K991847 · Argomed, Inc. · KNS · Aug 26, 1999 · Gastroenterology, Urology
Device Facts
Record ID
K991847
Device Name
THERMOFLEX SYSTEM
Applicant
Argomed, Inc.
Product Code
KNS · Gastroenterology, Urology
Decision Date
Aug 26, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Thermoflex system is intended for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia ("BPH"). It is indicated for use in men over the age of 50 years with prostate lengths between 2.0 cm and 6.4 cm who present symptoms of urinary outflow obstruction secondary to BPH.
Device Story
Thermoflex system provides thermal treatment for BPH-related urinary outflow obstruction. Device consists of heating console/pump and sterile, multi-balloon WIT prostatic catheter. Physician inserts catheter into bladder; locating balloon positioned at bladder neck; treatment balloon dilated within prostate to ensure tissue contact. Console heats water externally, circulating it through insulated catheter to treatment balloon. Heat transfers conductively to prostatic tissue, inducing coagulation necrosis and tissue devitalization. Pressure from dilation also relieves urethral obstruction. System used in clinical setting; physician-operated. Output is physical tissue ablation; healthcare provider monitors patient symptom improvement and flow rates post-treatment. Benefits include relief of urinary obstruction symptoms without traditional surgery.
Clinical Evidence
Multicenter, nonrandomized, prospective clinical study of a single patient group. Evaluated changes in BPH symptom severity, urine flow rates, prostate volume, and quality of life from pre- to post-treatment. Results showed clinically and statistically significant improvements across all metrics. Adverse events consistent with standard BPH catheterization procedures.
Technological Characteristics
Catheter-based thermal therapy system. Components: heating console/pump and sterile, multi-balloon WIT prostatic catheter. Sensing/actuation: conductive heat transfer via water-filled balloon dilation. Energy source: external water heating console. Connectivity: none. Sterilization: sterile catheter. Software: none described.
Indications for Use
Indicated for men >50 years old with prostate lengths 2.0-6.4 cm exhibiting symptoms of urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH).
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
{0}------------------------------------------------
AUG 26 1999
1991847 PG. I OF A
## Argomed, Inc.'s Thermoflex system
| Name of Device | Thermoflex™ Water-Induced Thermotherapy (WIT) System | |
|-----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|
| Common or Usual Name | Thermal Therapy BPH Catheter System or WIT System | |
| Classification Names | Electrosurgical Cutting and Coagulation Device (21 C.F.R. § 878.4400), Laser Surgical Instruments (21 C.F.R. § 878.4810), and Urethral Dilator (21 C.F.R. § 876.5520) | |
| Product Codes | GEI, GEX, and KOE | |
| Submitter | Argomed, Inc.<br>3800 Gateway Centre Boulevard<br>Suite 308<br>Morrisville, North Carolina 27560 | |
| Phone:<br>Facsimile:<br>Contact Person:<br>Date Prepared: | (919) 469-2340<br>(919) 469-2779<br>Richard B. Klein<br>May 28, 1999 | |
| Predicate Devices | | |
| Trade Name | Manufacturer | 510(k) |
| Transurethral Needle<br>Ablation ("TUNA") System | VidaMed, Inc. | K965199, K960918,<br>and K951245 |
| Optilase® Nd:YAG<br>Laser ("VLAP") system | Trimedyne, Inc. | K954597 |
| Prostatic Balloon<br>Dilation Catheter | Apex | K913477 |
## Intended Use
The Thermoflex system, the VidaMed TUNA system, and the Trimedyne VLAP system have the same intended use: to provide thermal treatment to the prostate to treat symptoms due to urinary outflow obstruction secondary to BPH. The Thermoflex system and the Apex catheter are intended to treat symptoms due to urinary outflow obstruction secondary to BPH. The Thermoflex system and the predicate devices are indicated for use in men over the
{1}------------------------------------------------
age of 50 years exhibiting the symptoms of urinary outflow obstruction. Both the Thermoflex system and the VidaMed TUNA system define limits on the length or size of the prostate prior to treatment to insure that thermal treatment is provided to the target tissue without exposure of non-target tissue. For the Thermoflex system, the minimum prostate length, to be determined by cystoscopic examination, is 2.0 cm, and the maximum length is 6.4 cm. For the VidaMed TUNA system, the prostate must be between 20 cc and 50 cc in size. Thus, the Thermoflex system has the same intended use and similar indications for use as these predicate devices.
## Substantial Equivalence
The Thermoflex system, the VidaMed TUNA system, the Trimedyne VLAP system, and the Apex Prostatic Catheter have the same intended use: the treatment of symptoms due to urinary outflow obstruction secondary to BPH. The Thermoflex system and the predicate devices have similar principles of operation and technological characteristics. Each device is inserted into the urethra and serves to relieve pressure placed on the urethra by the prostate. Relief of the pressure reduces the symptoms of urinary outflow obstruction. As opposed to surgical procedures, such as transurethral resectioning of the prostate ("TURP"), the Thermoflex system and the TUNA and VLAP predicate devices displace and necrose prostatic tissue by the application of heat to the prostatic tissue. The Thermoffex system and the Apex Prostatic Catheter both relieve pressure of the prostate on the urethra by the application of pressure induced by dilation of the catheter.
The technological differences between the Thermoflex system and its predicate devices are the combination of heat transfer with dilation, the method of heat generation, the control of the generated heat, and the point of energy delivery within the prostate.
The Thermoflex system combines pressure from dilation of the catheter with the introduction of heat into the prostatic tissue to induce coagulation necrosis. The pressure from the Thermoflex system dilation ballows the uniform distribution of heat to the prostate providing the therapeutic necrotic effect. With the Thermoflex system, heat is generated externally to the body, and then circulated to the prostate through an insulated catheter that protects nontarget tissues, such as the rectum, the bladder, and the penile urethra. The VidaMed TUNA and the Trimedyne VLAP systems generate heat within the prostate by the conversion of RF energy or radiant light into heat. Heat is controlled in the Thermoflex system by the lack of energy generation within the prostate. The heated water cannot increase in temperature or impart additional energy into the prostate, as no energy source is present within the body. Both the Thermoflex system and the Trimedyne VLAP system apply the heat directly to the prostatic urethral lumen, inducing coagulation necrosis. The VidaMed TUNA
{2}------------------------------------------------
system employs RF probes that are inserted through the urethral lumen into the prostate itself.
## Performance Characteristics
The Thermoflex system is a catheter based thermal therapy device for the treatment of symptoms due to urinary outflow obstruction secondary to BPH. The Thermoflex system combines a heating console/pump with a sterile, multiple balloon WIT prostatic catheter. Following insertion of the WIT prostatic catheter into the bladder through the urethra, a "locating" balloon is filled with a small volume of air and positioned within the bladder neck to insure proper placement of the treatment balloon within the prostate. A predetermined volume of water is added to the balloon dilation catheter ("treatment balloon"), causeing the treatment balloon to dilate within the prostate, creating pressure and insuring contact of the treatment balloon's surface with the prostatic tissue. Water is heated externally to the body and pumped through the thermally insulated catheter to the treatment site and back to the heating console/pump. At the treatment site, heat from the water is exposed to the urethra and prostatic tissue by means of physical contact of the treatment balloon with the target tissue. Heat from the water is transferred conductively to the tissue, inducing tissue ablation due to coagulation necrosis and tissue devitalization.
The equivalence of the Thermoflex system to the predicate devices in the treatment of symptoms of urinary outflow obstruction secondary to BPH was assessed in a multicenter, nonrandomized, prospective clinical study of a single patient group. The objectives, study endpoints, and methods of analysis focused upon assessing changes in severity of BPH symptoms and signs from pretreatment to post-treatment, and in a comparison of these endpoints to published safety and effectiveness information available for the TUNA system and VLAP system predicate devices. The clinical evaluation demonstrated that the Thermoflex system for Water-Induced Thermotherapy treatment of BPH induced clinically and statistically significant improvements in urine flow rates, prostate volume, prostate symptomatology, and quality of life assessments. The adverse events observed using this device are typical for devices used to treat BPH or for catheterization in general. Thus, the WIT procedure using the Thermoflex system is a safe and effective treatment for BPH ..
# Conclusion
The Thermoflex system has the same intended use and very similar principles of operation, technological characteristics, clinical efficacy, and clinical safety as a combination of several class II predicate devices, the VidaMed, Inc. Transurethral Needle Ablation ("TUNA") system (K965199, K960918, and K951245), the Trimedyne Optilase® Nd:YAG Laser ("VLAP") system (K954597), and the Apex Prostatic Balloon Dilatation Catheter (K913477). Thus, the
{3}------------------------------------------------
...
:
Thermoflex system is substantially equivalent to legally marketed devices intended to treat symptoms due to urinary outflow obstruction secondary to BPH.
{4}------------------------------------------------
Rockville MD 20850
Food and Drug Administration
9200 Corporate Boulevard
Image /page/4/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three wing-like shapes, positioned to the right of a circular arrangement of text. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" and is arranged around the upper portion of the circle.
AUG 26 1998
Mr. Richard B. Klein President & CEO ARGOMED, Inc. 3800 Gateway Centre Blvd. Suite 308 Morrisville, NC 27560
Re: K991847 Thermoflex™ Water-Induced Thermotherapy System Dated: May 28, 1999 Received: May 28, 1999 Regulatory Class: II 21 CFR §876.4300/Procode: 78 KNS 21 CFR §876.5520/Procode: 78 KOE 21 CFR §878.4810/Procode: 79 GEX
Dear Mr. Klein:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat. and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): _14991847
Device Name: Argomed, Inc. Thermoflex™ Water-Induced Thermotherapy System
Indications for Use:
The Thermoflex system is intended for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia ("BPH"). It is indicated for use in men over the age of 50 years with prostate lengths between 2.0 cm and 6.4 cm who present symptoms of urinary outflow obstruction secondary to BPH.
#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division S . Off) Division o productive, Abdominal, ENT, and Radiolo al Devices 510(k) Number .
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use _ (Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.