RC TACK

K992377 · Innovasive Devices, Inc. · MBI · Oct 8, 1999 · Orthopedic

Device Facts

Record IDK992377
Device NameRC TACK
ApplicantInnovasive Devices, Inc.
Product CodeMBI · Orthopedic
Decision DateOct 8, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Intended Use

The proposed RC Tack is intended for soft tissue reattachment to host bone for the following indications: SHOULDER Repair of rotator cuff tears Acromio-clavicular separation Biceps tenodesis Deltoid repair KNEE Extra-Capsular repairs Reattachment of medial collateral ligament Reattachment of lateral collateral ligament Reattachment of posterior oblique ligament Joint capsule closure Patellar ligament and tendon avulsion repairs Extra-capsular reconstruction ITB tenodesis ANKLE Lateral and medial instability Achilles tendon reconstruction and repair

Device Story

RC Tack is a sterile, single-use polymer bone anchor for soft tissue reattachment to bone in shoulder, knee, and ankle procedures. Device consists of three components: tip, sleeve, and pin. During deployment, pin is driven into sleeve, causing radial expansion of sleeve to achieve bone fixation. Used by surgeons in clinical settings. Performance demonstrated via bone model testing showing superior holding strength compared to predicate and reference devices.

Clinical Evidence

Bench testing only. Bone model testing compared ultimate holding strength of RC Tack against Innovasive ROC EZ Suture Bone Fastener and ConTack Fastener. Results demonstrated RC Tack holding strength was greater than both comparison devices.

Technological Characteristics

Polymer bone anchor; 4.5mm size; three-component design (tip, sleeve, pin); radial expansion fixation mechanism; sterile, single-use.

Indications for Use

Indicated for soft tissue to bone fixation in shoulder, knee, and ankle procedures, including rotator cuff repair, ligament reattachment, and tendon reconstruction.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for Innovative Devices. The logo consists of a large, stylized "A" formed by two overlapping triangles. To the right of the "A" is the word "INNOVASIVE" in a bold, italicized font. Below "INNOVASIVE" is the word "DEVICES" in a smaller, non-italicized font. #### OCT 8 1999 ## 510(K) SUMMARY ## 1. SUBMITTER: Innovasive Devices, Inc. 734 Forest St. Marlborough, MA 01752 Telephone: 508-460-8229 Fax: 508-460-6661 Contact: Kathleen Morahan, Regulatory Affairs Specialist Date Prepared: July 9, 1999 ## 2. DEVICE: Trade Name: RC Tack Common Name: Bone Anchor Classification Name: "Single/Multiple Component Bone Fixation Appliances and Accessories" ## 3. PREDICATE DEVICE: The Innovasive ROC EZ Suture Bone Fastener (K970089, K971922). ## 4. DEVICE DESCRIPTION: The RC Tack is a polymmer implant intended for soft tissue reattachment to host bone in the shoulder, knee, and ankle. The implant consists of three components: a tip, sleeve, and pin. Upon deployment, the pin is driven into the sleeve expanding the sleeve radially to gain bone fixation. The RC Tack is a sterile single use device offered in one size, 4.5mm. ## 5. INTENDED USE: The proposed RC Tack is intended for soft tissue reattachment to host bone for the following indications: ## SHOULDER {1}------------------------------------------------ Repair of rotator cuff tears Acromio-clavicular separation Biceps tenodesis Deltoid repair ### KNEE Extra-Capsular repairs Reattachment of medial collateral ligament Reattachment of lateral collateral ligament Reattachment of posterior oblique ligament Joint capsule closure Patellar ligament and tendon avulsion repairs Extra-capsular reconstruction ITB tenodesis #### ANKLE Lateral and medial instability Achilles tendon reconstruction and repair ### 6. COMPARISON OF CHARACTERISTICS: The proposed RC Tack and the predicate ROC EZ Suture Bone Fastener implants utilize the same materials. Both devices also utilize the same method of bone fixation: radially expansion of the sleeve. Lastly, the indications being requested for the proposed RC Tack are already cleared for the ROC EZ Suture Bone Fastener. #### 7. PERFORMANCE DATA: The following performance data was provided in support of the substantial equivalence determination: Bone Model Testing: the ultimate holding strength of the proposed RC Tack was compared to Innovasive's currently marketed ROC EZ Suture Bone Fastener and the ConTack Fastener. The ConTack Fastener was chosen as a second comparison device because it tack soft tissue to bone, although its materials and design differ from the proposed RC Tack. The proposed RC Tack holding strength was greater than that of the predicate device and the ConTack Fastener, demonstrating substantially equivalent performance between the devices. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized caduceus-like symbol, which is a staff with a serpent entwined around it, representing healing and medicine. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the symbol. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 8 1999 OCT Ms. Kathleen Morahan Regulatory Affairs Specialist Innovasive Devices Incorporated 734 Forest Street Marlborough, Massachusetts 01752 Re: K992377 Trade Name: RC Tack Bone Anchor Product Code: MBI Regulatory Class: II Dated: July 9, 1999 Received: July 15, 1999 Dear Ms. Morahan: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. {3}------------------------------------------------ #### Page 2 Ms. Kathleen Morahan If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # INDICATIONS FOR USE K992377 The RC Tack is intended for soft tissue to bone fixation for the following indications: ## SHOULDER Repair of rotator cuff tears Acromio-clavicular separation Biceps tenodesis Deltoid repair ## KNEE Extra-Capsular repairs Reattachment of medial collateral ligament Reattachment of lateral collateral ligament Reattachment of posterior oblique ligament Joint capsule closure Patellar ligament and tendon avulsion repairs Extra-capsular reconstruction ITB tenodesis ## ANKLE Lateral and medial instability Achilles tendon reconstruction and repair Prescription Use (Per 21 CFR 801.109) besle Division of General Restorative Device 510(k) Num
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...