BioMend Extend™ Absorbable Collagen Membrane is an absorbable, implantable material that is indicated for guided tissue regeneration procedures in periodontal defects to enhance regeneration of the periodontal apparatus.
Device Story
BioMend Extend™ is a white, compressed, non-friable, absorbable collagen membrane derived from bovine Achilles tendon. Used in dental surgery for guided tissue regeneration; acts as a physical barrier to retard epithelial downgrowth into periodontal defects while allowing nutrient passage. Implanted by clinicians; semi-occlusive design (0.004-micron pore size) integrates into surrounding tissue; completely absorbed within 18 weeks. Eliminates need for second surgical procedure for membrane removal. Provides hemostatic properties; can be cut to size in wet or dry state.
Clinical Evidence
Bench testing only. Biocompatibility studies conducted: cytotoxicity, dermal sensitization/irritation, ISO acute intracutaneous reactivity, acute systemic toxicity, genotoxicity, implantation/absorption, sub-chronic toxicity, muscular implantation, and hemolysis. Manufacturing process complies with FDA and European standards for animal tissue sourcing and viral inactivation.
Technological Characteristics
Type I collagen derived from bovine deep flexor (Achilles) tendon. Morphology: condensed laminated sheets with textured surface. Pore size: 0.004 microns. Form factor: flexible membrane, cuttable. Sterilization: Ethylene oxide (ETO). Biocompatible, non-pyrogenic, non-cytotoxic.
Indications for Use
Indicated for guided tissue regeneration in periodontal defects to enhance regeneration of the periodontal apparatus.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Integra LifeSciences Corporation Special 510(k): Device Modification BioMend Extend™ Absorbable Collagen Membrane
Kaqazale
## BioMend Extend™ Absorbable Collagen Membrane SUMMARY OF SAFETY AND EFFECTIVENESS
#### Submitter's name and address:
Integra LifeSciences Corporation 105 Morgan Lane Plainsboro, NJ 08536
#### Contact person and telephone number:
Judith E. O'Grady Senior Vice President, Regulatory Affairs, Quality Assurance and Clinical Research (609) 275-0500
#### Date Summary was prepared:
June 30, 1999
#### Name of the device:
| Proprietary Name: | BioMend Extend™ Absorbable Collagen Membrane |
|----------------------|------------------------------------------------------|
| Common Name: | Resorbable Periodontal Barrier |
| Classification Name: | Bone Filling Augmentation Material, Product Code LYC |
### Substantial Equivalence:
BioMend Extend™ Absorbable Collagen Membrane is substantially equivalent in function and intended use to the product, BioMend™ Absorbable Collagen Membrane which has been cleared to market under Premarket Notification 510(k) K924408.
#### Device Description:
BioMend Extend™ Absorbable Collagen Membrane is a white, compressed, non-friable matrix fabricated from collagen derived from bovine deep flexor (Achilles) tendon. Bovine tendon is known to be one of the purest sources of Type I Collagen that can be readily obtained and processed in commercial amounts. BioMend Extend is completely absorbable, eliminating the need for the second surgical procedure often required to remove a non-resorbable membrane. The collagen is currently used for general and dental surgery as absorbable hemostatic agents and absorbable wound dressings.
Under scanning electron microscopy, BioMend Extend™ has a morphology of condensed laminated sheets in cross-section and a textured surface. BioMend Extend" appears paperwhite in the dry state and translucent and non-slippery when wet. BioMend Extend" can be cut to any size or shape in the wet or dry state, without tearing or fragmenting.
BioMend Extend™ has an effective pore size of 0.004 microns, which will effectively retard epithelial downgrowth during early phases of healing. Being semi-occlusive, it allows essential nutrients to pass through the membrane. BioMend Extend™ incorporates into the surrounding tissue and is completely absorbed within 18 weeks.
BioMend Extend™ is sterilized by ethylene oxide gas.
{1}------------------------------------------------
### Statement of the intended use:
BioMend Extend™ Absorbable Collagen Membrane is an absorbable, implantable material that is indicated for guided tissue regeneration procedures in periodontal defects to enhance regeneration of the periodontal apparatus.
### Comparison of technological characteristics to predicate devices:
A table comparing characteristics of BioMend Extend™ Absorbable Collagen Membrane and the predicate device is provided in Table 1.
| | BioMend Extend™ Absorbable<br>Collagen Membrane | BioMend® Absorbable<br>Collagen Membrane |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Used for guided tissue<br>regeneration procedures in<br>periodontal defects to enhance<br>regeneration of the periodontal<br>apparatus. | Used for guided tissue<br>regeneration procedures in<br>periodontal defects to enhance<br>regeneration of the periodontal<br>apparatus. |
| Incorporates Same Basic<br>Design | Yes | Yes |
| Utilizes the same operating<br>principle | Cell occlusive (effective<br>pore size 0.004 microns)<br>Implantable<br>Resorbable<br>Hemostatic | Cell occlusive (effective<br>pore size 0.004 microns)<br>Implantable<br>Resorbable<br>Hemostatic |
| Incorporates same materials? | Yes, Type I Collagen | Yes, Type I Collagen |
| Sterilization Process | ETO | ETO |
| Biocompatible | Yes | Yes |
| Non-pyrogenic | Yes | Yes |
| Shelf Life | 24 months | 24 months |
| Product Size | 15 mm x 20 mm<br>20 mm x 30 mm<br>30 mm x 40 mm | 15 mm x 20 mm<br>20 mm x 30 mm<br>30 mm x 40 mm |
## Table I: BioMend Extend™ Absorbable Collagen Membrane and Predicate, BioMend® Absorbable Collagen Membrane Comparison Chart
{2}------------------------------------------------
### Safety
Biocompatibility studies have demonstrated BioMend Extend™ Absorbable Collage Membrane to be: noncytotoxic, nonpyrogenic, nonirritating, and nonsensitizing. The following studies were conducted:
- Cytotoxicity a)
- b) Dermal Sensitization /Irritation
- ISO Acute Intracutaneous Reactivity c)
- d) Acute Systemic Toxicity
- e) Genotoxicity
- Implantation/ Absorption f)
- Sub-Chronic Toxicity g)
- h) Muscular Implantation
- Hemolysis i)
The BioMend Extend™ Absorbable Collagen Membrane manufacturing process complies with the United States Food and Drug Administration and European Standards for animal tissue sourcing and viral inactivation.
#### Conclusion
BioMend Extend™ Absorbable Collagen Membrane is substantially equivalent to the unmodified device, BioMend® Absorbable Collagen Membrane. The modifications do not affect the intended use or fundamental scientific technology of the device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white circular seal. The seal contains the logo for the U.S. Department of Health & Human Services. The logo is a stylized image of three human figures, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged around the top of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 26 1999
Judith E. O'Grady, RN, MSN Senior Vice President Requlatory Affairs, Quality Assurance and Clinical Research Integra LifeSciences Corporation 105 Morgan Lane Plainsboro, New Jersey 08536
Re : K992216 BioMed® Extend™ Absorbable Collagen Trade Name: Membrane Unclassified Requlatory Class: Product Code: LYC Dated: June 30, 1999 Received: July 1, 1999
Dear Ms. O'Grady:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. O'Grady
this response to your premarket notification Please note: submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA asbeing of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to ens robet notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Director
Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radioloqical Health
Enclosure
{5}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
## 510(k) Number:
Device Name: BioMend Extend™ Absorbable Collagen Membrane
# Indications for Use:
BioMend Extend™ Absorbable Collagen Membrane is an absorbable, implantable material that is Dioinend Exema - 110001 views - Somiga procedures in periodontal defects to enhance regeneration of the periodontal apparatus.
# (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
| | Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--|--------------------------------------------------------|--|
|--|--------------------------------------------------------|--|
| Prescription Use | Or | Over-the-Counter Use ______ |
|---------------------|----|-----------------------------|
| (Per 21 CFR 801.109 | | |
Optional Format 1-2-96)
| | Susan Runne <div>(Division Sign-Off)</div> <div>Division of Dental, Infection Control,</div> <div>and General Hospital Devices</div> <div>"(k) Number 1990216</div> |
|--|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
3A
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.