← Product Code [HSD](/productcode/HSD) · K992525

# ACUMED MODULAR SHOULDER SYSTEM (K992525)

_Acu Med, Inc. · HSD · Feb 4, 2000 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K992525

## Device Facts

- **Applicant:** Acu Med, Inc.
- **Product Code:** [HSD](/productcode/HSD.md)
- **Decision Date:** Feb 4, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3690
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Addresses proximal fractures when used with interlocking screws and arthritis of the shoulder joint.

## Device Story

Acumed Modular Shoulder System is a semi-constrained modular humeral prosthesis. Components include cobalt chrome head, titanium body, and stem; modular design allows mixing/matching to replicate patient anatomy. Features include suture holes for tuberosity fixation and Morse taper stem connection. Interlocking screws provide rotational stability; cannulated stems facilitate placement over guide wires. Used by orthopedic surgeons in clinical settings for fracture management or arthritis. Targeting instrumentation assists screw placement. Device provides anatomic posterior offset restoration. Implants are provided sterile (gamma radiation) or non-sterile (steam sterilizable).

## Clinical Evidence

Bench testing only. Data provided includes verification of sterility (AAMI Method 1) and mechanical testing of the Morse taper locking strength.

## Technological Characteristics

Modular humeral prosthesis; materials: cobalt chrome (head), titanium (body/stem). Features Morse taper connection, suture holes, and interlocking screw fixation. Sterilization: gamma radiation (sterile version) or steam sterilization (instrumentation/non-sterile implants).

## Regulatory Identification

A shoulder joint humeral (hemi-shoulder) metallic uncemented prosthesis is a device made of alloys, such as cobalt-chromium-molybdenum. It has an intramedullary stem and is intended to be implanted to replace the articular surface of the proximal end of the humerus and to be fixed without bone cement (§ 888.3027). This device is not intended for biological fixation.

## Predicate Devices

- Atlas Modular Shoulder ([K961260](/device/K961260.md))
- Select Shoulder System ([K962224](/device/K962224.md))

## Reference Devices

- Cortical Bone Screws ([K942340](/device/K942340.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## K992525

## FEB 4 2000

## Appendix F - 510(k) Summary

This summary regarding 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.93.

| Classification Name:             | Prosthetic, shoulder, humeral (hemi-shoulder), uncemented, metallic     |
|----------------------------------|-------------------------------------------------------------------------|
| Common Name:                     | Modular shoulder system                                                 |
| Proprietary Name:                | Acumed Modular Shoulder System                                          |
| Proposed Regulatory Class:       | Class II                                                                |
| Device Product Code:             | HSD                                                                     |
| Manufacturing Facility:          | Acumed, Inc.                                                            |
|                                  | 10950 SW 5th Street, Suite 170                                          |
|                                  | Beaverton, OR 97005 U.S.A.                                              |
| Establishment Registration No .: | 3025141                                                                 |
| Contact:                         | Shari Jeffers                                                           |
| Labeling/Promotional Materials:  | See Appendix D                                                          |
| Substantial Equivalence:         | This device is similar in size, material, and intended use to the Atlas |
|                                  | Modular Shoulder (K961260) and the Select Shoulder System               |
|                                  | (K962224). This device is most similar in design to the Atlas Modular   |
|                                  | shoulder in that both devices consist of a head, body, and stem         |
|                                  | components, have holes for suturing, and have a Morse tapered stem      |
|                                  | Literature on the Atlas and Select Shoulder Systems are included in     |
|                                  | Appendix E.                                                             |

The Acumed Modular Shoulder System is a semi-constrained modular component designed to address proximal fractures when used with interlocking screws and arthritis of the shoulder joint to with or without bone cement. It incorporates a cobalt chrome modular head, a titanium body, and a selles of the components may be mixed and matched in order to best duplicate the patient's anktomy. Several strategic holes for suturing are in the body the surgeon to tie down the greater- and lesser-tuberosities, which are frequently displaced during these types of fractures. Interlocking screws identical to those cleared by K942340 Cortical Boxe Screws are intended to provide rotational interlocking with the stems are cannulated and are intended to facilitate placement of the component over a previously inserted gtride wire. The component sizing and the geometrical placement of the head and stem are intended to allow the surgeon to duplicate the anatomic posterior offset in the humeral head. Targeting instrumentation is provided to allow accurate of the screws through the component.

The Acumed Modular Shoulder System is provided both non-sterile. The sterile version is packaged in inner and outer PETO blisters with Tyvek lids. Sterility is achieved by a minimum of 2.5 megarads gamma radiation. Verification of sterlity is performed using the AAMI - Method 1. Sterllity assurance level is 10 . We make no claims as to the pyrogenicity of this product. Instrumentation is provided nonsterile in a tray. On file at Acumed is data which shows that the instrumentation and implant can be successfilly steam sterilized under specific process parameters which will obtain a resulting SAL of 104. Information regarding has been provided. Test data provided concludes that the taper selected for the Acumed Modular Shoulder System will provide adequate locking strength for this application.

Predicate devices that are substantially equivalent to the Acumed Modular System are the Atlas Modular Shoulder and the Intermedics Select Shoulder. All the devices mentioned above are manufactured from similar materials and have the same intended use. The design of the Acumed Shoulder is most similar to the Atlas Shoulder in that both devices consist of a head, body, and stem components, have holes for suturing, and have a Morse rapered stem. The Acumed Shoulder and the Atlas Shoulder have similar surgical techniques. Based on the similarities between the Acumed Modular System and the predicate devices studied, the safety and effectiveness of the Acumed Modular System is expected to be similar to the prodicate devices mentioned above.

10950 SW Sth Street, Suite 170, Beaverton, OR 97005 U.S.A. . (503) 627;9957 . ● Fax: (503) 520-9618 Page 11 © 1999, Acumed, Inc.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a circular design with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized human figure or bird in flight, composed of three curved lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 4 2000

Ms. Shari Jeffers Regulatory Affairs Manager Acumed. Inc. 10950 SW 5th Street, Suite 170 Beaverton, Oregon 97005

Re: K992525/S1

Trade Name: Acumed Modular Shoulder System Regulatory Class: II Product Code: HSD Dated: November 5, 1999 Received: November 8, 1999

Dear Ms. Jeffers:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{2}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely vours.

Neil R. Ogden
James E. Dillard III

James E. Di Acting Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

S10(k) Number (if known):_K992 S Z S

Device Name: ___ Acumed Shoulder Implany

Indications For Use:

Addresses proximal fractures when used with interlocking screws and arthritis of the shoulder joint.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

| Concurrence of CDRH, Office of Device Evaluation (ODE)         |             |
|----------------------------------------------------------------|-------------|
|                                                                | NPO for JZD |
| (Division Sign-Off)<br>Division of General Restorative Devices |             |
| 510(k) Number                                                  | K992525     |

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use_

. •

·

(Optional Format 1-2-96)

.

---

**Source:** [https://fda.innolitics.com/device/K992525](https://fda.innolitics.com/device/K992525)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
