS/PowerSculpt System consists mainly of a console, a handpiece and disposable cannulae. Cuently, an adapter is proposed for use with the current XPS handpieces, to convert the rotary action of the handpiece to a reciprocating action. The end result and performance though is equivalent to the predicate devices ..

As with the predicate devices, the small (approx. 3 mm), but rapid reciprocation of the cannula simulates the manual motion of the surgeon using a standard aspiration cannula. Not only does this reduce the manual effort required by the surgeon, it facilitates the penetration of the cannula through the tissue.

The XPS/PowerSculpt System is substantially equivalent to the currently marketed predicates as it has the same intended use and the same technological characteristics as the predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K992855](https://fda.innolitics.com/device/K992855)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
