K991829 · Biostar, Inc. · LLH · Nov 2, 1999 · Microbiology
Device Facts
Record ID
K991829
Device Name
CDTOX A OIA
Applicant
Biostar, Inc.
Product Code
LLH · Microbiology
Decision Date
Nov 2, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.2660
Device Class
Class 1
Indications for Use
The BioStar® CdTOX A OlA assay is an Optical ImmunoAssay test for the rapid qualitative detection of Clostridium difficile Toxin A in human fecal samples from patients suspected of having Clostridium difficile associated disease (CDAD). This test is intended for in vitro diagnostic use as an aid in the diagnosis of CDAD.
Device Story
CdTOX A OIA is an in vitro diagnostic test for qualitative detection of Clostridium difficile Toxin A in human fecal samples. Principle of operation: Optical ImmunoAssay (OIA) technology; antigen-antibody binding on a silicon wafer surface; mass enhancement via substrate addition increases molecular thin film thickness; change in optical thickness alters reflected light path; visually perceived as color change (purple spot on gold background). Used in clinical laboratories; performed by laboratory personnel. Output is a visual colorimetric result; aids clinicians in diagnosing CDAD. Benefits patients by providing rapid detection of C. difficile toxin.
Clinical Evidence
Clinical evaluation included 874 specimens compared against Premier EIA and cytotoxicity assay (CTA). Sensitivity 77.3%, specificity 93.8%, concordance 91.0% vs CTA. Concordance between CdTOX A OIA and Premier EIA was 95.0%. Reproducibility studies at multiple hospital and physician office labs showed 94.4% to 97.8% agreement.
Technological Characteristics
Optical ImmunoAssay (OIA) on silicon wafer surface. Manual test kit includes reaction tubes, conjugate, wash solution, substrate, test devices, controls, pipettes, and sampling brushes. Qualitative visual readout. No electronic components or software.
Indications for Use
Indicated for rapid qualitative detection of Clostridium difficile Toxin A in human fecal samples from patients suspected of having Clostridium difficile associated disease (CDAD).
Regulatory Classification
Identification
A microorganism differentiation and identification device is a device intended for medical purposes that consists of one or more components, such as differential culture media, biochemical reagents, and paper discs or paper strips impregnated with test reagents, that are usually contained in individual compartments and used to differentiate and identify selected microorganisms. The device aids in the diagnosis of disease.
{1}------------------------------------------------
# NOV - 2 1999
8.0 510(k) SUMMARY (page 1 of 4)
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: _ K991829
A. Safety and effectiveness information required per [5807.92(a)(1)]:
- SUBMITTER'S NAME: BioStar, Inc. .
- ADDRESS: 6655 Lookout Rd. Boulder, CO 80301 .
- TELEPHONE: (303) 530-3888 ext. 659 .
- . FAX: (303) 530-6601
- CONTACT PERSON: Curtis A. Carlson .
- DATE 510(k) SUMMARY PREPARED: May 28, 1999; revised October 27, 1999. ●
B. Safety and effectiveness information required per [\$807.92(a)(2)]:
- TRADE OR PROPRIETARY NAME: CATOX A OLA® .
- . COMMON NAME: C. difficile Toxin A Assay
- CLASSIFICATION NAME: Direct Antigen Detection Assay, Clostridium difficile .
C. Identification of legally marketed device to which we are claiming equivalence [§807.92(a)(3)]:
- TRADE OR PROPRIETARY NAME: PREMIER™ C. difficile Toxin A ●
- . REGULATORY CLASS: Class I
- PRODUCT CODE: 83LLH .
- MANUFACTURER: Meridian Diagnostics, Inc. .
- 510(k) NUMBER: K903456 ●
Note: Performance of the CdTOX A OIA product was also established versus cytotoxicity assay.
## D. Description of device [§807.92(a)(4)]:
## Principle of the Test:
The CdTOX A OIA test involves the qualitative detection of the enterctoxin (Toxin A) produced by Clostridium difficile. The Optical Immuno Assay technology enables the direction of a physical change in the optical thickness of molecular this change is a result of antigen-antibody binding on an optical surface (silicon wafer). When a sample containing Toxin A (antigen) from Clostridium difficile is placed directly on the optical surface, the immobilized specific antibody captures the antigen. After washing, the substrate is added, increasing the thickness (Mass Enhancement) of the molecular thin film. This change in thickness alters the reflected light path and is visually perceived as a color changes in optical thickness produce a distinct, visible color change. A positive result appears as a purple spot on the predominant gold background. When antigen is not present in the specimen, no binding takes place. Therefore, the optical thickness remains unchanged and the surface retains the original gold color indicating a negative result.
{2}------------------------------------------------
Biostar CdTOX A OLA 510(k) Summary. Revised October 27, 1999
### 8.0 510(k) SUMMARY (page 2 of 4)
### Device Components:
The CdTOX A OLA test kit contains the following:
Reaction Tubes Reagent 1: Conjugate Wash Solution Substrate Test Devices Positive Control Negative Control Transfer Pipettes Sampling Brushes
- E. Intended use of device [§807.92(a)(5)]:
The BioStar® CdTOX A OlA assay is an Optical ImmunoAssay test for the rapid qualitative detection of Clostridium difficile Toxin A in human fecal samples from patients suspected of having Clostridium difficile associated disease (CDAD). This test is intended for in vitro diagnostic use as an aid in the diagnosis of CDAD.
F. Technological characteristics [§807.92(a)(6)]:
| Technological Characteristic | Predicate Device (Premier TM Toxin A) | Our Device (CdTOX A OLA) |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The PremierTM C. difficile Toxin A assay is<br>an Enzyme Immunoassay for the Detection<br>of C. difficile Toxin A in Stool Specimen. | The BioStar® CdTOX A OIA assay is<br>an Optical ImmunoAssay test for the<br>rapid qualitative detection of<br><i>Clostridium difficile</i> Toxin A in<br>human fecal samples from patients<br>suspected of having <i>Clostridium<br/>difficile</i> associated disease (CDAD).<br>This test is intended for <i>in vitro</i><br>diagnostic use as an aid in the<br>diagnosis of CDAD. |
| Detection | Detects <i>C. difficile</i> Toxin A | Detects <i>C. difficile</i> Toxin A |
| Technology | Enzyme Immunoassay (EIA) Microwell | Optical Immunoassay (OIA) |
| Specimens Evaluated | Stool | Stool |
{3}------------------------------------------------
### 8.0 510(k) SUMMARY (page 3 of 4)
#### G. Summary of nonclinical testing [§807.92(b)(1)]:
#### Analytical Sensitivity:
The analytical sensitivity was determined by adding known concentrations of Toxin A to five specimens each of liquid, semi-solid, and solid stools. Seeded specimens were tested in duplicate by the CdTOX A OfA test. Results are summarized in the table below.
## Limit of Detection by Stool Type
| Toxin A Seeded Matrix | Limit of Detection (ng/g) |
|-----------------------|---------------------------|
| Buffer Control | 3 |
| Liquid Stool | 3-8 |
| Semi-Solid Stool | 4-8 |
| Formed Stool | 3-5 |
#### Analytical Specificity (Cross Reactivity):
To determine the analytical specificity of the CdTOX A OIA test, microorganisms were grown on the appropriate media and suspended to a concentration of approximately 1.0 X 10 cfu/mL or greater and tested in the absence and presence of Toxin A near the lower limit of detection of the assay. . Cryptosporidium parvum and Giardia lamblia were rested at 1.25 X 10° cysts reactivity or interference was observed.
#### Interfering Substances:
Interference testing was performed to determine the effect of substances which might be encountered in the collection of specimens. Whole blood, Red Blood Cells (RBC), White Blood Cells (WBC), mucous (25mg/g), and barium sulfate (58% w/w suspension) were tested and did not interfere with the CdTOX A OLA test. All assays were run per the procedure described in the package insert.
#### Sample/Toxin Stability
Toxin stability in liquid, semi-solid, and formed stool was judged by spiking Toxin A (Lee Laboratories) at levels near the analytical sensitivity of the assay and testing at the specified times. Spiked samples were initially held at room temperature for 2 hours then transferred to 2-8℃ storage. These results show that over the sample consistency types tested, the toxin is sufficiently stable to allow detection in samples handled as recommended in the package insert.
#### Stool Consistency Reactivity:
The reactivity of stools with liquid, semi-solid, or formed consistency was evaluated using the OdTOX A OIA test and microwell EIA test. The positivity rates were similar for the two assays across all three types of stool specimens. All of the specimens in the clinical performance evaluation of the CdTOX A OLA rest were submitted for C. difficile testing on the basis of clinical history of the patient, not the consistency of the specimen. Test results are illustrated in the table below.
## Reactivity Data From Four Clinical Sites
#### Stool Consistency
| | Liquid (445) | Semi-solid (298) | Formed (80) |
|---------------|--------------|------------------|-------------|
| CdTOX A OIA + | 42 (9%) | 54 (18%) | 21 (26%) |
| EIA + | 51 (11%) | 50 (17%) | 19 (24%) |
{4}------------------------------------------------
#### 8.0 510(k) SUMMARY (page 4 of 4)
- Summary of clinical testing [§807.92(b)(2)]: H.
#### Reproducibility
Reproducibility testing was conducted using a stool marix at three hospital laboratories. Ten blinded samples were tested at each site at three separate times. Samples consisted of negative, low, and moderate levels of C. difficile Toxin A. The concentration of toxin in the low samples was less than 2-fold over the minimum detectable limit. The overall reproducibility was 94.4% (85/90). An additional study was done using artificial stool matrix. Comparable samples were prepared and tested at four hospital laboratories and three physician office laboratories (POLs). Overall reproducibility in this study was 97.6% (205/210); at the hospitals, 97.5% (117/120) at the POLs, 97.8% (88/90).
#### Clincal Sensitivity and Specificity
A sudy comparing the CdTOX A OIA test and a commercial microwell immunoassay, the Premier EIA, to cytotoxicity assay (CTA) was performed at a central hospital lab. The sensitivity, specificity, and concordance of the CdTOX A OIA test was 77.3%, 93.8%, and 91.0% respectively. The sensitivity, specificity, and concordance of the Premier ELA test was 81.8%, 92.0%, and 90.3% respectively.
The CdTOX A OIA test was evaluated versus the Premier EIA at four clinical trial sites (central hospital labs) located in the Southwest, Mid-Atlantic and East Coast regions of the United States. A total of 874 specimens from patients suspected of having or with a history of CDAD were evaluated by the CdTOX A OLA test and the Premier ELA. Of these, 823 gave unequivocal results using the BlA method. Concordance between the Cd TOX A OlA test and the Premier EIA test was 95.0% (98+/+, 684 -/-); discordance was 5.0% (19 +/-, 22 -/+).
- I. Conclusions from nonclinical / clinical testing [§807.92(b)(3)]:
The results of the above described internal and external studies demonstrate that the CdTOX A OLA test is as safe and effective as the cleared predicate device.
No additional information has been requested by FDA at this time.
J. Additional information [§807.92(d)]:
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and two wavy lines representing its legs. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.
NOV - 2 1999
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Curtis A. Carlson, Ph.D. Director, Regulatory Affairs BioStar, Inc. 6655 Lookout Road Boulder, Colorado 80301
Re: K991829 Trade Name: CdTOX A OIA® Test Kit Regulatory Class: I Product Code: LLH Dated: October 7, 1999 Received: October 8, 1999
Dear Dr. Carlson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{6}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely vours.
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
#### INDICATIONS FOR USE STATEMENT
| 510(k) Number (if known): | K991829 |
|---------------------------|--------------|
| Device Name: | CdTOX A OIA® |
Indications For Use:
The BioStar® CdTOX A OlA assay is an Optical ImmunoAssay test for the rapid qualitative detection of Clostridium difficile Toxin A in human fecal samples from patients suspected of having Clostridium difficile associated disease (CDAD). This test is intended for in vitro diagnostic use as an aid in the diagnosis of CDAD.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Qffice of Device Evaluation (ODE)
Woody Dubois
(Division Sign-Off) Division of Clinical Laboratory Devices K991829 510(k) Number _
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.