# AXYA MEDICAL INC, BONE ANCHOR SYSTEM AND KIT- (BAK) (K992405)

_Axya Medical, Inc. · GAW · Sep 22, 1999 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K992405

## Device Facts

- **Applicant:** Axya Medical, Inc.
- **Product Code:** [GAW](/productcode/GAW.md)
- **Decision Date:** Sep 22, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.5010
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Bone Anchor System and Kit is indicated for securing soft tissue structures to bone anchors in surgical procedures where bone anchors are indicated. Examples of these procedures include: Shoulder procedures (e.g. Rotator cuff repair) Knee procedures (e.g. Patellar ligament / tendon avulsion repair) Ankle procedures (e.g. Achilles tendon reconstruction and repair) Foot procedures (e.g. Hallux valgus reconstruction) Elbow procedures (e.g. Biceps tendon reattachment) Bladder neck suspension (female urinary incontinence due to urethral hypermobility) The BAK is indicated for use in both traditional open surgery and endoscopic surgical proceduras.

## Device Story

System consists of electronic control module, reusable welding activator, and disposable suture welding sleeve kit with bone anchors and USP polypropylene monofilament suture. Used in OR by surgeons for soft tissue-to-bone attachment. Principle of operation: ultrasonic energy generates friction to melt and fuse polymer suture loops, replacing manual knot-tying. Welding sleeve design prevents heated suture contact with patient tissue; no ultrasonic generator components contact patient. Output is a welded suture loop with knot strength exceeding USP requirements. Benefits include secure fixation equivalent to manual knots without thermal injury risk.

## Clinical Evidence

No human clinical trials. Evidence based on in vitro bench testing and animal studies. Bench testing: cyclic fatigue, elongation, and knot strength testing of welded polypropylene sutures vs. USP requirements. Animal studies: New Zealand rabbit model (dorsal skin and bowel) compared welded sutures to manually knotted sutures; histopathology showed no significant safety or efficacy differences.

## Technological Characteristics

System includes electronic control module and reusable ultrasonic welding activator. Disposable kit contains bone anchors and USP polypropylene monofilament suture. Ultrasonic energy source (same as predicate ASLS/SWK). Sterilization via gamma irradiation (2.5 Mrads) or ethylene oxide gas. Materials have established history of medical use. No flux or welding rod used; weld formed by friction-induced polymer melting.

## Regulatory Identification

Nonabsorbable polypropylene surgical suture is a monofilament, nonabsorbable, sterile, flexible thread prepared from long-chain polyolefin polymer known as polypropylene and is indicated for use in soft tissue approximation. The polypropylene surgical suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be undyed or dyed with an FDA approved color additive; and the suture may be provided with or without a standard needle attached.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

## Predicate Devices

- Axya Suturing and Ligating System (ASLS) ([K980988](/device/K980988.md))
- Axya Suture Welding System and Kit (SWK) ([K983108](/device/K983108.md))
- Endo-Judge disposable suture placement system ([K932591](/device/K932591.md))

## Reference Devices

- UltraCision Harmonic Scalpel ([K895252](/device/K895252.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## SUMMARY OF SAFETY AND EFFECTIVENESS

## General Company Information

- Axya Medical, Inc. Name:
- 100 Cummings Center Address: Suite 444C Beverly, MA 01915

(978) 232 - 9997 Telephone: (978) 232 - 9998 Fax:

## General Device Information

- Product Name: Bone Anchor System and Kit (BAK)
- Classification: Accessory to an Endoscope - Class II

## Predicate Devices

Axya Suturing and Ligating System (ASLS) [510(k) K980988]

Axya Suture Welding System and Kit (SWK) [510(k) K983108]

## Description

The Axya Bone Anchor System consists of an electronic control module and a reusable welding activator with an integral connector cable. The Kit (BAK) consists of a disposable suture welding sleeve packaged together with appropriate and currently approved bone anchors and USP polypropylene monofilament suture. The Kit is intended for use with the reusable system components in order to complete the suture welding process for the attachment of soft tissues to bone anchors in surgical procedures where bone anchors are indicated

{1}------------------------------------------------

K992405
2 of 3

## Intended Use

The Bone Anchor System and Kit is indicated for securing soft tissue structures to bone anchors in surgical procedures where bone anchors are indicated. Examples of these procedures include:

| Shoulder procedures     | (e.g. Rotator cuff repair)                                  |
|-------------------------|-------------------------------------------------------------|
| Knee procedures         | (e.g. Patellar ligament / tendon avulsion repair)           |
| Ankle procedures        | (e.g. Achilles tendon reconstruction and repair)            |
| Foot procedures         | (e.g. Hallux valgus reconstruction)                         |
| Elbow procedures        | (e.g. Biceps tendon reattachment)                           |
| Bladder neck suspension | (female urinary incontinence due to urethral hypermobility) |

The BAK is indicated for use in both traditional open surgery and endoscopic surgical proceduras.

## Substantial Equivalence

This submission supports the position that the Axya Bone Anchor System and Kit (BAR) is substantially equivalent in design and function to the Axya Suturing and Ligating System (ASLS) [510(k) K980988], the Axya Suture Welding System and Kit (SWK) [510(k) K983108] and the Endo-Judge disposable suture placement system developed by Synergistic Medical Technologies and marketed by Ethicon Endosurgery [510(K) K932591], and that it is appropriate for its intended application. Suture applicators and suture placement devices which may be used in both endoscopic or traditional open surgical procedures have been classified under 21 CFR 876.1500. These devices are indicated for the placement of sutures to close elther traumatic or surgically produced wounds. The BAK is fabricated from materials with a substantial history of use in medical devices. Both the BAK and the predicate devices secure the suture loop with an ultrasonic weld.

The 510(k) Notice for the predicate ASLS system contains summaries of both in yitro studies which were conducted to evaluate the safety, efficacy and appropriateness of the suture welding system. These data are applicable to the Bone Anchor System because the design and components of the welding mechanism are the same as those used in both the ASLS and the SWK that were previously cleared for marketing. Data were presented which demonstrate that sutures placed by means of the suture welding process exhibit "knot strength" characteristics substantially above the USP requirements for the respective sizes and types of suture materials tested. These tests confirm that the polypropylene monofilament sutures placed with the ultrasonic suture welding technology are equivalent in holding strength (efficacy) to sutures placed with conventional knotting techniques.

Additional data presented in the Notice for the Bone Anchor System and Kit confirm that USP size 2 polypropylene monofilament sutures secured by a weld formed with the AxyaWeld Tips exceed the USP knot strength requirements, Data have been presented that demonstrate that strength requirements are met regardless of the angle of applied tension between the suture

{2}------------------------------------------------

and the bone anchor stem. Suture fatigue and elongation over the functional life have been evaluated in bench-test models and the welded sutures have been found to meet the suggested criteria following cyclic testing.

The safety of the suture delivery system was previously evaluated by placing both welded sutures and manually knotted sutures in the dorsal skin and in bowel tissue of New Zealand rabbits [510(k) K980988]. Positive (electrocautery contact) and negative (normal tissue) controls were used in the bowel study. In both animal models there was no histopathologic change seen at the suture implant sites where the suture was sealed with ultrasonic energy. The investigators concluded that there was no significant difference in safety of efficacy between the traditional method of suture placement and the technique which includes replacement of manual knot-tying with suture welding. The design of the AxyaWeld Sleeve also prevents the heated area of suture from coming into direct contact with tissues.

The ultrasonic energy source used to weld and secure the suture loop is the same generator used for the Automatic Suturing and Ligating System and that used for the Suture Welding System and Kit, and is similar to the energy source used in the UltraCision Harmonic Scalpel [510(k) K895252].

Because of design features of the BAK, no portion of the ultrasonic generator comes into contact with human tissues. Because of this, there is virtually no risk of causing a thermal injury to the patient. The suture material is heated and welded by friction and the weld is formed by melting and fusing the polymer. No "flux" or "welding rod" is employed and no new chemical entities are introduced or produced in the welding process.

The Axya Bone Anchor System and Kit are sterilized by exposure to gamma irradiation at 2.5 Mrads which provides a sterility assurance level of at least 10° or, alternatively, by means of an ethylene oxide gas sterilization process which also provides a SAL of 10°. The sterility process selected is dictated by the type of suture included with a given product configuration. These sterlity processes, the manufacturing process, and the packaging process are validated by the manufacturer.

Axya Medical, Inc. believes that the information provided establishes that similar legally marketed and pre-enactment devices have been used for the same types of clinical applications as the Axya Bone Anchor System and Kit. The materials from which the Axya device is fabricated have an established history of use in medical applications, and the specific materials used by Axya have been thoroughly tested in accordance with applicable FDA quidelines.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

# SEP 2 2 1999

Mr. Howard L. Schrayer Axya Medical, Inc. 100 Cummings Center, Suite 444C Beverly, Massachusetts 01915

Re: K992405

Trade Name: Axya Bone Anchor System and Kit Regulatory Class: II Product Code: GAW and MBI Dated: July 19, 1999 Received: July 20, 1999

Dear Mr. Schrayer:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

{4}------------------------------------------------

Page 2-Mr. Howard L. Schrayer

for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.

FDA notes that your device will contain sutures for which you have provided evidence that the suture characteristics are not altered by the sterilization process used in the device. However, you should be aware of the following additional information regarding the inclusion of a suture as a component of your device:

1. The labeling, packaging and method of sterilization of the suture cannot be changed without prior notification, review and clearance by FDA.

2. The supplier of the sutures used in your device cannot be changed without prior notification, review and clearance by FDA.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on the labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

#### 510(k) Number (if known): 510(k) K992405

Device Name: Axya, Bone Anchor System and Kit, (BAK)

Indications For Use:

The Bone Anchor System and Kit is indicated for securing soft tissue structures to bone anchors in surgical procedures where bone anchors are indicated. Examples of these procedures include:

| Shoulder procedures     | (e.g. Rotator cuff repair)                                  |
|-------------------------|-------------------------------------------------------------|
| Knee procedures         | (e.g. Patellar ligament / tendon avulsion repair)           |
| Ankle procedures        | (e.g. Achilles tendon reconstruction and repair)            |
| Foot procedures         | (e.g. Hallux valgus reconstruction)                         |
| Elbow procedures        | (e.g. Biceps tendon reattachment)                           |
| Bladder neck suspension | (female urinary incontinence due to urethral hypermobility) |

The BAK is indicated for use in both traditional open surgery and endoscopic surgical procedures.

#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use     | X |
|----------------------|---|
| Use                  |   |
| (Per 21 CFR 801.109) |   |

OR

Over-The-Counter

(Optional Format 1-2-96)

Division of General Restorative Devices
510(k) Number K992405

---

**Source:** [https://fda.innolitics.com/device/K992405](https://fda.innolitics.com/device/K992405)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).