← Product Code [MNH](/productcode/MNH) · K992739

# CLICK'X (K992739)

_Synthes Spine · MNH · Feb 8, 2000 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K992739

## Device Facts

- **Applicant:** Synthes Spine
- **Product Code:** [MNH](/productcode/MNH.md)
- **Decision Date:** Feb 8, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

When used as a posterior pedicle screw fixation system, the Synthes USS is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudoarthrosis). In addition, the Synthes USS is intended for treatment of severe spondylolisthesis (Grade 3 and 4) of the LS-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine with removal of the implants after the attainment of a solid fusion. The levels of pedicle screw fixation for these patients are L3-S2. When used as a posterior non-pedicle screw fixation system, the Synthes USS is intended for scoliotic, lordotic, or kyphotic deformities (such as scoliosis, Schuermann's disease), degenerative disc disease (back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), and fractures of the posterior thoracolumbar spine. The Click' X components of the Synthes USS, when used as a posterior pedicle screw fixation system, are intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities of the thoracic, lumbar and sacral spine: degenerative spondvlolisthesis with objective evidence of neurologic impairment, fracture, dislocation, spinal tumor and failed previous fusion (pseudoarthrosis). In addition, the Click' X components of the Synthes USS are intended for treatment of severe spondvlolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine with removal of the implants after the attainment of a solid fusion. The levels of pedicle screw fixation for these patients are L3-S2. In addition, when used with 3.5/6.0 mm parallel connectors, the Synthes USS can be linked to the CerviFix System. The Anterior Components of the USS are intended for anterolateral screw and/or staple fixation for the correction of anterolateral for deformities for the spine, lumbar scoliosis, pseudoarthrosis, and fracture or dislocation of the thoracolumbar spine (levels T8-L5).

## Device Story

Synthes USS is a spinal fixation system comprising rods, hooks, screws, and connectors; includes Click'X variable axis components. Designed for posterior pedicle/non-pedicle fixation and anterior anterolateral fixation. Click'X 3D heads attach to pedicle screws in situ; locking caps secure rods and variable axis feature. Used by surgeons in clinical settings to stabilize spinal segments during fusion procedures. Provides mechanical immobilization to facilitate bone healing; implants may be removed after fusion. Benefits patients by correcting deformities and stabilizing unstable spinal segments.

## Clinical Evidence

No clinical data provided; substantial equivalence based on design, material, and intended use comparisons to predicate devices.

## Technological Characteristics

Implants composed of titanium (ASTM F1295, ASTM F67) or stainless steel. System includes rods, hooks, pedicle screws, 3D heads, locking caps, and connectors. Mechanical fixation principle. No software or electronic components.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- Synthes USS ([K990745](/device/K990745.md))

## Reference Devices

- CerviFix System

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

FEB 8 2000

K992739

1812

# Summary of Safety and Effectiveness Information [510(k) Summary]

SYNTHES (U.S.A.) 1690 Russell Road Paoli, PA 19301

(610) 647-9700 Contact: Jonathan M. Gilbert 8/13/99

Device: Synthes USS, including Click'X components

The Synthes USS (including Click'X components) consists of rods, hooks, side-opening screws with collar and nut, variable axis screws with rod and screw connector, collar, locking ring and nut, transconnector systems, open and closed transverse bars, parallel connectors, Schanz screws, clamp with posterior nut and associated manual surgical instruments. The implants are composed of titanium (ASTM F1295, ASTM F67) or various grades of stainless steel.

The Click' X variable axis components of the USS system are composed of titanium (ASTM F1295, ASTM F67) and consist of rods, a transconnector system, pedicle screws, back-opening 3D Heads and Locking Cap. The back-opening 3D Head is clicked on the pedicle screw in situ and attaches to the rod with a locking cap. The locking cap consists of a ring to close the back-opening screw head and an inner set screw to fix the rod and the variable axis feature of the 3D Head.

The components of the predicate spinal system Synthes USS are currently marketed in the U.S. The indications for use, warnings and precautions for the subject components are similar to those cleared via K990745 and conform to labeling requirements set forth by 21 CFR 888.3070and FR Notice published 7/27/98 [Docket No. 95N-0176].

## Posterior Components

When used as a posterior pedicle screw fixation system, the Synthes USS is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudoarthrosis).

In addition, the Synthes USS is intended for treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine with removal of the implants after the attainment of a solid fusion. The levels of pedicle screw fixation for these patients are L3-S2.

When used as a posterior non-pedicle screw fixation system, the Synthes USS is intended for scoliotic, lordotic, or kyphotic deformities (such as scoliosis, Schuermann's disease), degenerative disc disease (back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), and fractures of the posterior thoracolumbar spine.

The Click' X components of the Synthes USS, when used as a posterior pedicle screw fixation system, are intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities of the thoracic, lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, spinal tumor and failed previous fusion (pseudoarthrosis).

{1}------------------------------------------------

K942739

2/12

In addition, the Click' X components of the Synthes USS are intended for treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine with removal of the implants after the attainment of a solid fusion. The levels of pedicle screw fixation for these patients are L3-S2.

In addition, when used with 3.5/6.0 mm parallel connectors, the Synthes USS can be linked to the CerviFix System.

## Anterior Components

The Anterior Components of the USS are intended for anterolateral screw and/or staple fixation for the correction of anterolateral lordotic deformities for the spine, lumbar scoliosis, pseudoarthrosis, and fracture or dislocation of the thoracolumbar spine (levels T8-L5).

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing right, stacked on top of each other, with flowing lines suggesting movement or connection.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

8 2000 FEB

Mr. Jonathan M. Gilbert Senior Regulatory Affairs Associate Synthes Spine 1690 Russell Road Paoli, Pennsylvania 19301

Re: K992739

Trade Name: Click'X (to be used as part of the Universal Spinal System) Regulatory Class: II Product Code: MNH and MNI Dated: November 10, 1999 Received: November 12, 1999

Dear Mr. Gilbert:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general control provisions of the Act. The general control provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

{3}------------------------------------------------

Page 2 - Mr. Jonathan M. Gilbert

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Kinnell Taylor

James E. Dillard III Acting Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications For Use Statement

Page 1 of 1

510(k) Number (if known): NA

Device Name: Synthes USS

Indications for Use:

## Posterior Components

When used as a posterior pedicle screw fixation system, the Synthes USS is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudoarthrosis).

In addition, the Synthes USS is intended for treatment of severe spondylolisthesis (Grade 3 and 4) of the LS-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine with removal of the implants after the attainment of a solid fusion. The levels of pedicle screw fixation for these patients are L3-S2.

When used as a posterior non-pedicle screw fixation system, the Synthes USS is intended for scoliotic, lordotic, or kyphotic deformities (such as scoliosis, Schuermann's disease), degenerative disc disease (back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), and fractures of the posterior thoracolumbar spine.

The Click' X components of the Synthes USS, when used as a posterior pedicle screw fixation system, are intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities of the thoracic, lumbar and sacral spine: degenerative spondvlolisthesis with objective evidence of neurologic impairment, fracture, dislocation, spinal tumor and failed previous fusion (pseudoarthrosis).

In addition, the Click' X components of the Synthes USS are intended for treatment of severe spondvlolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine with removal of the implants after the attainment of a solid fusion. The levels of pedicle screw fixation for these patients are L3-S2.

In addition, when used with 3.5/6.0 mm parallel connectors, the Synthes USS can be linked to the CerviFix System.

### Anterior Components

The Anterior Components of the USS are intended for anterolateral screw and/or staple fixation for the correction of anterolateral for deformities for the spine, lumbar scoliosis, pseudoarthrosis, and fracture or dislocation of the thoracolumbar spine (levels T8-L5).

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

| Prescription Use     | X | OR | Over-the-Counter Use |  |
|----------------------|---|----|----------------------|--|
| (Per 21 CFR 801.109) |   |    |                      |  |

![Signature](signature.jpg)
(Division Sign-Off)

RA-01 8/6/999

|                | General Restorative Devices |
|----------------|-----------------------------|
| 510(k) Number: | K99 2739                    |

9

---

**Source:** [https://fda.innolitics.com/device/K992739](https://fda.innolitics.com/device/K992739)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
