K991722 · Sunrise Medical Hhg, Inc. · CAW · Nov 3, 1999 · Anesthesiology
Device Facts
Record ID
K991722
Device Name
DEVILBISS 5 LITER OXYGEN CONCENTRATOR
Applicant
Sunrise Medical Hhg, Inc.
Product Code
CAW · Anesthesiology
Decision Date
Nov 3, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The revised DeVilbiss 5 Liter Oxygen Concentrator is designed to provide supplemental low flow oxygen therapy for patients suffering from COPD, cardiovascular disease, and lung disorders. The oxygen concentrator is used in home type environments: homes, patient care facilities, etc.
Device Story
Device is a 1-5 lpm pressure swing adsorption oxygen concentrator; uses molecular sieve technology to produce 93% +/-3% oxygen. Input is room air drawn via piston compressor; air passes through filters; pneumatic valve directs air into one of two sieve beds. Nitrogen adsorbed in pressurized bed; oxygen flows to patient. Simultaneously, other bed depressurizes to exhaust nitrogen. Revised model uses fixed-time cycle controlled by microprocessor; predicate used pressure-based cycle. Device used in homes/care facilities; operated by patients or caregivers. Output is continuous oxygen flow. High-pressure relief valve provides mechanical safety. Benefits include reliable supplemental oxygen delivery for respiratory conditions.
Clinical Evidence
Bench testing only. Performance evaluated by measuring oxygen purity output at 1-5 lpm. Both subject and predicate devices met the 90% minimum purity requirement, with theoretical maximum of 95.5%. No clinical data provided.
Technological Characteristics
Pressure swing adsorption oxygen concentrator; 1-5 lpm flow. Uses UOP Oxysiv-5 molecular sieve material. Single pneumatic valve system; fixed-time cycle control via microprocessor. Mechanical pressure relief valve for high-pressure safety. Dimensions/form factor not specified. Standalone operation. Electronics modified from predicate to support timed cycling; reduced alarm set (High Pressure, Long Cycle, No Flow removed).
Indications for Use
Indicated for patients with COPD, cardiovascular disease, or lung disorders requiring supplemental low flow oxygen therapy in home or patient care facility settings.
Regulatory Classification
Identification
A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).
{0}------------------------------------------------
## 6.0 510(k) Summary
Submitter's Name: Greg Clites
Address: Sunrise Medical HHG Inc Respiratory Products Division 100 DeVilbiss Drive P.O. Box 635 Somerset, PA 15501-0635
Phone #: 814-443-7441
Fax #: 814-443-7571
Contact Person: Greg Clites
Date Summary Prepared: May 14, 1999
Name of Device: DeVilbiss 5 Liter Oxygen Concentrator
Common Name: Oxygen Concentrator
Classification Name: Portable Oxygen Generator
Legally Marketed Device to which Equivalence is Claimed: DeVilbiss 5 Liter Oxygen Concentrator*
Manufacturer: Sunrise Medical HHG Inc Respiratory Products Division 100 DeVilbiss Drive P.O. Box 635 Somerset, PA 15501-0635
Predicate Device 510(k) Number: K953815
* The original FDA 510(k) submittal name for the predicate device was the "5 LPM Oxygen Concentrator".
### Device Description:
The revised DeVilbiss 5 Liter Oxygen Concentrator is a 1-5 lpm, pressure swing adsorption, oxygen concentrator based on molecular sieve technology. Room air is drawn into the concentrator via a piston style compressor. The air then passes through a series of filters that remove dust, bacteria, and other particulates. A pneumatic valve directs air into one of the two sieve beds. Nitrogen is adsorbed in the bed as the pressure increases while oxygen flows through, thereby producing an enriched oxygen product for the patient. Simultaneously in the other bed, nitrogen is desorbed as the pressure decreases and is exhausted into the atmosphere. A momentary intermediate pneumatic sequence ties the beds together with the exhaust blocked for an enhanced nitrogen purge. The cycle continues, providing a continuous flow of oxygen at a purity of 93% +/-3% to the patient. The concentrator is available with or without an oxygen monitoring device.
#### Intended Use of the Device:
The revised DeVilbiss 5 Liter Oxygen Concentrator is designed to provide supplemental low flow oxygen therapy for patients suffering from COPD, cardiovascular disease, and lung disorders. The oxygen concentrator is used in home type environments: homes, patient care facilities, etc.
{1}------------------------------------------------
#### Technological Characteristics:
The key technological characteristic differences between the revised DeVilbiss 5 Liter Oxygen Concentrator and the predicate DeVilbiss 5 Liter Oxygen Concentrator are summarized below:
The operating principle of the single pneumatic valve revised DeVilbiss 5 Liter Oxygen Concentrator is the same as the dual pneumatic valve predicate DeVilbiss 5 Liter Oxygen Concentrator, with the revised concentrator operates with a timed cycle and the predicate concentrator cycles based on the monitoring of system pressure. Operating principle:
Room air is drawn into the pressure swing adsorption molecular sieve technology oxygen concentrator by a compressor. Exiting the compressor, the air enters a four-way directional valve and is directed into one of two cylinders containing molecular sieve material. Nitrogen is adsorbed in the bed being pressurized, while oxygen flows through, thereby producing an enriched oxygen product for the patient. The other bed is simultaneously being depressurized, cleansing the bed of nitrogen. The process is alternated between the two beds assuring the patient a continuous supply of oxygen. The additional predicate concentrator valve serves to enhance the nitrogen purge with the UOP 5A-MG molecular sieve material to meet the required performance specification. The revised concentrator accomplishes this enhanced nitrogen purge of UOP Oxysiv-5 molecular sieve material via a momentary intermediate valve position, enabling the revised concentrator to meet the required performance specification. Both UOP 5A-MG and Oxysiv-5 are non-hazardous and non-toxic.
The predicate DeVilbiss 5 Liter Oxygen Concentrator cycles based on a fixed pressure setting detected in the oxygen accumulator tank. When a pc board mounted pressure transducer detects the set pressure in the accumulator tank, the microprocessor shifts the pneumatic valving. The revised concentrator cycles based on a fixed cycle time. The microprocessor shifts the pneumatic valving according to this fixed time.
While the method of shifting the valve to cycle the system is different, the operating principle of the pressure swing adsorption concentrator is the same for both the revised DeVilbiss 5 Liter Oxygen Concentrator and the predicate device. The oxygen purity requirement to the patient is satisfied for both products, hence there is no effect on safety and effectiveness.
The revised and predicate concentrator electronics have been designed with the same microprocessor. Since the functionality of the electronics has been modified for the revised concentrator with implementation of the timed cycle and an accompanying decrease in alarms, the processor is different. With the exception of the No Power and the Low Pressure alarms, the alarm features (High Pressure, Long Cycle, and No Flow) have been removed from the revised concentrator. New software documentation will be generated accordingly. A high pressure indication is most directly tied to the safety of the patient. This will mechanically be maintained via the use of a pressure relief valve on the revised concentrator. The Long and Short Cycle alarms are not necessary on the revised concentrator since it is controlled by a fixed cycle time The High Pressure, Long Cycle, and Short Cycle alarms shut the predicate concentrator down in the event that one of these conditions was detected by the software, in an effort to protect the concentrator from deterioration, but are not directly tied to the oxygen performance. Hence, the modified software and accompanying reduction in the patient alert alarms from the revised DeVilbiss 5 Liter Oxygen Concentrator do not present new issues of concentrator safety or effectiveness.
#### Performance Data:
The key performance descriptor for the oxygen concentrator devices is the oxygen purity performance at 1-5 lpm of oxygen output. The theoretical maximum oxygen purity which is capable from a molecular sieve concentrator is 95.5%, and the commonly accepted minimum level in the concentrator market is 90%. Based on tests performed, both the revised DeVilbiss 5 Liter Oxygen Concentrator and the predicate concentrator meet this requirement.
Conclusion: The oxygen purity levels for the revised and predicate concentrators are equivalent, hence there is no effect on safety and effectiveness.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the words "Public Health Service" in bold, black font. The words are stacked on top of each other, with "Public Health" on the top line and "Service" on the bottom line. The text appears to be extracted from a document or sign.
Image /page/2/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, often used as a symbol of medicine. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus symbol.
NOV -3 1999 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Greq Clites Sunrise Medical HHG Inc. Respiratory Products Division 100 DeVilbiss Drive P.O. Box 635 Somerset, PA 15501-0635
Re: K991722 DeVilbiss 5 Liter Oxygen Concentrator Regulatory Class: II (two) Product Code: 73 CAW Dated: August 31, 1999 Received: September 1, 1999
Dear Mr. Clites:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to leqally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in In addition, FDA may publish further announcements regulatory action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 - Mr. Greg Clites
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
feccions. Hwacteristanse for,
Ceila M. Witten, Ph.D., M.D.
Acting Director
Division of Cardiovascular,
Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 of 1
510(k) Number (if known): __ K991722
Device Name: DeVilbiss 5 Liter Oxygen Concentrator
Indications For Use:
The DeVilbiss 5 Liter Oxygen Concentrator is intended for use as an oxygen concentrator to provide supplemental low flow oxygen therapy in the home, nursing homes, patient care facilities, etc.
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
JoAu Westerman
(Division Sign-Off)
Division of Cardiovascular, Respiratory,
and Neurological Devices
510(k) Number K971722
Prescription Use x (Per 21 CFR 801.109) Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.