SYNERGY POROUS SIZE 8 HIP STEM

K991485 · Smith & Nephew, Inc. · LPH · Jul 12, 1999 · Orthopedic

Device Facts

Record IDK991485
Device NameSYNERGY POROUS SIZE 8 HIP STEM
ApplicantSmith & Nephew, Inc.
Product CodeLPH · Orthopedic
Decision DateJul 12, 1999
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 888.3358
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Total hip components are indicated for uncemented use only in individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative joint disease (NIDJ) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; old, remote osteomyclitis with an extended drainage-free period; nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; fernoral osteotomy, or Girdlestonc resection; fracture dislocation of the hip; and correction of deformity. The Synergy Porous Size 8 Hip Stem is designed for single use only.

Device Story

Synergy Porous Size 8 Hip Stem is a metallic femoral component for total hip arthroplasty. Device is implanted by orthopedic surgeons in clinical settings to replace damaged hip joints. It functions as a structural prosthesis, designed for uncemented fixation. It is compatible with modular cobalt chrome or ceramic femoral heads via a 12/14 taper. The device provides mechanical stability and restores joint function for patients with degenerative or traumatic hip conditions. It is intended for single use only.

Clinical Evidence

No clinical data provided; substantial equivalence is based on technological characteristics and design similarities to legally marketed predicate devices.

Technological Characteristics

Material: Titanium alloy (Ti-6Al-4V, ASTM F1472). Design: Porous tapered hip stem. Compatibility: Modular cobalt chrome or ceramic femoral heads with 12/14 taper. Fixation: Uncemented. Single-use only.

Indications for Use

Indicated for individuals undergoing primary and revision hip surgery for trauma, inflammatory joint disease (e.g., rheumatoid arthritis), noninflammatory degenerative joint disease (e.g., osteoarthritis), avascular necrosis, traumatic arthritis, slipped capital epiphysis, fused hip, pelvic fracture, diastrophic variant, remote osteomyelitis, nonunion, femoral neck/trochanteric fractures, femoral osteotomy, Girdlestone resection, fracture dislocation, or deformity correction.

Regulatory Classification

Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K991485 ## 510(k) Summary Submitter's name: Submitter's address: Submitter's telephone number: Contact person: Date summary prepared: Trade or proprietary device name: Common or usual name: Classification name: Smith & Nephew, Inc. 1450 Brooks Road, Memphis, TN 38116 901-399-6487 David Henley April 23, 1999 Synergy Porous Size 8 Hip Stem Syncrgy Forous ont - Porous Tapered Hip Stem 21 CFR 888.3358 hip joint metal/polymer/metal semi-21 CFR 866.5578 mp John mean posted uncemented prosthesis-Class II 87LPH ## Substantially Equivalent Legally Marketed Devices - Global Taper "Tapered" Hip Stem Smith & Nephew . - Revision (Echelon) Hip Stem Smith & Nephew . � - Versys® Hip Stem Zimmer . - Perfecta IMCTM Hip Stem Wright Medical � - AML® Hip Stem DePuy ● #### Device Description Device Description The Synergy Porous Size 8 Hip Stem is manufactured from titanium material (Ti-6AI-4V, ASTM F1472) The Synergy Porous Size o Hip Stein is nambidered from in them. and is designed for use with existing cobalt chrome or ceramic modular femoral heads with a 12/14 taper. ## Device Intended Use Device Intended Use I of a mp components are indicated ancements associations hips damaged as a result of trauma, surgery where other treatments or devices have failed in rehabilitation inint di surgery where other treatments of devices have in rendemannatory degeneration in disease inflammatory Joint disease such as medification in the recosis, traumatic arthoritos (NIDD) of any of Its composite uraginses such as estocal director of tremote of the province of the provimal silpped capital epiphysis, fused inf, fracture of the period tractures of the proximal an extended drainage-free period; nonunion, femoral neck fractures fractores of the pro all extended dramage-free period, nonanageable using other techniques; femoral osteotomy, or femal with nead in Forrement dislocation of the hip; and correction of deformity. The Synergy Porous Size 8 Hip Stem is designed for single use only. ## Technological characteristics: The Synergy Porous Size 8 Hip Stem is similar to the legally marketed devices listed above. All of these I he Synergy I Grous Gree of Thp Seem is Smith, as with of the Synergy Porous Size 8 Hip Stem and have the same technological characteristics. #### Performance characteristics: x croomance that the Synergy Porous Size 8 Hip Stem is substantially equivalent to legally marketed devices. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The eagle is positioned to the right of a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a sans-serif font. The text is arranged to follow the curve of the circle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUL 1 2 1999 Mr. David Henley Clinical/Regulatory Affairs Specialist Smith & Nephew, Inc. 1450 Brooks Rd. Memphis, Tennessee 38116 Re: K991485 Trade Name: Synergy porous Size 8 Hip Stem Regulatory Class: II Product Code: LPH and LZO Dated: April 27, 1999 Received: April 28, 1999 Dear Mr. Henley: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commercc prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {2}------------------------------------------------ Page 2 -- Mr. David Henley This letter will allow you to begin marketing your device as described in your 510(k) promated I his letter will anow you to begin marketing your device to a legally marketed nothication. The PDA inding of succemar vqur device and thus, permits your device to procced to the market. If you desire specific advice for your device on our labeling regulation (21 CFF Part 801 and II you desire specific advice for your dones), please contact the Office of Compliance at additionally 609.10 for in vitty diagnostions on the promotion and advertising of your device, (301) 594-4037. Addinonally, for questions on assisted on associated the regulation prease contact the Office or Comphanee at (301) of the other (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Information on your responsionnes ander the rice may of -2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications Statement Synergy Porous Size 8 Hip Stem Total hip components are indicated for uncemented use only in individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative joint disease (NIDJ) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; old, remote osteomyclitis with an extended drainage-free period; nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; fernoral osteotomy, or Girdlestonc resection; fracture dislocation of the hip; and correction of deformity. signature Orvision Sid vision of General Re 510(k) Number Prescription Use (Per 21 CFR 801.109) Syncrgy Porous Sz. 8 Hip Stem CONFIDENTIAL
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...