K991376 · Disetronic Medical Systems · LZG · Jun 15, 1999 · General Hospital
Device Facts
Record ID
K991376
Device Name
DAHEDI INSULIN INFUSION PUMP
Applicant
Disetronic Medical Systems
Product Code
LZG · General Hospital
Decision Date
Jun 15, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The DAHEDI Insulin Infusion Pump is intended for the controlled delivery of insulin as prescribed by a physician. It is indicated for patients with insulin dependent diabetes Mellitus who do not have optimum blood glucose control on conventional insulin injection therapy. Patients for insulin pump therapy must be highly motivated to perform self glucose monitoring on a frequent and regular basis, as well as adhere to a proper diet and exercise regiment. Patients must be capable of operating the pump. They must also have access to the educational training, support, and follow-up of health care professional experienced in insulin pump therapy.
Device Story
Ambulatory, battery-operated insulin infusion pump; treats insulin-dependent diabetes mellitus. Device uses motor-driven piston to displace insulin from reservoir through subcutaneous catheter; delivers basal rates and boluses. User inputs dosage parameters via pump buttons; monitors therapy via pump interface. Optional IR adapter enables data transfer of 42-day history to IBM-compatible PC. Used by patients in home/daily settings; requires physician prescription. Benefits include precise, continuous insulin delivery to improve glycemic control. Device is compact, watertight, and shock-resistant.
Clinical Evidence
Bench testing only. Device tested for safety and performance in accordance with IEC 601-2-24 (infusion pump safety) and IEC 601-1/601-1-2 (general safety and electromagnetic interference). Results demonstrated that system specifications and functional requirements were met.
Technological Characteristics
Ambulatory, battery-operated pump. Proprietary hardwired digital control chip for functional/safety parameters. Compatible with standard female luer lock subcutaneous infusion sets. Watertight, shock-resistant housing. Connectivity via optional IR adapter for PC data transfer. Tested to IEC 601-2-24 and IEC 601-1 standards.
Indications for Use
Indicated for patients with insulin-dependent diabetes mellitus requiring insulin pump therapy due to suboptimal blood glucose control on conventional injections. Patients must be motivated, capable of self-monitoring glucose, adherent to diet/exercise, and have access to professional training and support.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
Disetronic H-Tron Plus V 100 Insulin Infusion Pumps (K973044)
Related Devices
K022831 — MODIFICATION TO DISETRONIC D-TRONPLUS INSULIN INFUSION PUMP · Disetronic Medical Systems AG · Sep 11, 2002
K021725 — DISETRONIC D-TRON INSULIN INFUSION PUMP · Disetronic Medical Systems AG · Aug 2, 2002
K023471 — DISETRONIC H-TRONPLUS INSULIN INFUSION PUMP · Disetronic Medical Systems AG · Oct 29, 2002
K994186 — DISETRONIC D-TRON INSULIN INFUSION PUMP · Disetronic Medical Systems · Dec 30, 1999
K043000 — MODIFICATION TO D-TRONPLUS · Disetronic Medical Systems AG · Dec 1, 2004
Submission Summary (Full Text)
{0}------------------------------------------------
JUN 1 5 1999
# 510(K) SUMMARY
## 9.0 Summary of Safety and Effectiveness
K991376
- 9.1 Trade/Proprietary Name: DAHEDI Insulin Infusion Pump
- Insulin Infusion Pump and Accessories 9.2 Common/Usual Name:
- Infusion Pump 9.3 Classification Name:
- 9.4 Substantial Equivalence: The DAHEDI Insulin Infusion Pumps are substantially equivalent to the Disetronic H-Tron Plus V 100 Insulin Infusion Pumps (K973044).
#### 9.5 Device Description
The DAHED! Insulin Infusion Pump is a very small, ambulatory, battery operated portable pump that is designed for the treatment of Insulin Dependent Diabetes Mellitus. The DAHEDI Insulin Infusion Pump infuses small amounts of insulin through a catheter into the patient via subcutaneous administration twenty-four hours a day. It is compact, watertight and shock resistant to maximize patient convenience. The labeling provided in the appendices contain detailed pictorial representations, descriptions and instructions that are adequate to facilitate evaluation of the nature and operation of the device.
The pump uses a commercially available insulin reservoir that holds reqular insulin into which the pump drives a piston rod that displaces precise amounts of insulin at exact times throughout the day. Insulin is delivered from the insulin reservoir of the pump to the body through a thin, plastic tube called an "infusion set." The infusion set has a small needle or flexible catheter that the user inserts under the skin (most commonly in the subcutaneous tissue of the abdomen). Insulin passes through the infusion set into a needle or a soft Teflon cannula into the body. The pump is compatible with all commercially available subcutaneous insulin infusion sets with female luer lock connectors.
The insulin reservoirs and catheters are sterile and single use (disposable). The pump itself is not sterile as it does not come in contact with the insulin.
Insulin delivery is accomplished through the reservoir piston mechanism. The piston is advanced by means of a motor driving a piston forward into the cartridge. The frequency of the motor revolution is controlled by the pump according to the information entered into the pump by the user or his care provider. The rate of the basal infusion, initiation of boluses, any dosage limitations or modifications can be simply entered using the buttons on the pump. Additionally, the user controls infusion start, infusion stop and has access to information important to the proper monitoring of the pump and the therapy. An optional IR adapter allows the user to send 42 days of history into a standard IBM compatible Personal computer.
#### 9.6 Intended Use
The Indications for use and intended use are the same as those for the predicate device.
{1}------------------------------------------------
### 9.0 Summary of Safety and Effectiveness Continued
#### 9.7 Technological Characteristics
The DAHEDI Insulin Infusion Pump incorporates a proprietary digital control chip that has been hardwired to operate all functional and safety control parameters. This function has been accomplished by software resident in a microprocessor in other equivalent devices. All other functional and technological characteristics are similar to the predicate device.
#### 9.8 Performance Data
The DAHEDI Insulin Infusion Pump has been designed and tested in accordance with IEC 601-2-24 of the International Electrotechnical Commission Technical Committee No. 62: Electrical Equipment in Medical Practice Sub-Committee 62D: Electromedical Equipment -Draft - Part 2: Particular requirements for safety of infusion pumps and controllers. IEC 601-2-24 incorporates the requirements of IEC 601-1 for all general safety requirements including IEC 601-1-2 Electromagnetic Interference (EMI).
Disetronic has adhered to Good Design, Manufacturing and Quality Assurance procedures. All test results demonstrate that the system specifications and functional requirements were met.
#### 9.9 Conclusion
Based on the functional comparison and the performance and safety testing, Disetronic has determined that the DAHEDI Insulin Infusion Pump is substantially equivalent to the Disetronic H-TRON Plus V 100 Insulin Infusion Pump.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract design of three human profiles facing right, with flowing lines above and below them. The design is in black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 15 1999
Disetronic Medical Systems, Incorporated c/o Lee Leichter P/L Biomedical 7690 Cameron Circle Fort Meyers, Florida 33912
Re: K991376 DAHEDI Insulin Infusion Pump Trade Name: Regulatory Class: II Product Code: LZG Dated: April 21, 1999 Received: April 21, 1999
Dear Lee Leichter:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister.
{3}------------------------------------------------
Page 2 - Lee Leichter
Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to ene regaracion choicess, "https://www.bbcananty-by-sother general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fdaggov/cdrh/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Directbr Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
14991376 510(k) File Number:
DAHEDI Insulin Infusion Pump Device Name:
The DAHEDI Insulin Infusion Pump is intended for the controlled Indications For Use: delivery of insulin as prescribed by a physician.
> It is indicated for patients with insulin dependent diabetes Mellitus who do not have optimum blood glucose control on conventional insulin injection therapy. Patients for insulin pump therapy must be highly motivated to perform self glucose monitoring on a frequent and regular basis, as well as adhere to a proper diet and exercise regiment. Patients must be capable of operating the pump. They must also have access to the educational training, support, and follow-up of health care professional experienced in insulin pump therapy.
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.19)
(Division Sign-Off)
Division of Dental, Infection Control,
and General Hospital Devices
510(k) Number***_****_*
OR
Over-The-Counter Use***_****_*
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.