SIGMA DIAGNOSTICS INFINITY CHOLESTEROL REAGENT, MODELS 401-25, 401-100P, 401-500P, 401-2000P, 402-20, 02-1OOP

K992803 · Sigma Diagnostics, Inc. · CHH · Sep 13, 1999 · Clinical Chemistry

Device Facts

Record IDK992803
Device NameSIGMA DIAGNOSTICS INFINITY CHOLESTEROL REAGENT, MODELS 401-25, 401-100P, 401-500P, 401-2000P, 402-20, 02-1OOP
ApplicantSigma Diagnostics, Inc.
Product CodeCHH · Clinical Chemistry
Decision DateSep 13, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1175
Device ClassClass 1

Intended Use

Sigma Diagnostics INFINITY™ Cholesterol Reagent is intended for the in vitro quantitative, diagnostic determination of cholesterol in human serum or plasma. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.

Device Story

In vitro diagnostic reagent kit for quantitative cholesterol measurement in human serum or plasma. Principle: enzymatic hydrolysis of cholesterol esters by cholesterol esterase to cholesterol and fatty acids; oxidation of free cholesterol by cholesterol oxidase to cholest-4-en-3-one and hydrogen peroxide; peroxidase-catalyzed reaction of hydrogen peroxide with hydroxybenzoic acid and 4-aminoantipyrine to form a measurable quinoneimine dye. Quantitation performed via spectrophotometry at 500-550 nm (or 600/650 nm for bichromatic analyzers). Used in clinical laboratory settings by trained personnel. Output provides cholesterol concentration levels, aiding clinicians in diagnosing hyperlipoproteinaemias and assessing risks related to liver, biliary, intestinal, thyroid, and adrenal functions.

Clinical Evidence

Bench testing only. Correlation study performed using 126 plasma samples comparing INFINITY Cholesterol Reagent to Sigma Procedure No. 352 (K872295), yielding a regression equation of y = 1.00x - 0.3.

Technological Characteristics

Enzymatic colorimetric assay. Reagents: cholesterol esterase, cholesterol oxidase, peroxidase, hydroxybenzoic acid, 4-aminoantipyrine. Detection: spectrophotometric (500-550 nm). Stable liquid formulation. No specific materials of construction or software/connectivity described.

Indications for Use

Indicated for in vitro quantitative diagnostic determination of cholesterol in human serum or plasma for the diagnosis and treatment of disorders involving excess blood cholesterol and lipid/lipoprotein metabolism disorders.

Regulatory Classification

Identification

A cholesterol (total) test system is a device intended to measure cholesterol in plasma and serum. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K992803 ### SEP 1 3 1999 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS INFINITYTM Cholesterol Reagent, Procedure 401 Sigma Diagnostics INFINITY™ Cholesterol Reagent is intended for the in vitro quantitative, diagnostic determination of cholesterol in human serum and plasma. Measurement of serum cholesterol levels can serve as an indicator of liver function, biliary function, intestinal absorption, propensity toward coronary artery disease, thyroid function and adrenal disease. Cholesterol levels are important in the diagnosis and classification of hyperlipoproteinaemias. Stress, age, gender, hormonal balance and pregnancy affect normal cholesterol levels. 12 This reagent is based on the formulation of Allain et al. and the modification of Roeschlau with further improvements to render the reagent stable in solution. - Cholesterol esters are enzymatically hydrolyzed by cholesterol esterase (CE) to cholesterol and free 1. farry acids. Cholesterol Esters CE Cholesterol + Fatty Acids - Free cholesterol, including that originally present, is then oxidized by cholesterol oxidase (CO) to 2. cholest-4-en-3-one and hydrogen peroxide .. Cholesterol + O2 _CO _CO _ Cholest-4-en-3-one + H2O2 - The hydrogen peroxide combines with hydroxybenzoic acid (HBA) and 4-aminoantipyrine (AAP) 3. in the presence of peroxidase (POD) to form a chromophore (quinoneimine dye) which may be quantitated at 500-550 nm. For bichromatic analyzers the blank wavelength should be set to 600 or 650 am. 2H2O2 + HBA + 4AAP POD Quinoneimine Dye + 4H2O The Sigma Diagnostics INFINITY™ Cholesterol Reagent Kit (Procedure No. 401) is substantially equivalent to, and is the same product as the TRACE Scientific Cholesterol Reagent Kit cleared by FDA as K962890. Correlation studies to Sigma Diagnostics Cholesterol Reagent, Procedure No. 352 (K872295) using plasma samples yielded a regression equation of: INFINITY Cholestero] = 1.00 (Cholesterol 352) - 0.3 (N=126) ### References - Searcy R.L. "Diagnostic Biochemistry." McGraw-Hill, New York, NY. 1969. 1. - Ellefson R.D. and Caraway W.T. "Fundamentals of Clinical Chemistry." Ed. Tictz N.W. 1976; p 2. રેણવાડી રહ્યારે તે અને અને અને અને અને અને અને અને અને અને અને અને અને અને અને અને અને અને અને અને અને અને અને અને અને અને પાસની દર્ભાગના પાસની દર્ભાગના દિવસ દિવસ દિવસ દિવસ - Allain C.C., Poon L.S., Chan C.S.G., Richmond W. and Fu P.C. Clin. Chem., 1974; 20:470-475. 3. . - Roeschlau P., Bernt E. and Gruber W.A. Clin. Chem. Clin. Biochem. 1974; 12:226. 4. Contact Person: William R. Gilbert II, Ph.D. Manager, Scientific Affairs Sigma Diagnostics 545 South Ewing St. Louis, MO 63103 314-286-6693 {1}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with three curved lines above them, resembling a bird in flight. SEP 1 3 1999 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 William R. Gilbert, Ph.D. Manager, Scientific Affairs SIGMA DIAGNOSTICS® 545 South Ewing Avenue St. Louis, Missouri 63103 K992803 Re: > Trade Name: INFINITY™ CHOLESTEROL Reagent (Procedure No. 401) Regulatory Class: I reserved Product Code: CHH Dated: August 16, 1999 Received: August 19, 1999 Dear Dr. Gilbert: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {2}------------------------------------------------ ## Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrl/dsma/dsmamain.html". Sincerely yours. Steven Putman Steven I, Gutman, M.D. M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ 510(k) Number (if known): Device Name: Sigma Diagnostics INFINITY™ Cholesterol Reagent # Indications For Use: . Sigma Diagnostics INFINITY™ Cholesterol Reagent is intended for the in vitro quantitative, diagnostic determination of cholesterol in human serum or plasma. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders. Seán Cogger (Division Sign-Off) Division of Clinical Laboratory Devices Division of Clinical Laboratory Devices 510(k) Number . (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) **Prescription Use** (Per 21 CFR 801.109) ✓ ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ OR Over-The-Counter Use _
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%