K991929 · Intratherapeutics, Inc. · FGE · Aug 4, 1999 · Gastroenterology, Urology
Device Facts
Record ID
K991929
Device Name
INTRASTENT
Applicant
Intratherapeutics, Inc.
Product Code
FGE · Gastroenterology, Urology
Decision Date
Aug 4, 1999
Decision
SESU
Submission Type
Special
Regulation
21 CFR 876.5010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The IntraStent™ Biliary Stent is indicated as a palliative treatment for malignant neoplasms in the biliary tree.
Device Story
IntraStent™ is a balloon-expandable biliary stent; constructed from 316L stainless steel hypotube with laser-cut open-lattice design. Device provided unmounted; manually crimped onto noncompliant PTA balloon catheter by clinician. Upon inflation, stent expands to conform to duct inner luminal surface; retains expanded state after balloon deflation. Used in clinical setting for palliative treatment of malignant biliary strictures. Output is mechanical support of biliary duct. Benefits patient by maintaining duct patency in presence of malignant neoplasms.
Clinical Evidence
Bench testing only. Performance testing verified IntraStent™ met same characteristics as predicate ITI Stent (K980290). Pyrogenicity testing performed on lot-to-lot basis supports nonpyrogenic claim.
Indicated for palliative treatment of malignant neoplasms in the biliary tree.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{1}------------------------------------------------
## 510(k) Summary (Page 1 of 2)
K991929
Product Name: IntraStent™ Common Name: Biliary catheter
| Submitter's Name: | IntraTherapeutics, Inc, 651 Campus Drive<br>St. Paul, MN 55112 |
|-------------------|------------------------------------------------------------------------------------------------------------|
| Official Contact: | Cathy Yohnk<br>Senior Clinical Research/Regulatory Affairs Associate<br>Tel. 651-697-2003 Fax 651-697-2080 |
Summary Preparation Date: June 7, 1999
This summary is provide in compliance with section 513(I(X)(A) of the Act and This summary is provide in comphance with contained in this premarket notification submission.
The product trade and common name are IntraStent™ and biliary catheter, respectively.
Comments of the career of CAL CED 50768 5010 as a biliary cathercrand This is a Class II product classified under 21 CFR 88768.5010 as a biliary catheter and This is a Class II product classince ander is claimed to IntraTherapeuics, Inc. ITT Stent (K980290) and Cordis Corporation PALMAZ™, PALMAZ™, PALMAZ-SCHATZ™, and (K980290) and Cordis Corporation (KB05720, K911581, K964688, K980653).
The IntraStent™ is a balloon expandable stainless steel stent with an open attisce design. The device is provided unmounted, to be manually crimped onto a noncompliant PTA I he device is provided uninounced, to be manadis computer luminal surface and inflation the crimped stent expands to conform to the duct inner luminal surface and retains the expanded state upon balloon deflation.
The intended use is "as a palliative treatment for malignant neoplasms in the biliary tree".
Summary of technological characteristics: The IntraStent™ provides a line extension to Sundhaly of technological characteristics and predicate stents are balloon expandable the 111 Stein (1596290). The miracted series of slots/apertures into a 316L stainless steel hypotube. The IntraStent™ and predicate stent cuts are made with a laser. Both the IntraStent™ and predicate stents are cleaned, electro-polished, packaged in a double sterile barrier and sterilized. The IntraStent™ is ethylene oxide sterilized and the sterne "oarrier" and "stermzed: "The "madelem" and predicate stents are provided
<sup>*</sup>This document uses the term "substantial equivalence" as intended in 21 CFR 807.87, and not as defined in Title 36 of the US Code.
{2}------------------------------------------------
## 510K Summary (Page 2 of 2)
unmounted, additionally the predicate Palmaz™ stent can be provided premounted on a PTA balloon catheter. For the unmounted stent, the Palmaz™ stent offers a separate nonsterile crimping tool while the ITI product does not. Both the IntraStent™ and predicate stents include use of manual compression using one's thumbs and forefingers to compress the stent onto the delivery balloon catheter.
Bench tests were performed to verify that the IntraStent™ met the same performance characteristics as the predicate ITI Stent (K980290). Pyrogenicity testing performed on a lot to lot basis supports the nonpyrogenic claim.
The IntraStent™ is substantially equivalent to the currently marketed ITI Stent and the Cordis Corporation/Johnson & Johnson PALMAZTM, PALMAZ-SCHATZ™, and PERFLEX™ Balloon-Expandable Biliary Stent as a palliative treatment for malignant strictures of the biliary tree. As demonstrated the IntraStent™ is identical in materials, indication for use and technological characteristics. Performance testing (bench) further supports a substantial equivalence claim. The collective evidence therefore provides assurance that the IntraStent™ meets the requirements that are considered acceptable for the intended use.
{3}------------------------------------------------
Image /page/2/Picture/2 description: The image shows a circular logo with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around the edge. In the center of the circle is a stylized image of an eagle with its wings spread. The eagle is facing to the left and has three lines representing its body.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG - 4 1999
Ms. Cathy Yohnk Ms. Cathy Yollik
Senior Clinical Research / Regulatory Affairs Associate IntraTherapeutics, Inc. 651 Campus Drive St. Paul, Minnesota 55112
K991929 Re: IntraStent™ Biliary Stent Regulatory Class: II 21 CFR 876.5010 Product Code: 78 FGE Dated: July 8, 1999 Received: July 9, 1999
Dear Ms. Yohnk:
We have reviewed your Section 510(k) notification of intent to market the device referenced
we have the states and the besice is whetentially equivalent (for the indications We have reviewed your Section 510(k) notification of intention of the indications for use
above and we have detectived the device is substantially equivalent in interstate co above and we have determined the device is substanted in interstate commerce
stated in the enclosure) to legally marketed prodicate devices marketed in interstate comments, o stated in the enclosure) to legally marketed predical Device Amendments, or to devices that
prior to May 28, 1976, the enactment date of the Rederal Food. Drug, and Cosmetic prior to May 28, 1976, the enactment date of the Federal Food, Drug, and Cosmetic
have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosm have been reclassified in accordance with the provisions of the general controls provisions of
Act (Act). You may, therefore, market the device, subject to the Act include Act (Act). You may, therefore, market the device, subject covinsions of the Act include
the Act and the limitations described below. The general controly provise, labeling, the Act and the limitations described below. The general of the states of the station of the subsidiary, and the supplied requirements against misbranding and adulteration.
The Office of Device Evaluation has determined that there is a reasonable likelihood that this The Office of Device Evaluation nas detectimed that the proposed labeling and that such use
device will be used for an intended use not identified in the Act, the following device will be used for an intended use not lucinities an the Section 513 (i) (I) (E) of the Act, the following
could cause harm. Therefore, in accordance with Section of could cause harm. "Therororo, in accorder of the device's labeling:
The safety and effectiveness of this device for use in the vascular system have not been established.
Furthermore, the indication for billary use must be prominently displayed in all labeling, including Furthermore, the indication for unlary asc mast or promotional materials, in close pouch, box, and career labels, of a similar point size, and in bold print.
{4}------------------------------------------------
Page 2 - Ms. Cathy Yohnk
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations (Fremarket Approval), it may of subject to sae of Federal Regulations, Title 21, Parts 800 to 895. allecting your device can be found in in assumes compliance with the Current Good A substantially oquivalent develments, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS Medical Devices. Gelleral regulation (FDA) will verify such assumptions. Failure to mspections, the Pood and Drag result in regulatory action. In addition, FDA may publish Comply with the GMI Togation inaly in the Federal Register. Please note: this itimel announcements concerning your averaission does not affect any obligation you might have
response to your premarket notification submission does not at Redict Rediction response to your premarket notification sector cevices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
The FDA finding of substantial equivalence of your device to a legally marketed predicated and The LDA inding of substantial oquridential of traits your device to proceed to the market. This resurts in a clussinoution for your device as described in your 510(k) premarket notification if the limitation statement above is added to your labeling, as described.
Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. I herefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.
If you desire specific information about the application of other labeling requirements to your 11 you desire specific manen as as a many as a many and and stores), please contact the Office of Compliance at (301) 594-4616. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html"
Sincerely yours,
Sincerely yours,
Susan Alpert, Ph.D., M.D. Director Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page _ l of 1
510(k) Number (if known): K991929
Device Name: IntraStent™ Biliary Stent
FDA's Statement of the Indications For Use for device:
The IntraStent™ Biliary Stent is indicated as a palliative treatment for malignant neoplasms in the biliary tree.
Prescription Use / OR
(Per 21 CFR 801.109)
Over-The-Counter Use
Lit
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Device ટાપ્ડા
510(k) Number K991929/5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.