EXPRESSEYE

K992473 · Optomotor Laboratories · GWN · Aug 12, 1999 · Neurology

Device Facts

Record IDK992473
Device NameEXPRESSEYE
ApplicantOptomotor Laboratories
Product CodeGWN · Neurology
Decision DateAug 12, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.1460
Device ClassClass 2
AttributesPediatric

Indications for Use

ExpressEye is used to measure the saccadic movements of the eye and classify a subject's performance in different tasks as normal or abnormal. Subjects are patients above the age of 6 years with suspicion of problems in dynamic vision and/or optomotor control.

Device Story

ExpressEye measures horizontal eye movements to assess optomotor control. The device includes a headband with infrared light-emitting/receiving elements and integrated laser visual stimuli (Class 2). A handheld processor controls stimuli and analyzes eye movement signals; data can be transferred to a PC. The system tracks corneal reflection via infrared light to monitor saccadic movements, fixation quality, and reaction times. Used in clinical settings to classify patient performance as normal or abnormal. Output provides objective analysis of eye movement traces, which may inform training procedures. The device benefits patients by identifying dynamic vision deficits through quantitative assessment of reflexive and voluntary eye control.

Clinical Evidence

Bench testing only. Testing confirmed compliance with 21 CFR Part 1040 for Class 2 lasers (0.2 mW output), adherence to ACGIH spectral radiance limits for infrared light, and FCC Class A electromagnetic compatibility standards. Biocompatibility of headband materials is confirmed per US/European regulations.

Technological Characteristics

System includes plastic headband, infrared LEDs (940 nm), and Class 2 mini-lasers (0.2 mW). Sensing via infrared corneal reflection. Handheld processor for stimulus control and signal analysis. Connectivity via optional PC data transfer. Electromagnetic shielding included. Biocompatible headband materials.

Indications for Use

Indicated for patients >6 years old with suspected dynamic vision or optomotor control problems. Used to measure horizontal saccadic eye movements in response to visual stimuli to assess fixation, reflexive, and voluntary eye movement control.

Regulatory Classification

Identification

A nystagmograph is a device used to measure, record, or visually display the involuntary movements (nystagmus) of the eyeball.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ 6979423 # AUG 12 1999 7-15-1999 # 510(k) Summary Submitted by: Optomotor Laboratories BBL Fischer Tivolistr. 11 D - 79104 Freiburg Germany +49 761 2922600 Phone: Phone: +49 761 2922699 Fax: #### Contact: Prof. Dr. rer. nat. Burkhart Fischer Trade Name: ExpressEye # Common Name: Eve Movement Recorder #### Classification: Class II, 21 CFR 882.1460, Nystagmograph #### Device Description: ExpressEye measures horizontal eye movements and analyses the fixation quality, direction and reaction time of fast eye movements. The device consists of a head band, which carries light sources for visual stimuli, an amplifier, and infrared light emmitting/receiving elements. A hand held processor controls the stimuli and analyses the eye movement signals. Optional the user can transfer the data to a personal computer. #### Intended Use: ExpressEye is used to measure the saccadic movements of the eye and classify a subject's performance in different tasks as normal or abnormal. Subjects are patients above the age of 6 years with suspicion of problems in dynamic vision and/or optomotor control. ## Substantial Equivalence: House Infrared/Video ExpressEye is substantially equivalent to the The Electronystagmograph System by Eye Dynamics, Inc. Both devices measure the movements of the eye by infrared light corneal reflection. Both instruments allow to display fast and slow eye movements. While ExpressEye analyses fast eye movements (saccades) the Nystagmograph permits viewing of the eyes and recording on video tape. Most technical details such as infrared illumination are the same. ExpressEye additionally generates visual target stimuli and provides appropriate analysis. {2}------------------------------------------------ 5992423 5 Summary of comparison with the predicate Device: | Item | ExpressEye | House IR/Video | |------------------|----------------------------------------------------------|----------------------------------------------| | Intended use | (Identical, with additional data analysis in ExpressEye) | | | Sensor | video camera | phototransistors | | Light source | infrared LEDs<br>Peak Wavelength<br>= 940 nm | infrared LEDs<br>Peak Wavelength<br>= 950 nm | | Headmounting | plastic head band | eye goggles | | Target projector | mini Lasers with<br>low energy (Class II) | not present | Summary and conclusion of technical tests: ## Laser: The device meets the performance standard 21 CFR Part 1040 (light emitting products) part 10 (laser products). The lasers are classified as Laser Class 2. The use of the lasers to project visual targets on a surface in front of the test person is save, because the power of the lasers is only 0.2 milliwatts, 5 times less than permitted for a laser class 2 device. ## Infrared light: the infrared light used by the device has a spectral radiance inside the Threshold Limit Value recommended by the American Conference of Governmental Industrial Hygienists. ## Electromagnetic Interference: The electromagnetic compatibility was measured by a laboratory approved by the U.S. FCC. They certified that the radiation is below the level allowed for class A. Therefore the following text appears in the device manual: Warning - This is a Class A product. In a domestic environment this product may cause radio interference in which case the user may be required to take adequate measures. The disturbance of the device by outside radiation found in the measurement is eliminated by added shielding. #### Biocompatibility: The materials of the head band are declared biocompatible and comformant to European and US regulations by the manufacturer of the head band. The additional acquirement of variables from the eye movement traces does not affect the safety because it results in recommendations of training procedures, not in any kind of medication. {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines beneath them, resembling water or fabric. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 AUG 1:2 1999 Dr. Burkhart Fischer Optomotor Laboratories Tivolistr. 1 1 D-79104 Freiburg Germany Re: K992473 Trade Name: ExpressEye Regulatory Class: II Product Code: GWN Dated: July 15, 1999 Received: July 26, 1999 Dear Dr. Fischer: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4}------------------------------------------------ # Page 2 – Dr. Burkhart Fischer This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, bastula Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # 1 Indications for Use ExpressEye can be used in any case of problems in the optomotor domain. The measured horizontal eye movements of a person made in response to generated stimuli are analysed to nontonial cyc inovential or a person's fixation and reflexive and voluntary eye movement control. (Division Si Division of Ger 1992473 510(k) Number **Prescription Use** **(Per 21 CFR 801.109)**
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...