← Product Code [GZJ](/productcode/GZJ) · K992709

# 1818 TENS UNIT (K992709)

_138 Medical Supplies, Inc. · GZJ · Nov 9, 1999 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K992709

## Device Facts

- **Applicant:** 138 Medical Supplies, Inc.
- **Product Code:** [GZJ](/productcode/GZJ.md)
- **Decision Date:** Nov 9, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.5890
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

This TENS device is indicated for the symptomatic relief and management of chronic (long-term) intractable pain and as an adjunctive treatment in the management of post surgical and post traumatic acute pain problems.

## Device Story

1818 TENS Unit transmits electrical pulses through skin to peripheral nerves; blocks pain signals traveling to brain. Device used for symptomatic relief/management of chronic intractable pain and adjunctive treatment for post-surgical/post-traumatic acute pain. Operation involves electrical stimulation; intended for clinical or home use. Healthcare providers or patients use output to manage pain symptoms, potentially reducing reliance on other pain management interventions.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

Transcutaneous Electrical Nerve Stimulation (TENS) device. Complies with AAMI/ANSI NS-4, IEC 60601-1, and IEC 60601-1-2 standards. Electrical pulse generation for peripheral nerve stimulation.

## Regulatory Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

## Predicate Devices

- Diamond Medical TX-3 TENS Unit ([K883188](/device/K883188.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

EXHIBIT #1

# 510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 510(k) number is:________________________________________________________________________________________________________________________________________________

#### 1. Submitter's Identification:

138 Medical Supplies Inc. 313 Butler Street Brooklyn, NY 11217

# Date Summary Prepared:

August 9, 1999 Contact: Ms. Isabelle Liang

#### 2. Name of the Device:

1818 TENS Unit

#### റ്റ് Predicate Device Information:

The 1818 TENS Unit is substantially equivalent to the Diamond Medical TX-3 TENS Unit, K#883188.

#### 4. Device Description:

The device is used to transmit electrical pulses through the skin to the underlying peripheral nerves to help in the blockage of the pain signal traveling to the brain.

#### 5. Intended Use:

This TENS device is indicated for the symptomatic relief and management of chronic (long-term) intractable pain and as an adjunctive treatment in the management of post surgical and post traumatic acute pain problems.

{1}------------------------------------------------

#### 6. Comparison to Predicate Devices:

The 1818 TENS Unit is substantially equivalent to the Diamond Medical TX-3 TENS Unit.

### 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:

All required sections of the AAMI/ANSI NS-4 Standard were met. All required IEC 60601-1 and IEC 60601-1-2 testing was met.

#### 8. Discussion of Clinical Tests Performed:

Not Applicable

#### 9. Conclusions:

The 1818 TENS Unit has the same intended use and technological characteristics as the TX-3 TENS Unit device. Moreover, bench testing contained in this submission and clinical testing supplied demonstrate that technological characteristics do not raise any new questions of safety or effectiveness. Thus, the 1818 TENS Unit is substantially equivalent to the predicate device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three parallel lines that curve and resemble a bird or a symbol representing health and human services.

Public Health Service

NOV - 9 1999

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

138 Medical Supplies, Inc. c/o Ms. Susan D. Goldstein-Falk MDI Consultants, Inc. 55 Northern Boulevard, Suite 200 Great Neck, New York 11021

Re: K992709 Trade Name: 1818 TENS Unit Regulatory Class: II Product Code: GZJ Dated: August 9, 1999 Received: August 11, 1999

Dear Ms. Goldstein-Falk:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Ouality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (OS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

## Page 2 - Ms. Susan D. Goldstein-Falk

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Celia M. Witten, Ph.D., M.D.

Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Page 2 - Ms. Susan D. Goldstein-Falk

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Celia M. Witten, Ph.D., M.D.

Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) NUMBER (IF KNOWN): __K 99 a 7 o 7 138 Medical Supplies Inc. 1818 TENS Unit DEVICE NAME: INDICATIONS FOR USE:

This TENS device is indicated for the symptomatic relief and management of chronic (long-term) intractable pain and as an adjunctive treatment in the management of post surgical and post traumatic acute pain problems.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)

Concurrent of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter-Use (Optional Format 1-2-95)

Pacella

---

**Source:** [https://fda.innolitics.com/device/K992709](https://fda.innolitics.com/device/K992709)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
