Retrospective clinical serum samples from 60 patients were used to correlate serial HER-2/neu measurements with clinical status (progression vs. lack of progression) to validate the assay's utility in monitoring metastatic breast cancer.
Retrospective study; Metastatic breast cancer; Clinical monitoring; Disease progression
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective clinical evaluation; Follow-up/Duration: 6-12 month period
Patients with metastatic breast cancer; Sample Size: 60 patients; Number of Sites: 3
Not applicable for this study
Correlation of serial HER-2/neu changes with clinical status (progression vs. lack of progression)
Indications for Use
The Bayer Immuno 1™ Her-2/neu Assay is an in vitro, diagnostic device intended for use in the quantitative determination of HER-2/neu in human serum. HER-2/neu values obtained may be used in the follow-up and monitoring of patients with metastatic breast cancer. HER-2/neu values should be used in conjunction with information available from clinical and other diagnostic procedures in the management of breast cancer. The clinical utility of the serum measurement of HER-2/neu as a prognostic indicator for early detection of recurrence and in the management of patients on immunotherapy regimens has not been fully established.
Device Story
Assay measures HER-2/neu protein in human serum samples; utilizes immunoassay technology with monoclonal antibodies (NB-3 conjugated to fluorescein; Fab' fragment of TA-1 conjugated to alkaline phosphatase). Samples, calibrators, or controls incubated at 37°C with conjugates; magnetic particles coated with anti-fluorescein antibody added for binding; complex washed to remove unbound molecules; colorimetric substrate added. Absorbance measured at 405 and 450 nm; rate of conversion proportional to HER-2/neu concentration. Used in clinical laboratory settings by trained personnel. Output provides quantitative HER-2/neu levels; clinicians use data to monitor disease progression in metastatic breast cancer patients. Benefits include objective tracking of clinical status over time.
Clinical Evidence
Retrospective clinical study of 60 metastatic breast cancer patients over 6-12 months. Primary endpoint: correlation between serial serum HER-2/neu changes and clinical status (progression vs. lack of progression). A ≥15% increase in HER-2/neu was used as the threshold for progression. Results showed 66/111 progression events correctly identified and 109/142 non-progression events correctly identified. Total CV <3%. Analytical sensitivity (MDC) 0.11 ng/mL.
Technological Characteristics
Immunoassay using monoclonal antibodies (NB-3, TA-1). Reagents: fluorescein-conjugated antibody and alkaline phosphatase-conjugated Fab' fragment. Separation via anti-fluorescein coated magnetic particles. Detection: colorimetric substrate, absorbance at 405/450 nm. Range: 0-250 ng/mL. Calibration: 6-level set (0-250 ng/mL). Controls: 3-level set. Software: cubic-through-zero curve-fitting algorithm.
Indications for Use
Indicated for quantitative measurement of HER-2/neu protein in human serum for follow-up and monitoring of patients with metastatic breast cancer. Used in conjunction with other clinical/diagnostic procedures. Not established for early detection of recurrence or management of immunotherapy regimens.
Regulatory Classification
Identification
A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.
Special Controls
*Classification.* Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.
Submission Summary (Full Text)
{0}------------------------------------------------
# SEP 2 9 2000
# 510(K) SUMMARY
## FOR THE
# BAYER IMMUNO 1™ HER-2/neu ASSAY
This summary of 510(k) safety and effectiveness information is being submitted in Fills summary of a requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K992228
#### GENERAL INFORMATION 1.
Trade Name:
Bayer Immuno 1™ HER-2/neu Assay
Classification Name:
Tumor-Associated Antigen Immunological Test Systems
Fredric Cline
Fredrick Clerie Director Regulatory Affairs Bayer Corporation Business Group Diagnostics 511 Benedict Avenue Tarrytown, NY 10591 Phone 914-524-2954 FAX 914-524-2500
9-22-2000
Date
{1}------------------------------------------------
This premarket notification is to add the Bayer Immuno 1™HER-2/neu Assay to the intended use of the Bayer Immuno 1™System. In this 510(k) application, the performance and clinical safety and effectiveness of the Bayer Immuno 1 HER-2/neu Assay for the management (monitoring) of metastatic breast cancer patients has been established by external clinical studies in the target population of longitudinal metastatic breast cancer patients and by comparison to accepted diagnostic procedures in accordance with the "Guidance Document For Submission of Tumor Associated Antigen Premarket Notifications, 510(k), to the FDA." Clinical evaluations of the Bayer Immuno 1 HER-2/neu Assay at three US clinical trial sites demonstrated clinical safety and effectiveness. These studies validated clinical performance characteristics and the comparison to accepted diagnostic procedures.
#### INDICATIONS FOR USE 2.
The Bayer Immuno 1™ HER-2/neu Assay is an in vitro diagnostic assay intended to quantitatively measure HER-2/neu protein in human serum on the Bayer Immuno 1 M System. HER-2/neu values obtained may be used in the follow-up and monitoring of patients with metastatic breast cancer. HER-2/neu values should be used in conjunction with information available from clinical and other diagnostic procedures in the management of breast cancer. The clinical utility of serum measurement of HER-2/neu as a prognostic indicator for early detection of recurrence and in the management of patients on immunotherapy regimens has not been fully established.
#### DEVICE DESCRIPTION 3.
The Bayer Immuno I™ HER-2/neu Assay utilizes a well-established immunoassay technology in which one monoclonal HER-2/neu antibody (NB-3) is conjugated to fluorescein (designated Reagent 1, or R1) and the Fab' fragment of another monoclonal HER-2/neu antibody (TA-1) is conjugated to alkaline phosphatase (Reagent 2, or R2). The R1 and R2 conjugates are reacted with patient sample, calibrator, or control and are incubated at 37℃ on the system. Immuno 1 Magnetic Particles coated with an antifluorescein antibody (mIMP™ Reagent) are then added and a second incubation occurs during which the antibody complex is bound. The magnetic particles complexed with the
{2}------------------------------------------------
immunological sandwich are then washed to separate unbound molecules, and a colorimetric substrate is added. The rate of conversion of substrate to a compound with absorbance at 405 and 450 nm is measured is proportional to the concentration of HER-2/neu in the sample. A cubic-through-zero curve-fitting algorithm is used to generate standard curves.
The assay has a range of zero to 250 ng/mL. The Bayer SETpoint™ HER-2/neu Calibrators consist of a set of six calibrator levels at 0, 10, 25, 60, 125, and 250 ng/mL. The Bayer TESTpoint™ HER-2/neu Controls consist of a set of three control levels at approximately 15, 50 and 100 ng/mL.
#### SUMMARY OF STUDIES 4.
Non-clinical studies were performed to validate the performance of the method according to the protocol entitled "Non-clinical Testing Protocol for the Evaluation of the of the Bayer Immuno 1™ HER-2/neu Assay." Protocols were performed at the Bayer Corporation laboratories in Tarrytown, NY and Elkhart, IN. These studies included evaluation of interfering substances, cross-reactivity, heterophilic antibodies, calibration linearity, sample linearity, parallelism (sample dilution), hook effect, reproducibility, and reagent lot-to-lot variation.
The clinical evaluation of Immuno 1 HER-2/neu Assay as an aid in the management of breast cancer patients with Stage IV metastatic disease, during the course of disease and therapy, was conducted at three US clinical trial sites.
#### Characterization of the Antigen. 4.1
The calibrator antigen used in the Bayer Immuno 1 HER-2/neu assay is derived from a recombinant 3T3 mouse cell line 3-30. HER-2/neu p105 antigen is harvested from the tissue culture media and concentrated 10-fold. Western blot analysis shows a single dominant band with a molecular weight of 105,000 Daltons consistent with the HER-2/neu extracellular domain.
{3}------------------------------------------------
#### Characterization of the Antibodies 4.2
The NB-3 monoclonal (part no. 7591MR) and TA-1 monoclonal anti-HER-2/neu (part no. 7590MR) are used in the preparation of the R1 and R2 reagents for the Bayer Immuno 1m HER-2/neu Assay. The procedures for the preparation of the NB-3 R1 Reagent fluorescein and TA-1 R2 Reagent alkaline phosphatase conjugates are standard protocols similar to those used for conjugate preparation for other 510(k) Bayer Immuno 1 assays. Bayer Corporation (Business Group Diagnostics) in Elkhart is the approved supplier for the antibodies and antibody conjugates.
#### Assay Performance 4.3
#### Specificity: Interference 4.3.1
The recovery of HER-2/neu from patient samples was studied before and after spiking the serum samples with the potentially interfering substance. Each potential interferent was tested at a maximum concentration.
The Immuno 1 HER-2/neu Assay was performed on serum samples or pools of serum to which were added various concentrations of triglycerides, hemoglobin, immunoglobulin, bilirubin, albumin or cholesterol. HER-2/neu values were also measured in serum samples after spiking with either an individual chemotherapeutic drug, "Over the Counter" (OTC) drug, vitamin or HERCEPTIN® (Trastuzumab), trademark of Genetech BioOncology, South San Francisco, CA. None of the potential endogenous or exogenous interferents demonstrated any significant interfering effects on HER-2/neu recovery.
#### Cross-Reactivity 4.3.2
Possible cross-reactions in the Immuno 1 HER-2/neu assay were studied by comparing HER-2/neu recoveries in patient samples with Human Epidermal Growth Factor. The maximum effect seen with this crossreactant was not significant (<1%).
{4}------------------------------------------------
#### Heterophilic Antibodies 4.3.3
To investigate the effectiveness of the assay's reagent formulation in minimizing heterophilic antibody interferences, patient samples with HAMA, RF titers, or autoimmune diseases were tested for possible interference in the HER-2/neu assay. The observed HER-2/neu recoveries indicated a lack of significant heterophilic interference in the assay and demonstrated the effectiveness of the reagent formulation in minimizing these interferences.
#### Linearity 4.3.4
To determine the linearity of this assay, five individual serum samples from breast cancer patients were diluted (100%, 75%, 50%, 25%, and 0%) with a pool of human sera. Recoveries of the intermediate dilutions were all between 95 and 102 percent of the expected values. These results demonstrate the linearity of HER-2/neu recoveries over the entire calibration range.
#### Hook Effect (Antigen Excess) 4.3.5
Extremely high concentrations of HER-2/neu seen in some malignant conditions may cause a "hook effect" in an assay. An excess of analyte saturates both label and capture antibody and causes the reported concentration to "hook" back into the assay range rather than be flagged as above range. HER-2/neu antigen was diluted in Level 1 Calibrator at concentrations of 300 ng/mL to 10,000 ng/mL. Results clearly demonstrated the lack of a hook effect in the Immuno 1 HER-2/neu Assay at Her-2/neu values ≤10,000 ng/mL.
#### Parallelism (Dilution Studies) 4.3.6
As a further verification of assay linearity, and to qualify the Level 1 Calibrator as a sample diluent, five patient serum samples containing a high level of HER-2/neu were diluted (100%, 75%, 50%, 25%, and 0%) with Level 1 Calibrator. Linear regression analysis for the determination of deviations from linearity for each of these clinical samples showed no deviation from linearity. The recovery of HER-2/neu assay values ranged from 90% to 103%. These data demonstrate that Level 1 calibrator is an acceptable diluent for high samples, with accurate
{5}------------------------------------------------
recovery of diluted values. This study was repeated using the same samples to qualify Immuno 1 Sample Diluent B as a sample diluent. The recovery of HER-2/neu assay values ranged from 100% to 110%. These data demonstrate that Immuno 1 Sample Diluent B is also an acceptable diluent for high samples, with accurate recovery of diluted values.
#### 4.3.7 Reproducibility
Intra- and inter-assay reproducibility were evaluated at three clinical trial sites for Bayer SETpoint Complexed Her-2/neu Calibrators, Controls and a human serum pool with HER-2/neu concentration approximately 15 ng/mL. Imprecision data pooled across Immuno 1 HER-2/neu reagent lots and systems(sites) showed maximal total coefficients of variation (%CV) of 2.4% over the range of the assay method. This is well within acceptable limits for an assay of this type.
| | Number of<br>Runs | Number of<br>Replicates | Mean | Within Run | | Total | |
|--------------|-------------------|-------------------------|-------|------------|-----|---------|-----|
| | | | ng/mL | STD DEV | %CV | STD DEV | %CV |
| Serum Pool | 80 | 400 | 15.6 | 0.27 | 1.7 | 0.28 | 1.8 |
| Calibrator 2 | 80 | 640 | 10.0 | 0.24 | 2.4 | 0.24 | 2.4 |
| Calibrator 3 | 80 | 640 | 24.9 | 0.51 | 2.0 | 0.52 | 2.1 |
| Calibrator 4 | 80 | 640 | 59.3 | 1.09 | 1.8 | 1.12 | 1.9 |
| Calibrator 5 | 80 | 640 | 123.2 | 2.33 | 1.9 | 2.38 | 1.9 |
| Calibrator 6 | 80 | 490 | 245.2 | 4.16 | 1.7 | 4.26 | 1.7 |
| Control L3 | 80 | 400 | 108.6 | 1.77 | 1.6 | 1.79 | 1.7 |
#### 4.3.8 Sensitivity (Detection Limit)
Sensitivity of the Immuno 1 HER-2/neu Assay was evaluated at three clinical trial sites by determining the Minimum Detectable Concentration (MDC). An MDC of 0.11 ng/mL was observed when assaying 480 replicates of the Immuno 1 HER-2/neu zero calibrator using two Immuno 1 HER-2/neu calibrator lots and three Immuno 1 HER-2/neu reagent lots.
{6}------------------------------------------------
#### 4.4 CLINICAL STUDIES
The clinical utility of the Bayer Immuno I HER-2/neu assay in monitoring patients with metastatic breast cancer was evaluated using retrospective serum samples from three clinical sites in the United States. Serum HER-2/neu values were measured in 60 patients with metastatic breast cancer over a 6-12 month period. These results were then separated into groups that either showed HER-2/neu values that paralleled the clinical course of disease, or HER-2/neu values that did not parallel the clinical course as follows: For all patients whose pretreatment serum HER-2/neu values exceeded 15 ng/mL, all serial measurements were analyzed visit-to-visit and serial changes in serum HER-2/neu were correlated with changes in clinical status. For each pair of serial measurements, an increase of equal or greater than 15% was considered to indicate progression, and a change of less than 15% increase was considered to indicate a lack of progression. Results presented in Table 1 show the overall correspondence of the serial HER-2/neu changes and changes in clinical status.
### Table 1: Correspondence of Serial HER-2/neu Changes and
| | Change In Clinical Status | | |
|------------------------|---------------------------|------------------------|-------|
| Change In<br>HER-2/Neu | Progression | Lack Of<br>Progression | Total |
| ≥ 15% increased | 66 | 33 | 99 |
| < 15% increase | 44 | 109 | 154 |
| Total | 111 | 142 | 253 |
| | Clinical Status |
|--|-----------------|
| | |
| | |
{7}------------------------------------------------
# Distribution of HER-2/neu Concentrations; Sensitivity and Specificity
The Immuno 1 HER-2/neu Assay was used to estimate the clinical (crosssectional) sensitivity in patients with breast cancer and characterize the frequency distribution of Immuno 1 HER-2/neu assay values in a population of breast cancer patients by stage of disease.
Specificity of the Immuno 1 HER-2/neu Assay was determined in patients with benign breast diseases, other non-malignant diseases, and in normal healthy individuals.
The Upper Limit of 15 ng/ml is comparable to the Upper Limit seen with manual kits used for research purposes, and is also equivalent to the 15 ng/ml Upper Limit found by Bayer in a previous study.
## Conclusions from the Clinical Studies
The results of this retrospective clinical trial demonstrate that the Bayer Immuno 1™ HER-2/neu assay is reproducible, and is safe and effective for the management and follow-up of patients with metastatic breast cancer.
HER-2/neu is the first cellular oncogene, which has been shown useful in this clinical setting. Data collected from this study show that changes in serum HER-2/neu concentrations over time in metastatic breast cancer patients reflect changes in clinical status such as progression of disease.
The reproducibility of the Immuno 1 HER-2/neu assay is outstanding with total CVs of less than 3%. The detection limit of 0.1 ng/mL is acceptable for the intended use of this assay. This demonstrates that this assay should provide reliable and reproducible results when tested by different laboratories using different manufactured lots of reagents at different times.
#### રું. CONCLUSIONS DRAWN FROM ALL THE STUDIES
## Valid Scientific Evidence
The conclusions drawn from these studies are based upon valid scientific evidence. Data were gathered following a well-designed protocol, in a
{8}------------------------------------------------
research laboratory operating under the principles of Good Clinical Clinical data were gathered during well controlled Practices. investigations conducted by qualified experts. Patient case histories were well documented. The results of this study are comparable to literature reports of experiences with HER-2/neu assays.
## Method Performance
Immuno 1 HER-2/neu results are highly reproducible with a maximum inter-assay %CV pooled over reagent lots and clinical sites of 2.4% over the range of the assay. Other performance characteristics including analytical sensitivity and specificity, cross-reactivity, linearity, antigen excess hook effect meet the accepted specifications set for an assay of this type.
### Safety and Effectiveness
These clinical studies confirm the safety and effectiveness of the Immuno I HER-2/neu Assay as an aid in the follow-up and management of metastatic breast cancer patients. The correspondence between Immuno 1 HER-2/neu concentrations and the patients' clinical course of disease demonstrate that the Immuno 1 HER-2/neu Assay may be used in conjunction with other clinical indicators to confirm disease progression in metastatic breast cancer patients.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three curved lines representing its wings.
SEP 2 9 2000
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Mr. Fredrick Clerie Director Regulatory Affairs Bayer Corporation Business Group Diagnostics 511 Benedict Avenue Tarrytown, New York 10591
Re: K992228
> Trade Name: Bayer Immuno 1™ HER-2/neu Assay Regulatory Class: II Product Code: NCW Dated: July 6, 2000 Received: July 10, 2000
Dear Mr. Clerie:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{10}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket This letter will anow you to oogin manating your device to a legally marketed nouncation. The I D7X Intamg of bassification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and IT you desire specific actrice for your assic devices), please contact the Office of Compliance at additionally 807.10 for mi vitio diagnestions on the promotion and advertising of your device, (301) 594-4560. Practiceliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Information on your rosponsionness and its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
# 001001
Page 1 of 1
510(k) Number (if known):
Device Name: Bayer Immuno 1TM HER-2/neu Assay
Indications For Use:
The Bayer Immuno 1™ Her-2/neu Assay is an in vitro, diagnostic device intended for use in the quantitative determination of HER-2/neu in human serum. HER-2/neu values obtained may be used in the follow-up and monitoring of patients with metastatic breast cancer. HER-2/neu values should be used in conjunction with information available from clinical and other diagnostic procedures in the management of breast cancer. The clinical utility of the serum measurement of HER-2/neu as a prognostic indicator for early atine) of and in the management of patients on immunotherapy regimens has not been fully established.
# (PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K992228
Prescription Use ✓
OR
Over-the-counter Use
Prescription Use (Per 21 CFR 801.109)
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.