THERMALIGHT 2000

K992630 · Jennifermaxx, Inc. · GEX · Sep 10, 1999 · General, Plastic Surgery

Device Facts

Record IDK992630
Device NameTHERMALIGHT 2000
ApplicantJennifermaxx, Inc.
Product CodeGEX · General, Plastic Surgery
Decision DateSep 10, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The THERMALIGHT® 2000 is intended to destroy the dermal papilla of a hair.

Device Story

Thermalight 2000 is a non-invasive epilator for hair removal. Device uses pulsed light energy to destroy hair dermal papilla via thermolysis. Operates by applying light energy to individual hair follicles at skin surface using an optical fiber tip; tip does not penetrate skin. System includes power supply, energy storage, and xenon pulsed light source. Designed for single-hair treatment, similar to conventional needle electrolysis but using light instead of electrical energy. Intended to provide safe, painless hair removal without skin damage. Device is used by clinicians to treat unwanted body hair.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Xenon pulsed light source; optical fiber delivery system; power supply with energy storage; non-invasive tip; standalone unit.

Indications for Use

Indicated for removal of unwanted body hair through the application of pulsed light energy applied to the hair follicle at the surface of the skin (non-invasive photo-epilation).

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ SEP 1 0 1999 ## K992630 Jennifer MAXX 150 Fifth Avenue New York, NY 10010 212 645 9212 ## SUMMARY OF SAFETY AND EFFECTIVENESS #### SPONSOR IDENTIFICATION: A. Jennifer Maxx, Inc. 150 Fifth Avenue, Suite 218 New York City, NY 10011 Telephone: 212 645-9212 #### ESTABLISHMENT REGISTRATION NUMBER: Pending B. #### OFFICIAL CONTACT PERSON C. Norman F. Estrin, Ph. D., RAC President Estrin Consulting Group, Inc. 9109 Copenhaver Drive Potomac , MD 20854 Tel. : (301) 279 -2899 Fax : (301) 294-0126 #### D. DATE OF PREPARATION OF THIS SUMMARY: September 2, 1999 0:00005 #### PROPRIETARY (TRADE) NAME: THERMALIGHT®® 2000 E. - COMMON NAME: Epilator F. - G. CLASSIFICATION NAME AND REFERENCE : Unclassified - PROPOSED REGULATORY CLASS: н. {1}------------------------------------------------ #### DEVICE PRODUCT CODE: I. #### PANEL: General & Plastic Surgery J. #### DESCRIPTION OF DEVICE: K. The Thermalight2000 is an instrument which is designed to painlessly, safely, and effectively remove hair by the application of pulsed light energy to the hair follicle. It operates similar to needle electrolysis machine in that it removes hair, one hair at a time by thermolysis. Using the Thermalight2000, thermolysis is achieved by the application of light energy to the hair growth support tissue instead of the application of electrical energy employed by conventional needle electrolysis units. #### INTENDED USE: L. The THERMALIGHT® 2000 is intended to destroy the dermal papilla of a hair. #### INDICATIONS FOR USE: M. The THERMALIGHT®® 2000 is indicated for removal of unwanted body hair through the application of pulsed light energy applied to the hair follicle at the surface of the skin (non-invasive photo-epilation). #### PREDICATE DEVICE: N. The THERMALIGHT®® 2000 is substantially equivalent to the D' Plume XXII Photocoagulator (Epilation Machine) (K895687) #### 0. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS: Both the THERMALIGHT®® 2000 and the D' Plume XXII Photocoagulator (Epilation Machine) (K895687) have the same intended use and both are indicated for removal of unwanted body hair through the application of pulsed light energy applied to the hair follicle at the surface of the skin (non-invasive photo-epilation). The THERMALIGHT®® 2000 treats one hair at a time just as does conventional needle electrolysis and as does the D' Plume XXII Photocoagulator. Both machines employ a power supply to charge an energy storage device, feeding a xenon pulsed light source to produce a visible light energy pulse. Both machines also employ an optical fiber to conduct the light from the xenon flash lamp to the surface of the skin. The energy delivered to the skin surface is identical and the usage of the two units are identical. The THERMALIGHT®® 2000 incorporates more modern design techniques and newer components which provide a more reliable unit, with more economical, and more consistent performance. #### P. SAFETY: Unlike the machines which treat large areas at a time, the THERMALIGHT®® 2000 is safe and can not generate enough power to damage the skin. The light energy is applied by means of a optical fiber tip which is placed at the entrance to the hair follicle. The tip never enters below the skin. There is no pain, no skin damage, and there is no risk of introducing disease. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The symbol is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement. The text is written in all capital letters. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP 1 0 1999 Jennifer Maxx, Inc. c/o Norman F. Estrin, Ph.D., RAC President Estrin Consulting Group, Inc. 9109 Copenhaver Drive Potomac, Maryland 20854 Re: K992630 Trade Name: Thermalight 2000 Regulatory Class: II Product Code: GEX and KNW Dated: June 14, 1999 Received: June 15, 1999 Dear Dr. Estrin: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ ### Page 2 - Norman F. Estrin, Ph.D., RAC This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours. picasso Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # CONFIDENTIAL Page 1 of 1 ## 510(k) Number: K992630 ## Device Name: THERMALIGHT 2000 Indications for Use: The THERMALIGHT 2000 is indicated for removal of unwanted body hair through the application of pulsed light energy applied to the hair follicle at the surface of the skin (noninvasive photo-epilation). ## (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-the-Counter Use (Optional Format 1-2-96) (Division)Sign-Off Division of General Restorative Devices 510(k) Numbe
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...