← Product Code [FTM](/productcode/FTM) · K992537

# PERI-STRIPS-SLEEVE, PERI-STRIPS-STRIP AND PERI-STRIPS DRY (NON-NAOH-TREATED AND NAOH-TREATED (K992537)

_Bio-Vascular, Inc. · FTM · Feb 22, 2000 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K992537

## Device Facts

- **Applicant:** Bio-Vascular, Inc.
- **Product Code:** [FTM](/productcode/FTM.md)
- **Decision Date:** Feb 22, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.3300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Peri-Strips is intended for use as a prosthesis for the surgical repair of soft tissue deficiencies using surgical staplers. Peri-Strips can be used to reinforce staple lines during lung resections including pneumonectorny, pneumoreduction, pneumectorny, lobectomies, segmentectomies (segmental resections), wedge resections, blebectomies, bronchial resections, and other lung incisions and excisions of lung and bronchus. It can also be used for the reinforcement of the gastric staple line during the bariatric surgical procedures of gastric bypass and gastric banding. Other soft tissue deficiencies amenable to repair with Peri-Strips include defects of the abdominal and thoracic wall, gastric banding, muscle flap reinforcement, rectal and vaginal prolapse, reconstruction of the pelvic floor, and hernias (including diaphragmatic, femoral, incisional, inguinal, lumbar, paracolostomy, scrotal, and umbilical hernias). Peri-Strips may be used with anastomotic staplers (when tissue division is desired) or with non-anastomotic staplers (when no tissue division is desired) according to the above indications.

## Device Story

Peri-Strips and Peri-Strips Dry are surgical mesh patches derived from glutaraldehyde-crosslinked bovine pericardium. Configurations include tubular sleeves (for stapler use) and flat strips. Peri-Strips Dry undergoes vacuum dehydration and uses a hydrogel (PSD Gel) for stapler attachment. Devices are used by surgeons in OR settings to reinforce staple lines during lung and bariatric surgeries or to repair soft tissue deficiencies. The device acts as a mechanical reinforcement layer, sutured or affixed to surgical staplers to provide structural support to tissue during and after stapling. This reduces potential for staple line leaks or tissue tearing, benefiting patients by improving surgical outcomes in complex resections and bariatric procedures.

## Clinical Evidence

Bench and animal testing only. Testing demonstrated no difference in safety or efficacy between NaOH-treated and non-NaOH-treated bovine pericardium tissue. No clinical trial data presented.

## Technological Characteristics

Material: Glutaraldehyde-crosslinked bovine pericardium. Configurations: Sleeve (tubular) or strip (flat). Processing: Optional 1M NaOH treatment, rinsing, and neutralization; optional vacuum dehydration for 'Dry' version. Accessory: Hydrogel (PSD Gel) for Dry version. Application: Used with surgical staplers.

## Regulatory Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

## Predicate Devices

- Peri-Strips - Sleeve ([K983162](/device/K983162.md))
- Peri-Strips - Strip ([K983162](/device/K983162.md))
- Peri-Strips Dry ([K971048](/device/K971048.md))
- Peri-Strips Dry ([K983162](/device/K983162.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a logo with the letters 'bvi' inside of a circle. The circle is surrounded by a shape that looks like two overlapping circles. The logo is black and white and has a simple design.

FEB 2 2 2000

# 2637

Bio-Vascular, Inc.

| 510(k) Summary of Safety and Effectiveness |                                                                                                                                                                                                                                                   |
|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant Name and Address:                | Bio-Vascular, Inc.<br>2575 University Avenue, Suite 180<br>Saint Paul, Minnesota 55114 USA                                                                                                                                                        |
| Contact:                                   | Mary Kay Kessinger Sobcinski, RN, MHA<br>Manager of Clinical Affairs<br>Phone: 651-603-3703<br>FAX: 651-642-9018<br>e-mail: m.sobcinski@biovascular.com                                                                                           |
| Alternate Contact:                         | Mary L. Frick<br>Director of Regulatory Affairs, Clinical Affairs, and Quality<br>Assurance<br>Phone: 651-603-3803<br>FAX: 651-642-9018<br>e-mail: m.frick@biovascular.com                                                                        |
| Date Prepared:                             | February 18, 2000                                                                                                                                                                                                                                 |
| Trade & Common Names:                      | Peri-Strips® - Sleeve, Peri-Strips® - Strip and Peri-Strips<br>Dry™ (non-NaOH-treated); Peri-Strips® - Sleeve and Peri-<br>Strips® - Strip and Peri-Strips Dry™ (NaOH-treated)                                                                    |
| Device Classification Name:                | mesh, surgical, polymeric                                                                                                                                                                                                                         |
| Substantial Equivalence:                   | • Peri-Strips® - Sleeve (K983162), Peri-Strips® - Strip<br>(K983162) and Peri-Strips Dry™ (K971048) [non-NaOH-<br>treated];<br>• Peri-Strips® - Sleeve (K983162), Peri-Strips® - Strip<br>(K983162) and Peri-Strips Dry™ (K983162) [NaOH-treated] |

## Device Descriptions:

Peri-Strips (sleeve and strip) and Peri-Strips Dry are surgical mesh patches derived from bovine pericardium that is cross-linked with glutaraldehyde.

Peri-Strips comes in two configurations: Peri-Strips-sleeve and Peri-Strips-strip. Each unit of Peri-Strips-sleeve is composed of two strips of bovine pericardial tissue. After the final manufacturing step, each strip of pericardium is sewn to a polyolefin strip with polypropylene suture to create a tubular sleeve. The tubular configuration facilitates usage of the product with surgical staplers. The polyolefin strip and sutures are removed after the Peri-Strips-sleeve is secured in place. Peri-Strips-strip is a flat piece of bovine pericardium that is sutured to the stapler by the surgical team.

For Peri-Strips Dry (PSD), the bovine pericardial patch is manufactured in the form of a strip and undergoes an additional manufacturing step of dehydration by vacuum drying. The strip is affixed to the surgical stapler with the aid of a hydrogel (PSD Gel) supplied with the product.

Product may be treated with 1 molar sodium hydroxide (1M NaOH) for 60-75 minutes at 20-25°C, rinsed with deionized (DI) water, and neutralized with citrate solution, followed by a final DI water rinse. Bio-Vascular believes that product subjected to the NaOH treatment performs in a substantially 

{1}------------------------------------------------

equivalent manner to non-NaOH-treated product, and that there is no difference in safety and efficacy between treated and non-treated tissue.

## Statement of Intended Use:

The indications for Peri-Strips (sleeve and strip) and Peri-Strips Dry are being expanded to include reinforcement of the gastric staple line during the bariatric surgical procedures of gastric bypass and gastric banding.

#### Summary/Comparison of Technological Characteristics:

These products are the same as the predicate devices except for the added indication of bariatric surgery.

The devices are comprised of bovine pericardium crosslinked with glutaraldehyde. They have an extensive clinical history of safe and effective use when used as surgical patches. Bench testing and animal testing show no difference in safety and efficacy between NaOH-treated and non-NaOHtreated tissue. Specifications of r NaOH-treated and non-NaOH-treated tissue are identical.

#### Safety and Effectiveness Conclusions:

These devices are substantially equivalent to the predicate devices with regard to safety and efficacy. No new questions of safety and efficacy are raised by the bariatric sugery indication.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white circular seal for the Department of Health & Human Services - USA. The seal features the department's name in a circular pattern around the edge. In the center of the seal is a stylized symbol that resembles an eagle or a bird in flight, composed of three curved lines.

FEB 2 2 2000

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Mary Kay Kessinger Sobcinski, RN, MHA Manager of Clinical Affairs Bio-Vascular, Inc. 2575 University Avenue, Suite 180 Saint Paul, Minnesota 55114

Re: K992537 Trade Name: Peri-Strips® – Sleeve

> Peri-Strips® - Strip Peri-Strips® - Dry Regulatory Class: II

Product Code: FTM Dated: November 23, 1999 Received: November 24, 1999

Dear Ms. Kessinger Sobcinski:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

# Page 2 - Ms. Mary Kay Kessinger Sobcinski, RN. MHA

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

James F. Dillard III

James E. Dillard III Acting Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if Known):

Image /page/4/Picture/2 description: The image shows a sequence of numbers and letters in a bold, stylized font. The sequence appears to be 'k992537'. The characters are black against a white background, creating a high contrast. There is also a black bar on the left side of the image.

Device Name:

Peri-Strips Dry® Staple Line Reinforcement (NaOH-treated product)

#### Indications for Use:

Peri-Strips Dry is intended to reinforce staple lines during lung and bronchus resections including; pneumonectomy, pneumoreduction, pneumectorny, segmental resections), wedge resections, blebectomies, lobectomies, bullectomies, bronchial resections and other lung incisions and excisions of lung and bronchus. It can also be used for the reinforcement of the gastric staple line during the bariatric surgical procedures of gastric bypass and gastric banding.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
Per 21 CFR 801.109

OR

Over-The-Counter Use

(Division Sign-Off)
Division of General Restorative Devices
510(k) Number. K992537

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the text "Indications for Use" at the top. Below that, the text "510(k) Number (if known):" is on the left, followed by a blacked-out section. To the right of the blacked-out section is the number "K992537" in large, bold font. The number appears to be handwritten.

Device Name:

Peri-Strips Dry® Staple Line Reinforcement (non-NaOH-treated product)

## Indications for Use:

Peri-Strips Dry is intended to reinforce staple lines during lung and bronchus resections including: pneumonectorny, pneumoreduction, pneumectornies, (segmental resections), wedge resections, blebectomies, lobectomies, bullectomies, bronchial resections and other lung incisions and excisions of lung and bronchus. It can also be used for the reinforcement of the gastric staple line during the bariatric surgical procedures of gastric bypass and gastric banding.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use

OR

Over-The-Counter Use

pcalles

General Resto 510(k) Numbe

{6}------------------------------------------------

Image /page/6/Picture/0 description: This image shows the indications for use of a medical device. The 510(k) number, if known, is K992537. The device name is Bori String Staple Line Reinforcement.

Device Name;

Peri-Strips® Staple Line Reinforcement - Strip Configuration (NaOH-treated product)

## Indications for Use:

Peri-Strips is intended for use as a prosthesis for the surgical repair of soft tissue deficiencies using surgical staplers. Peri-Strips can be used to reinforce staple lines during lung resections including pneumonectorny, pneumoreduction, pneumectorny, lobectomies, segmentectomies (segmental resections), wedge resections, bullectornies, blebectomias, bronchial resections, and other lung incisions and excisions of lung and bronchus. It can also be used for the reinforcement of the gastric staple line during the bariatric surgical procedures of gastric bypass and gastric banding. Other soft tissue deficiencies amenable to repair with Peri-Strips include defects of the abdominal and thoracic wall, gastric banding, muscle flap reinforcement, rectal and vaginal prolapse, reconstruction of the pelvic floor, and hernias (including diaphragmatic, femoral, incisional, inguinal, lumbar, paracolostomy, scrotal, and umbilical hernias). Peri-Strips may be used with anastomotic staplers (when tissue division is desired) or with non-anastomotic staplers (when no tissue division is desired) according to the above indications.

# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
Per 21 CFR 801.109

OR

Over-The-Counter Use

bcosell

(Division Sign-Off)
Division of General Restorative Devices K992537
510(k) Number

{7}------------------------------------------------

# Indications for Use

510(k) Number (if Known):

Device Name:

Peri-Strips® Staple Line Reinforcement - Strip Configuration (non-NaOHtreated product)

#### Indications for Use:

Peri-Strips is intended for use as a prosthesis for the surgical repair of soft tissue deficiencies using surgical staplers. Peri-Strips can be used to reinforce staple lines during lung resections including pneumonectorny, pneumoreduction, pneumectomy, lobectomies, segmentectomies (segmental resections), wedge resections, bullectornies, bronchial resections, and other lung incisions and excisions of lung and bronchus. It can also be used for the reinforcement of the gastric staple line during the bariatric surgical procedures of gastric bypass and gastric banding. Other soft tissue deficiencies amenable to repair with Peri-Strips include defects of the abdominal and thoracic wall, gastric banding, muscle flap reinforcement, rectal and vaginal prolapse, reconstruction of the pelvic floor, and hernias (including diaphragmatic, femoral, incisional, inguinal, lumbar, paracolostomy, scrotal, and umbilical hemias). Peri-Strips may be used with anastomotic staplers (when tissue division is desired) or with non-anastomotic staplers (when no tissue division is desired) according to the above indications.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
Per 21 CFR 801.109

OR Over-The-Counter Use*_*_

of General Restorative De

{8}------------------------------------------------

510(k) Number (if known):

Image /page/8/Picture/1 description: The image shows the words "Indications for Use" followed by the number K992537. The words "Indications for Use" are in a smaller font than the number. The number is written in a bold, handwritten font. The image is black and white.

Device Name:

Peri-Strips® Staple Line Reinforcement - Sleeve Configuration (non-NaOHtreated product)

## Indications for Use:

Peri-Strips is intended for use as a prosthesis for the surgical repair of soft tissue deficiencies using surgical staplers. Peri-Strips can be used to reinforce staple lines during lung resections including pneumonectorny, pneumoreduction, pneumectorny, lobectomies, segmental resections), wedge resections, bullectomies, bronchlal resections, and other lung incisions and excisions of lung and bronchus. It can also be used for the reinforcement of the gastric staple line during the barlatric surgical procedures of gastric bypass and gastric banding. Other soft tissue deficiencies amenable to repair with Peri-Strips include defects of the abdominal and thoracic wall, gastric banding, muscle flap reinforcement, rectal and vaginal prolapse, reconstruction of the pelvic floor, and hernias (including diaphragmatic, femoral, incuinal, lumbar, paracolostomy, scrotal, and umbilical hernias). Peri-Strips may be used with anastomotic staplers (when tissue division is desired) or with non-anastomotic staplers (when no tissue division is desired) according to the above indications,

# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
Per 21 CFR 801.109

![alt_text](data:image/png;base64,iVBORw0KGgoAAAANSUhEUgAAABQAAAAUCAYAAACNbyblAAAAAXNSR0IArs4c6QAAAERlWElmTU0AKgAAAAgAAYdpAAQAAAABAAAAGgAAAAAAA6ABAAMAAAABAAEAAKACAAQAAAABAAAAUKADAAQAAAABAAAAUgAAAAA5r8rAAAAJUlEQVQ4EWNgGAXDEwYkYJgYmYGBYQYGBiRgmBgZgYGBiRgmBgYAAH0gA/40+4AAAAASUVORK5CYII=)

OR

Over-The-Counter Use

casillas

(Division Sign-Off)
Division of General Restorative Devices K992537
510(k) Number

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the words "Indications for Use" in the upper left corner. To the right of the words is a large, bold, handwritten number "K99 2537". There is also a black line under the words "Indications for Use".

510(k) Number (if known):

Device Name:

Peri-Strips® Staple Line Reinforcement - Sleeve Configuration (NaOH-treated product)

#### Indications for Use:

Peri-Strips is intended for use as a prosthesis for the surgical repair of soft tissue deficiencies using surgical staplers. Peri-Strips can be used to reinforce staple lines during lung resections including pneumonectorny, pneumoreduction, pneumectorny, lobectomies, segmentectomies (segmental resections), wedge resections, blebectomies, bronchial resections, and other lung incisions and excisions of lung and bronchus. It can also be used for the reinforcement of the gastric staple Ilne during the bariatric surgical procedures of gastric bypass and gastric banding. Other soft tissue deficiencies amenable to repair with Peri-Strips include defects of the abdominal and thoracic wall, gastric banding, muscle flap reinforcement, rectal and vaginal prolapse, reconstruction of the pelvic floor, and hernias (including diaphragmatic, femoral, incisional, lumbar, paracolostomy, scrotal, and umbilical hernias). Peri-Strips may be used with anastomotic staplers (when tissue division is desired) or with non-anastomotic staplers (when no tissue division is desired) according to the above indications.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

**Prescription Use**
Per 21 CFR 801.109

OR

Over-The-Counter Use

castellanos

eral Restorative Devices

---

**Source:** [https://fda.innolitics.com/device/K992537](https://fda.innolitics.com/device/K992537)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
