The DePuy Rockwood Clavicle Pin is intended to be used to repair an acute fracture, mal-union or non-union of the clavicle.
Device Story
The DePuy Rockwood Clavicle Pin is a metallic bone fixation fastener used by orthopedic surgeons to stabilize clavicle fractures. The device consists of a stainless steel pin with machine threads on one end (featuring a trocar point) and cancellous threads on the other, provided pre-assembled with two locking nuts. During surgery, the pin is inserted into the bone; one nut is tightened to compress the fracture site and prevent migration, while the second nut locks the first in place. The trocar point design allows for either medial or lateral insertion. By providing rigid internal fixation, the device facilitates bone healing in acute fractures, mal-unions, or non-unions, potentially reducing recovery time and improving clinical outcomes for patients.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on design, material, and intended use similarities to legally marketed predicate devices, supported by literature documenting the use of similar pin designs in clavicle fracture treatment.
Technological Characteristics
Constructed from 316L stainless steel. Features a threaded pin design with a trocar point and two locking nuts. Available in four diameters (2.5, 3.0, 3.8, 4.5mm) and a single length (152mm). Mechanical fixation device; no energy source, software, or connectivity.
Indications for Use
Indicated for patients requiring surgical repair of acute clavicle fractures, mal-unions, or non-unions.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
JUL 14 1999
## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
| SPONSOR OF THIS 510(K): | DePuy, Inc.<br>a Johnson & Johnson company<br>700 Orthopaedic Drive<br>P.O. Box 988<br>Warsaw, Indiana 46581-0988 |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| MANUFACTURER: | DePuy OrthoTech™<br>1905 North McArthur Drive<br>Tracy, California 95376 |
| 510(K) CONTACT: | Sally Foust<br>DePuy Orthopaedics, Inc.<br>Sr. Regulatory Associate<br>(219) 372-7455; FAX (219) 267-7098<br>E-mail: sfoust2@dpyus.jnj.com |
| TRADE NAME: | DePuy® Rockwood™ Clavicle Pin |
| COMMON NAME: | Fracture Pin |
| CLASSIFICATION: | Class II per 888.3040, Smooth or threaded metallic bone<br>fixation fastener |
| DEVICE CODE: | 87 JDW |
| EQUIVALENT DEVICES: | DePuy Hagie & Modified Hagie Pins (pre-enactment)<br>DePuy Tibia Bolt (pre-enactment)<br>Onyx Medical Corp. (K903258) and O'Tec Corp. (K905347)<br>Hagie Pins<br>Syntec-Trichung Knowles Pin (K983757) |
### DEVICE DESCRIPTION AND INTENDED USE:
The DePuy Rockwood Clavicle Pin consists of three sub-components, a pin and two nuts, and is provided pre-assembled. The clavicle pin has machine threads on one end and cancellous threads on the other end. The clavicle pin has a trocar point on the machine thread end and two nuts on the cancellous thread end. The pin is available in four diameters, 2.5, 3.0, 3.8 and 4.5mm, in one length, 152mm. The two locking nuts have slightly varying outer diameters. One nut is tightened on the pin to compress the fracture and prevent pin migration. The second nut is used to lock the first nut in place. The two sub-component nuts are also available separately.
The DePuy Rockwood Clavicle Pin is intended to be used to repair an acute fracture, mal-union or non-union of the clavicle.
{1}------------------------------------------------
### BASIS OF SUBSTANTIAL EQUIVALENCE:
The subject Rockwood Clavicle Pin is similar in design, material and intended use to the pre-enactment DePuy Hagie/modified Hagie Pins. All are trocar pointed, threaded pins manufactured from stainless steel that are indicated for use as the bone fixation fastener in bone-to-bone fixation of fractures. One difference between the Rockwood Clavicle Pin and the Hagie Pins is that the Rockwood is designed with the trocar point located on the machine thread end of the pin whereas the pre-enactment Hagie Pins have the trocar point on the cancellous thread end of the pin. This design of the Rockwood Clavicle Pin provides the surgeon with a clavicle pin that can be inserted from either a medial or lateral direction whereas the pre-enactment Hagie pins can only be inserted from a lateral direction.
The subject Rockwood Clavicle Pin is also similar in design, material and intended use to the preenactment DePuy Tibia Bolt, FDA cleared Hagie Pins (K903258, K905347) and Syntec-Trichung Knowles Pin (K983757). All are trocar pointed, threaded pins manufactured from stainless steel that are indicated for use as the bone fixation faster in bone-to-bone fixation of fractures.
Published and unpublished literature documents the use of Hagie, modified Hagie, and Knowles pins in the treatment of clavicle fractures thereby justifying the use of the very similarly designed Rockwood Clavicle Pin in the clavicle.
Based on similarities of design, materials and intended use, DePuy believes that the subject Rockwood Clavicle Pin is substantially equivalent to pre-enactment devices and FDA cleared devices currently on the market.
| | Rockwood<br>Clavicle Pin | Pre-Enactment<br>Hagie &<br>Modified Hagie<br>Pin | Pre-<br>Enactment<br>Tibial Bolt | K903258<br>K905347<br>Hagie Pins | K983757<br>Syntec-<br>Trichung<br>Knowles Pin |
|---------------|--------------------------|---------------------------------------------------|----------------------------------|----------------------------------|-----------------------------------------------|
| Material | 316L SS | 316L SS | 316L SS | 316L SS | 316L SS |
| Use | FX Fixation | FX Fixation | FX Fixation | FX Fixation | FX Fixation |
| Fracture Site | Clavicle | Multi | Tibia | Multi | Multi |
| Product Code | 87 JDW | 87 JDW | 87 JDW | 87 JDW | 87 JDW |
| Threaded | Yes | Yes | Yes | Yes | Yes |
| Trocar Point | Yes | Yes | Yes | Yes | Yes |
| Sizes | 5 | 2 | 1 | NA | 2 |
| Nuts | 2 | 1 | 2 | NA | NA |
The following table summarizes the similarities:
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three lines forming its body and head. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 14 1999
Ms. Sally Foust Senior Regulatory Associate DePuy Orthopaedics, Inc. P.O. Box 988 700 Orthopaedic Drive Warsaw, Indiana 46581-0988
Re: K991649
> Trade Name: DePuy Rockwood Clavicle Pin Regulatory Class: II Product Code: JDW Dated: May 12, 1999 Received: May 13, 1999
Dear Ms. Foust:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (OS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 -- Ms. Sally Foust
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
510(k) Number (if known)_
Device Name: DePuy® Rockwood™ Clavicle Pin
Indications for Use:
The DePuy Rockwood Clavicle Pin is intended to be used to repair an acute fracture, mal-union or nc union of the clavicle.
Concurrence of CDRH, Office of Device Evaluation
Prescription Use_ X (Per 21 CFR 801.109)
OR Over-The-Counter Use
(Division Sign-Off)
Division of General Restorative Devices
510(k) Number K991649
-
-------------------------------------------------------------------------------
# 000002
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.