XPS STRAIGHTSHOT OR MODEL 2000 MICRODEBRIDER SYSTEM OR XPS POWERSCULPT
K992855 · Xomed, Inc. · QPB · Jan 21, 2000 · General, Plastic Surgery
Device Facts
Record ID
K992855
Device Name
XPS STRAIGHTSHOT OR MODEL 2000 MICRODEBRIDER SYSTEM OR XPS POWERSCULPT
Applicant
Xomed, Inc.
Product Code
QPB · General, Plastic Surgery
Decision Date
Jan 21, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The XPS/PowerSculpt System with reciprocating adapter and suction cannula is indicated for the removal of soft tissue and fluid from the body during general surgical procedures including suction lipoplasty for aesthetic body contouring.
Device Story
System consists of power control unit, handpiece, reciprocating adapter, and suction cannulae. Device transforms rotary action of handpiece into reciprocating motion via adapter to facilitate cannula movement through adipose tissue. Used in surgical settings by surgeons to perform tissue debridement and aspiration. Reciprocating motion simulates manual surgeon movement, reducing manual effort and aiding tissue penetration. Output is physical removal of soft tissue and fluid via suction. Benefits include reduced surgeon fatigue and improved efficiency in adipose tissue dissection.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological characteristics and design similarities to predicate devices.
Technological Characteristics
System comprises an AC-powered control unit, handpiece, removable reciprocating adapter, and suction cannulae. The adapter converts rotary handpiece motion to reciprocating motion. Design is similar to existing powered aspiration devices. No specific material standards or software algorithm details provided.
Indications for Use
Indicated for removal of soft tissue and fluid from the body during general surgical procedures, including suction lipoplasty for aesthetic body contouring.
Regulatory Classification
Identification
A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.
Special Controls
*Classification.* Class II (special controls). Consensus standards and labeling restrictions.
Predicate Devices
MicroAire PAD-100 Power Aspiration Device (K981922)
NuMED Power Cannula (510k unknown)
Wells Johnson Company Aspiration Cannulae (K832520)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Xomed, Inc. David Timlin Manager Regulatory Affairs 6743 Southpoint Dr. North Jacksonville, Florida 32216-0980
June 8, 2021
Re: K992855
Trade/Device Name: Xps Straightshot Or Model 2000 Microdebrider System Or Xps Powersculpt Regulation Number: 21 CFR 878.5040 Regulation Name: Suction lipoplasty system Regulatory Class: Class II Product Code: OPB
Dear David Timlin:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated January 21, 2000. Specifically, FDA is updating this SE Letter because FDA has created a new product code to better categorize your device technology.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Cindy Chowdhury, OHT4: Office of Surgical and Infection Control Devices, 240-402-6647, Cindy.Chowdhury@fda.hhs.gov.
Sincerely,
# Cindy Chowdhury -S
Cindy Chowdhury, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three stripes representing its wings.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 21 2000
Mr. David Timlin Manager, Regulatory Affairs Xomed, Inc. 6743 Southpoint Drive North Jacksonville, Florida 32216-0980
Re: K992855 Trade Name: XPS/PowerSculpt System Regulatory Class: II Product Code: MUU Dated: December 8, 1999 Received: December 14, 1999
Dear Mr. Timlin:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{2}------------------------------------------------
Page 2 -- Mr. David Timlin
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Hinet Sagon
Ser James E. Dillard III Acting Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
| 510(k) Number (if known): | K992855 |
|---------------------------|--------------------------|
| Device Name: | XPS / PowerSculpt System |
| Indications for Use: | |
The XPS/PowerSculpt System with reciprocating adapter and suction cannula is indicated for the removal of soft tissue and fluid from the body during general surgical procedures including suction lipoplasty for aesthetic body contouring.
(Please do not write below this line - continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Or
Over-the-Counter Use
(Optional Format 1-2-96)
Russell Sage
(Division Sion-Division of Ger 510(k) Numb
052
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "XOMED" in bold, black letters. To the right of the word is a circle with two arrows pointing towards each other. There is a small "R" in a circle in the upper right corner of the circle with the arrows.
K992855 (P.1 of 2)
510(k) Summary
- 1.0 Date Prepared August 20, 1999
- 2.0 Submitter (Contact) David Timlin Xomed Surgical Products Jacksonville, FL (904) 279-7532
# 3.0 Device Name
| Proprietary Name: | XPS StraightShot or Model 2000 Microdebrider System<br>or XPS PowerSculpt<br>(The proposed product tradename has not been finalized<br>and may be changed at a later date) |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name(s): | Electrical surgical debriders, Tissue aspirator, Suction<br>lipoplasty system and cannula |
| Classification Name: | Surgical instrument, AC powered motors and<br>accessories / attachments<br>Suction lipoplasty system |
### 5.0 Device Classification
Surgical instrument, AC powered motors and accessories / attachments Procode 87HWE Class II ; 21 CFR 878.4820 Tier 1
Suction lipoplasty system Procode 79MFF Class II ; 21CFR 878.5040 Tier 2
## 6.0 Device Description
The Power Control Unit, handpiece and principle of operation remain essentially the same as described in K984363. The standard blades that are used for tissue debridement and aspiration will be replaced by suction cannulae equivalent in design to predicate suction lipoplasty cannulae. Lastly, a removable adapter, similar to the adapter cleared for rasping in K983025, will be provided to translate the rotary action of the handpiece to a reciprocating action. The
{5}------------------------------------------------
992855 (R.A of 3)
resulting reciprocation will facilitate the movement of the cannula through the adipose tissue.
#### 7.0 Intended Use
The Xomed XPS / PowerSculpt System with reciprocating adapter and suction rno Achea XF Or the removal of soft tissue and fluid from the body during general surgical procedures including suction lipoplasty for aesthetic body contouring.
#### Substantial Equivalence 8.0
The XPS/PowerSculpt System, with expanded indications for suction lipoplasty, is substantially equivalent to the following predicate devices that are currently marketed to dissect and remove adipose tissue:
| Powered Systems<br>MicroAire PAD-100 "Power Aspiration Device"<br>(MicroAire Surgical Instruments) | K981922 |
|----------------------------------------------------------------------------------------------------|--------------|
| NuMED Power Cannula<br>(NuMED or United American Medical) | 510k unknown |
| Aspiration Cannulae<br>Various cannulae (and aspirators)<br>(Wells Johnson Company) | K832520 |
Like the predicate powered systems, the XPS/PowerSculpt System consists mainly of a console, a handpiece and disposable cannulae. Cuently, an adapter is proposed for use with the current XPS handpieces, to convert the rotary action of the handpiece to a reciprocating action. The end result and performance though is equivalent to the predicate devices ..
As with the predicate devices, the small (approx. 3 mm), but rapid reciprocation of the cannula simulates the manual motion of the surgeon using a standard aspiration cannula. Not only does this reduce the manual effort required by the surgeon, it facilitates the penetration of the cannula through the tissue.
The XPS/PowerSculpt System is substantially equivalent to the currently marketed predicates as it has the same intended use and the same technological characteristics as the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.