K991567 · Shelhigh, Inc. · FTM · Jul 14, 1999 · General, Plastic Surgery
Device Facts
Record ID
K991567
Device Name
SHELHIGH NO-REACT TISSUE REPAIR PATCH/UROPATCH.
Applicant
Shelhigh, Inc.
Product Code
FTM · General, Plastic Surgery
Decision Date
Jul 14, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Tissue repair patch, for the surgical repair of soft tissue deficiencies which include: defects of the abdominal and thoracic wall, gastric banding , rectal and vaginal prolapse, urethral sling, reconstruction of the pelvic floor, and hernias, or defects of the diaphragm, femoral, incisional, inquinal, lumbar, and umbilical hernia.
Device Story
UroPatch is a surgical tissue repair patch composed of glutaraldehyde-fixed bovine pericardium treated with the No-React detoxification process. It is stored in 2% benzyl alcohol. The device is soft, pliable, and provides tensile strength and suture retention for soft tissue repair. It is intended for use by surgeons in clinical settings for the reconstruction of pelvic floor, hernia repair, and other soft tissue deficiencies. The patch acts as a permanent implant to support or replace deficient tissue, facilitating surgical repair and potentially improving patient outcomes by providing a biocompatible scaffold.
Clinical Evidence
Animal study conducted to compare the Shelhigh No-React patch against conventional glutaraldehyde-treated patches. Results demonstrated that the No-React patch exhibits a higher level of biocompatibility and cytocompatibility compared to the conventional treatment.
Technological Characteristics
Material: Glutaraldehyde-fixed bovine pericardium. Treatment: No-React detoxification process. Storage: 2% benzyl alcohol. Properties: Soft, pliable, high tensile strength, suture retention. Form factor: Membrane/patch. Energy source: None (passive implant).
Indications for Use
Indicated for patients requiring surgical repair of soft tissue deficiencies, including abdominal/thoracic wall defects, gastric banding, rectal/vaginal prolapse, urethral sling, pelvic floor reconstruction, and various hernias (diaphragm, femoral, incisional, inguinal, lumbar, umbilical).
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
K991567
# Class II 510(K) Summary Shelhigh No-React® UROPATCH™
This summary of the 510(k) information is being submitted as required by section 807.92(a).
### 1. Proprietary and Common Name:
Shelhigh No-React® Tissue repair patch; Proprietary name:
UroPatch™
Mesh , surgical Common name:
# II. Regulatory Class:
Class II device
### III. Intended Use
Tissue repair patch, for the surgical repair of soft tissue deficiencies which include: defects of the abdominal and thoracic wall, gastric banding , rectal and vaginal prolapse, urethral sling, reconstruction of the pelvic floor, and hernias, or defects of the diaphragm, femoral, incisional, inquinal, lumbar, and umbilical hernia.
#### IV. Product_Description
The Shelhigh No-React® Tissue repair patch/ UroPatch™ is made of glutaraldehyde fixed boyine pericardium. rinsed with the detoxification process No-React®. It is stored in 2% Benzyl alcohol like the Shelhigh No-React® pericardial patch. The material exhibits good tensile strength, shrink temperature, excellent biocompatibility and suture retention. It is soft and pliable making it convenient to implant. Bovine pericardial material has been used successfully as a tissue patch for pericardial closure. Glutaraldehyde processed bovine pericardium has a long history of success as a permanently implanted material.
### V. Substantial Equivalence
The Shelhigh No-React® Tissue repair patch/UroPatch™ is equivalent to the Shelhigh No-React® pericardial patch K974914, and K981756 currently manufactured by Shelhigh Inc.
{1}------------------------------------------------
# VI. Comparison with Predicate Device
The Shelhigh No-React® Tissue repair patch/UroPatch™. like the Shelhigh No-React® pericardial patch is a glutaraldehyde cross-linked bovine pericardium membrane which exhibits substantially equivalent physical/mechnical properties as measured by suture retention, tensile strength, and shrink temperature. They both have identical flexibility, wall thickness and both are stored in bezyle alcohol.
# VII.Nonclinical / Animal Tests
An animal study was conducted to evaluate Shelhigh No-React® vs. the conventional glutaraldehyde treated patch, the Shelhigh No-Reac® patch shows higher evel of biocompatibility.
### VIII. Conclusions
The non clinical /Animal testing data showed that the Shelhigh No-React®Tissue repair patch/ UroPatch™. has higher level of cytocompatibility when compared with the conventional glutaraldehyde treated and storage. The UroPatch™ is substantially equivalent to the predicate device.
Page 2 of 2 of Summary
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three horizontal lines extending from its head, representing the department's mission to promote health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle symbol. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 14 1999
Shlomo Gabbay, M.D. Chief Scientific Advisor Shelhigh, Inc. 67 E. Willow Street Millburn, New Jersey 07041
K991567 Re:
> Trade Name: No-React® Tissue Repair Patch/UroPatch Regulatory Class: II Product Code: FTM Dated: May 3, 1999 Received: May 5, 1999
Dear Dr. Gabbay:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 - Shlomo Gabbay, M.D.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
[signature]
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page _1_of __1__1__
510(k) Number (if known): _ (99156 7
Device Name: Shelhigh No-React® Tissue repair patch/UROPATCH™ Indications For Use:
Tissue repair patch, for the surgical repair of soft tissue deficiencies which include: defects of the abdominal and thoracic wall, gastric banding , rectal and vaginal prolapse, urethral sling, reconstruction of the pelvic floor, and hernias, or defects of the diaphragm, femoral, incisional, inguinal, lumbar, and umbilical hernia.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of General Re
| 510(k) Number | K971567 |
|---------------|---------|
|---------------|---------|
| Prescription Use<br>21CFR 801.109) | <div>X</div> | OR | Over-The-Counter Use (Per<br>(Optional Format 1-2-96)) |
|------------------------------------|--------------|----|--------------------------------------------------------|
|------------------------------------|--------------|----|--------------------------------------------------------|
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.