EXCEL GE BLOOD GLUCOSE TEST STRIPS & CODE STRIP

K991480 · Selfcare, Inc. · CGA · Sep 17, 1999 · Clinical Chemistry

Device Facts

Record IDK991480
Device NameEXCEL GE BLOOD GLUCOSE TEST STRIPS & CODE STRIP
ApplicantSelfcare, Inc.
Product CodeCGA · Clinical Chemistry
Decision DateSep 17, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2

Intended Use

Excel™ GE Blood Glucose Test Strips quantitatively measure glucose in fresh capillary whole blood and give results that are comparable to laboratory methods. They are intended for in vitro diagnostic (external) use by people with diabetes for self-monitoring of blood glucose. The Excel™ GE Blood Glucose Test Strips are designed to be used with Glucometer Elite® Blood Glucose Meters only.

Device Story

Excel™ GE Blood Glucose Test Strips are electrochemical sensors for quantitative glucose measurement in fresh capillary whole blood. Used by patients with diabetes at home for self-monitoring of blood glucose levels; also used by professionals. Strips are inserted into Glucometer Elite® Blood Glucose Meters; the meter performs an amperometric measurement of the blood sample. Results are displayed on the meter screen, allowing patients and clinicians to monitor diabetes control. The device provides quantitative data comparable to laboratory methods, aiding in clinical decision-making regarding diabetes management.

Clinical Evidence

Clinical and laboratory studies compared Excel™ GE Test Strips to YSI laboratory methods using capillary whole blood. Accuracy study included 128 patients with diabetes across three clinical centers. Professional user accuracy (n=252) showed r=0.9877-0.9949; lay user accuracy (n=128) showed r=0.9883-0.9941. Precision study (n=20 per level) across 4 glucose levels demonstrated CVs ranging from 1.8% to 3.0%. Results confirm performance equivalence to the predicate.

Technological Characteristics

Electrochemical test strips for amperometric glucose measurement. Designed for use with Glucometer Elite® Blood Glucose Meters. In vitro diagnostic use. No specific material standards or software architecture details provided.

Indications for Use

Indicated for people with diabetes for self-monitoring of blood glucose using fresh capillary whole blood. Designed for use with Glucometer Elite® Blood Glucose Meters.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K991480 # Section 5 # 510(k) SUMMARY (Summary of Safety and Effectiveness) ### Submitted by: Carol Adiletto Director of Clinical Affairs Selfcare, Inc. 200 Prospect Street Waltham, MA 02453-3457 USA Phone: (781) 647-3900 (781) 647-3939 Fax: ### Contact Person: Carol Adiletto Phone (781) 647-3900 x124 ### Summary Prepared: March 27, 1999 # Name of the device: Excel™ GE Blood Glucose Test Strips and Code Strip ### Classification name(s): Excel™ GE Blood Glucose Test Strips are a Class II device (21 CFR § 862.1345) for home use # Classification of predicate device(s): Excel™ GE Blood Glucose Test Strips and Code Strip are not materially different from the predicate device, Glucometer Elite® Blood Glucose Test Strips and Code Strip. The predicate device is distributed by Bayer and was cleared for use in the United States by K964630, K951537 and K924499. ## Description of the device/intended use(s): Excel ™ GE Blood Glucose Test Strips quantitatively measure glucose in fresh capillary whole blood and give results that are comparable to laboratory methods. They are intended for in vitro diagnostic (external) use by people with diabetes for self-monitoring of blood glucose. The Excel™ GE Blood Glucose Test Strips are designed to be used with Glucometer Elite® Blood Glucose Meters only. {1}------------------------------------------------ # Statement of How the Technological Characteristics of the Device Compare to the Predicate device: The technological characteristics of both devices are the same since they are electrochemical devices that quantitatively measure glucose amperometrically and have the same safety and effectiveness. They have the same intended use, i.e. use by people with diabetes for self-monitoring of blood glucose. The new device is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood. The Excel™ GE Blood Glucose Test Strips and Code Strip are intended for use outside the body (in vitro diagnostic use) by diabetics at home as an aid to monitor the effectiveness of diabetes control. ## Summary of Performance Data: Laboratory and clinical studies demonstrate that the Excel™ GE Blood Glucose Test Strips and Code Strip for use with Glucometer Elite® Blood Glucose Meters provides equivalent performance to the Glucometer Elite® Blood Glucose Test Strips and Code Strip for use with Glucometer Elite® Blood Glucose Meters. Precision: A laboratory study of 20 samples (n=20), using Excel™ GE Blood Glucose Test Strips and Glucometer Elite® glucose meters to test whole blood samples at 4 different glucose levels (Glucose Level), showed the following performance: | Precision | Elite® Model 3901A | | | | Elite® Model 3901M | | | | |------------------|--------------------|----------------------------|---------------|-----------|--------------------|----------------------------|---------------|-----------| | Glucose<br>Level | n | mean<br>glucose<br>(mg/dL) | SD<br>(mg/dL) | CV<br>(%) | n | mean<br>glucose<br>(mg/dL) | SD<br>(mg/dL) | CV<br>(%) | | 1 | 20 | 47 | 1.2 | 2.6 | 20 | 43 | 1.2 | 2.7 | | 2 | 20 | 77 | 2.3 | 3.0 | 20 | 83 | 2.2 | 2.7 | | 3 | 20 | 145 | 3.2 | 2.2 | 20 | 160 | 3.6 | 2.2 | | 4 | 20 | 360 | 6.4 | 1.8 | 20 | 331 | 7.0 | 2.0 | | Accuracy | Elite® Model 3901A | Elite® Model 3901M | |---------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------| | Accuracy of Professional users<br>compared to YSI using capillary<br>whole blood on 128 patients with<br>diabetes at three clinical centers | n = 252<br>y = 0.91x + 0.18<br>r = 0.9877<br>Sy.x = 11.45<br>Range = 51 - 374 mg/dL | n = 253<br>y = 0.93x + 1.24<br>r = 0.9949<br>Sy.x = 7.55<br>Range = 51 - 374 mg/dL | | Accuracy of Lay users compared<br>to YSI using capillary whole<br>blood on 128 patients with<br>diabetes at three clinical centers | n = 128<br>y = 0.92x + 6.99<br>r = 0.9883<br>Sy.x = 11.38<br>Range = 51 - 374 mg/dL | n = 128<br>y = 0.95x + 3.92<br>r = 0.9941<br>Sy.x = 8.31<br>Range = 51 - 374 mg/dL | {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines forming its body and head. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 SEP 17 1999 Ms. Carol Adiletto Director of Clinical Affairs Selfcare, Inc. 200 Prospect Street Waltham, Massachusetts 02453-3457 Re: K991480 > Trade Name: Excel™ GE Blood Glucose Test Strips and Code Strip Regulatory Class: II Product Code: CGA Dated: August 9, 1999 Received: August 11, 1999 Dear Ms. Adiletto: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ # Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Steven Putman Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Section 4 #### Indications for Use Statement 5991480 510(k) Number (if known): Device Name: Excel™ GE Blood Glucose Test Strips and Code Strip ### Indications for Use: Excel™ GE Blood Glucose Test Strips quantitatively measure glucose in fresh capillary whole blood and give results that are comparable to laboratory methods. They are intended for in vitro diagnostic (external) use . by people with diabetes for self-monitoring of blood glucose. The Excel™ GE Blood Glucose Test Strips are designed to be used with Glucometer Elite® Blood Glucose Meters only. Cooper (Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number K99480 (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.019) OR Over-The-Counter Use
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%