← Product Code [MQS](/productcode/MQS) · K992227

# CRIT-LINE MONITOR III (CLM III) (K992227)

_In-Line Diagnostics Corp. · MQS · Jul 30, 1999 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K992227

## Device Facts

- **Applicant:** In-Line Diagnostics Corp.
- **Product Code:** [MQS](/productcode/MQS.md)
- **Decision Date:** Jul 30, 1999
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.5820
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The intended use of the CRIT-LINE III Monitor is as a non-invasive hematocrit, oxygen saturation and access blood flow-measuring device. The CRIT-LINE MONITOR III, (CLM III) is a non-invasive hematocrit, oxygen saturation and percent change in blood volume monitor used in the treatment of hemodialysis patients. In addition, the CLM III estimates access recirculation and access blood flow in hemodialysis patients.

## Device Story

CLM III monitor utilizes microprocessor-based system to process optical data from blood chamber integrated into dialysis tubing circuit. Device measures hematocrit, oxygen saturation, and blood volume; calculates access blood flow (ABF) internally via software modification. Used in clinical dialysis settings by healthcare providers to monitor patient status during treatment. Output provides real-time physiological data to assist clinicians in managing hemodialysis therapy and assessing vascular access performance.

## Clinical Evidence

Bench testing and clinical data comparison. 29 data points collected across two clinical sites (London, Ontario and Salt Lake City, Utah) compared internally calculated ABF values against external calculation method. Results showed a correlation coefficient of 0.94, with an average difference of 46 ml/min and a standard deviation of 200 ml/min.

## Technological Characteristics

Microprocessor-based monitor with integrated chip select logic, serial communication, timing, and watchdog circuits. Connects to In-Line Diagnostics Blood Chamber within dialysis tubing circuit. Non-invasive optical sensing principle. Software-based calculation of ABF. No hardware changes from predicate K972470.

## Regulatory Identification

A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.

## Predicate Devices

- CRIT-LINE MONITOR III ([K972470](/device/K972470.md))

## Reference Devices

- External method for Access Blood Flow measurement ([K982412](/device/K982412.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the date JUL 30 1999 at the top left. Below the date is the text "- IN-LINE DIAGNOSTICS-" with a logo in the middle. The logo is a black square with a white zig-zag line in the middle.

# 510(k) SUMMARY

K99222

## 1. Submitter's Information

Matthew L. Haynie In-Line Diagnostics (IDC) 117 West 200 South Farmington, UT 84025 Tel: 801-451-9000 Fax: 801-451-9007

# 510(k) Summary Prepared By:

Same as above

## 2. Date 510(k) Summary Prepared:

June 25th, 1999

#### Name of Device: 3.

CRIT-LINE MONITOR III (CLM II)

## Common Name:

Non-invasive hematocrit, blood volume and oxygen saturation monitor

## Classification Name:

Hemodialysis system monitor accessory

## 4. Identification of legally marketed device which the submitter claims equivalence:

The ABF values which were calculated by the CLM III have been compared to the approved method of ABF measurement where measured hematocrit values of the CLM III are gathered and calculated via an external source (i.e. calculator, spreadsheet, etc.) in order to obtain an Access Blood Flow value.

The 510 (k) # for the external source method of determining Access Blood Flow is K982412.

{1}------------------------------------------------

Kinzler
Page 2 of 3

It is IDC's intention to show in this submission that the ABF values gathered by the CLM III are substantially equivalent to the values resulting from the already approved method of ABF determination using the CLM III and an external source (i.e. calculator, spreadsheet, etc.)

### Description of the Subject Devices: 5.

The CLM III consists of a state-of-the-art microprocessor which has all of the chip select logic, serial communication, timing and watchdog circuits incorporated within it. The CLM III is used in conjunction with the In-Line Diagnostics Blood Chamber. The blood chamber is connected to and becomes part of the dialysis tubing circuit. The sensor from the CLM III is connected to the blood chamber which reads critical blood parameters as blood passes through the blood chamber.

## Intended use of the Subject Device: 6.

The intended use of the CRIT-LINE III Monitor is as a non-invasive hematocrit, oxygen saturation and access blood flow-measuring device.

## 7. Technological Characteristics of the Subject Devices:

Since the CLM III has not changed in any way except for the software modification concerning the measurement of Access Blood flow, please refer to the original CLM III 510(k) submission for a complete device description (see #K972470).

## 8. Discussion of Clinical Tests Performed:

ABF measurements were validated by comparing statistical data between ABF values calculated internally by the CLM III and ABF values gathered by taking CLM III hematocrit measurements and placing these measurements into a formula where they could be calculated externally by a calculator or spreadsheet program.

Sixteen data points were taken on April 8th and April 9th, 1999 at Victoria Hospital in London Ontario Canada and an additional 13 data points were taken on June 22nd, and June 24th, 1999 at Central Valley Dialysis in Salt Lake City Utah.

A correlation coefficient statistically generated was used as the comparison ' criteria to evaluate the internally calculated CLM III ABF data against CLM III. ABF data gathered via the external method. A coefficient value near 1 (i.e. . 90 or greater) denotes a strong correlated relationship of the data. Twenty-nine data points were gathered with a resulting coefficient value of .94. The average difference between the two methods of ABF measurement was 46 ml/min with a standard deviation of 200 ml/min.

{2}------------------------------------------------

K992227

fig. 3 of 3

11:41 PM

.

. .

## Conclusions 9.

In conclusion, based on comparison with the legally marketed CLM III external method for Access Blood Flow Measurement, the subject CLM III is safe and effective for internal ABF measurement and performs as well as the legally marketed CLM III external method.

{3}------------------------------------------------

Image /page/3/Picture/15 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird-like figure.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 30 1999

Mr. Matthew L. Haynie Director of Quality Assurance/Regulatory Affairs In-Line Diagnostics Corporation 117 West 200 South P.O. Box 685 Farmington, UT 84025-0685

Re: K992227 CRIT-LINE III Monitor for Access Blood Flow (Modification) Dated: June 25, 1999 Received: July 2, 1999 Requiatory Class: II 21 CFR §876.5820/Procode: 78 MQS

Dear Mr. Haynie:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class III (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regult in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), piease contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours.

CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat. and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known): K992227

Device Name: CRIT-LINE III MONITOR

i/

Indications for Use:

The CRIT-LINE MONITOR III, (CLM III) is a non-invasive hematocrit, oxygen saturation and percent change in blood volume monitor used in the treatment of hemodialysis patients. In addition, the CLM III estimates access recirculation and access blood flow in hemodialysis patients.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription for Use _ (Per 21 CFR 801.109)

OR

Over the Counter Use

(Optional Format 1-2-96)

David A. Seymore

vision Sign-Off) vision of Reproductive, Abdominal, El and Radiological De 510(k) Number

---

**Source:** [https://fda.innolitics.com/device/K992227](https://fda.innolitics.com/device/K992227)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
