Retrospective clinical performance data from IVF clinics in Norway; Clinical performance data from the Carl von Linne Clinic in Sweden
The sponsor provided retrospective clinical data (live birth rates per embryo replacement) from multiple European IVF clinics to demonstrate that the device performs well in routine clinical settings and compares favorably to national averages.
Retrospective analysis of IVF clinic performance in Norway and Sweden; Retrospective observational analysis of clinical outcomes; Follow-up/Duration: 1996-1997; Study Period: 1996-1997
Patients undergoing ICSI procedures at specific IVF clinics in Norway and Sweden; Number of Sites: 4 (3 in Norway, 1 in Sweden)
National average success rates
Live birth rates per embryo replacement
Indications for Use
For use in assisted reproduction techniques, (ART), such as IVF, GIFT or similar in vitro procedure to aid in facilitating pregnancy. Medi-Cult Hyaluronidase is for the removal of cumulus and corona radiatacells which must be removed in order to facilitate access to the oocyte and minimize contamination of the injected needle before ICSI.
Device Story
Medi-Cult Hyaluronidase is a ready-to-use HEPES buffered salt solution supplemented with hyaluronidase, synthetic serum replacement (SSR), human serum albumin (HSA), and antibiotics (penicillin and streptomycin). It is used in clinical IVF laboratories by embryologists or trained technicians during assisted reproduction procedures. The solution is applied to oocytes to enzymatically remove cumulus and corona radiata cells; this removal facilitates access to the oocyte for ICSI (Intra Cytoplasmic Sperm Injection) and minimizes needle contamination. The device is intended to aid in facilitating pregnancy. Clinical performance is monitored via live birth rates and laboratory success metrics compared against national or clinic-specific averages.
Clinical Evidence
Clinical evidence consists of retrospective data from European IVF clinics (Norway and Sweden). Data from Norway (1996-1997) reported live birth rates per embryo replacement across three clinics (ranging from 15.9% to 28.6%). Data from the Carl von Linne Clinic in Sweden (1997) reported a 32% birth rate per embryo replacement, exceeding the national average of 25%. No serious adverse events or product complaints were reported over a 1.5-year period.
Technological Characteristics
HEPES buffered salt solution; contains hyaluronidase (bovine testes derived), Synthetic Serum Replacement (SSR), Human Serum Albumin (HSA), Penicillin, and Streptomycin. Ready-to-use liquid formulation. Quality control testing includes bioburden, filter integrity, sterility, pH, osmolarity, and endotoxin levels.
Indications for Use
Indicated for use in assisted reproduction techniques (ART) including IVF, GIFT, or similar in vitro procedures to facilitate pregnancy by removing cumulus and corona radiata cells from oocytes prior to ICSI.
Regulatory Classification
Identification
Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
Submission Summary (Full Text)
{1}------------------------------------------------
# JUN 2 0 2000
Page 1 of 2
## 510(k) Summary Summary of Safety and Effectiveness Information Supporting a Substantially Equivalent Determination
### (K991334)
#### The product:
-"Medi-Cult Hyaluronidase" Cat.No. 1011
Indications for use:
"Medi-Cult Hyaluronidase" is used in assisted reproduction techniques, (ART), such as IVF, GIFT or similar In Vitro procedure to aid in facilitating pregnancy.
"Medi-Cult Hyaluronidase" is for the removal of cumulus and corona radiatacells which must be removed in order to facilitate access to the oocytes and minimize contamination of the injected needle before ICSL
#### Product formulation:
Sperm Preparation Medium supplemented with Hyaluronidase. It is a HEPES buffered salt solution supplemented with Hyaluronidase. It is ready-to-use and contains Synthetic Serum Replacement (SSR) and Human Serum Albumin (HSA). The Hyaluronidase raw material is derived from Bovine testes in the USA. It contains antibiotics - Penicillin and Streptomycin.
A number of trials have shown that the products performs well, (see Clinical Testing section). A number of publications in peer-reviewed books or journals have presented data using Medi-Cult media. Often more than one procent from Medi-Cult has been used in the studies listed.
Product testing control contents:
- Bioburden, production-test ・
- -Integrity filter testing, production-test
- -Sterility, QC-test
- pH, QC-test
- -Osmolarity, QC-test
- -Endotoxin, QC-test
The culture media from Medi-Cult have been used by many European IVF-units since the end of the 1980's. (Produced and distributed by GEA-Biotech 1987 to 1989 and by Medi-Cult als from 1989). The first Medi-Cult products for use in conjunction with Intra Cytoplasmic Sperm Injection (ICSI) were introduced in 1993. These products were designed in accordance with the specification of the clinic that developed ICSI (Palermo et al. Lancet 1992 340 8810 17-18).
A number of publications in peer- reviewed books or journals have presented data using Medi-Cult products for ICSI. Often more than one product from Medi-Cult has been used in the studies.
The clinical result of a given IVF-clinic will depend on the patient population treated, the clinical procedures, the laboratory routines and on the various culture media used. A comparison of the clinical performance of clinics using only Medi-Cult products for ICSI with other IVP client using similar clinical and laboratory routines will give an indicate on the performance of the Medi-Cult culture media. In countries were we have access to both national data and data from in incul collected clinics, clinics using Medi-Cult products generally performs well.
{2}------------------------------------------------
K991334
Page 2 of 2
The IVF-units in the Nordic countries use similar clinical and laboratory protocols and their patient populations are comparable. The clinical data from most Nordic countries are available. In some countries the performance of individual IVF-clinics are available.
The largest IVF clinics in Norway have been using only Medi-Cult media since the end of the 1980'ties and products from Medi-Cult for ICSL The data collected since the introduction of ICSI in Norway in 1996 are shown below as live birth rates per started cycle. Since all clinics (except one) are using products from Medi-Cult, a meaningful comparison to the national average success rate is not meaningful.
Table. Live birth rates per embryo replacement obtained with ICSI in Norway
| Clinic | 1996 | 1997 |
|--------------------------------------|------|------|
| Dept of Ob & Gyn, Univ. of Trondheim | 18,4 | 22,4 |
| The National Hospital, Oslo | 15,9 | 18,4 |
| Haugesund Hospital | 28,6 | 27,1 |
In Sweden, the Carl von Linne Clinic uses only products from Medi-Cult for ICSI and for culture of gametes and embryos. It is the only Swedish clinic using Medi-Cult products. For the year 1997 their birth rate per embryo replacement was 32% their clinical results were the best in Sweden and well above the national average of 25%.
There have been no registered complaints for the product and no evidence in the last 1.5 year that the product has been the cause of any serious adverse events in connection with its intended use.
Thus based on the Federal Register Notice (Final Rule, Vol. 63, No. 175, page 48429, September 10, 1998), "Obstetric and Gynecologic Devices; Reclassification and Classification of Medical Devices Used for In-Vitro Fertilization and Related Assisted Reproduction Procedures" effective on October 13, 1998 and the supportive clinical data we feel that the safety and effectiveness of the product for its intended use is shown in the present submission.
Prepared and Submitted by:
Ronald G. Leonardi, Ph. D. 12/14/99
Date
Ronald &. Leonardi, Ph. D. President R & R Registrations P.O. Box 262069 San Diego CA 92196 1-619-586-0751
{3}------------------------------------------------
Image /page/2/Picture/15 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
## JUN 2 0 2000
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Medi-Cult A/S c/o Ronald G. Leonardi, Ph.D. President R & R Registrations P.O. Box 262069 San Diego, CA 92196-2069
Re: K991334 Medi-Cult Hyaluronidase Dated: May 22, 2000 Received: May 23, 2000 Requiatory Class: II 21 CFR §884.6180/Procode: 85 MQL
Dear Dr. Leonardi:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the Federal Food, Grug, and Cosmelic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual recistration, listing of devices, good manufacturing and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalion assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notfication submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predication for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for question and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmaldsmamain.html".
Sincerely yours,
Daniel G. Schultz, M.D.
Captain, USPHS Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{4}------------------------------------------------
# B4. Indication for Use Statement
510(k) Number (if know) K991334
Device Name: Medi-Cult Hyaluronidase
Indications for use:
For use in assisted reproduction techniques, (ART), such as IVF, GIFT or similar in vitro procedure to aid in facilitating pregnancy.
Medi-Cult Hyaluronidase is for the removal of cumulus and corona radiatacells which must be removed in order to facilitate access to the oocyte and minimize contamination of the injected needle before ICSI.
## PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|------------------------------------------------------------------------|
| Prescription Use | <span style="text-decoration: overline;">✓</span> |
| (Per 21 CFR 801.109) | OR |
| | Over-The-Counter Use <span style="text-decoration: underline;"></span> |
| | (Optional Format 1-2-96) |
| | (Division Sign-Off) |
| | Division of Reproductive, Abdominal, ENT,<br>and Radiological Devices |
510(k) Number K991334
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.